The VALEO vascular stent for cardiovascular lesions in children.

Caroline Ovaert1, Deborah Luciano, Jean Gaudart, Ophélie Boulogne, Anass Assaidi, Issam Kammache, Alain Fraisse.   

Abstract

AIMS: The VALEO vascular stent is a stainless steel, pre-mounted, open-cell stent. Redilation to large diameters and low profile are advantages in growing children. Clinical experience is scarce. This study aimed to analyse our experience with the use of the VALEO vascular stent in children. METHODS AND
RESULTS: Between June 2010 and December 2012, 41 VALEO stents were inserted during 30 transcatheter (group 1) and three perioperative procedures (group 2). Data were retrospectively analysed. Median age at implantation was 3.8 years (four days - 23 years), and median weight was 13.3 kg (2.2-53 kg). Indications were: pulmonary artery (n=23), pulmonary vein (n=1) or subhepatic vein (n=1) stenosis, and ductus arteriosus stenting (n=8). Stent placement was achieved in all but one. Acute complications (n=11) included stent dislodgement in two patients and fracture in three patients, with vessel predilatation as a risk factor. Median "stent" follow-up reached 7.83 months (0.4-34.3 months) and included 26 recatheterisations (23 patients, median interval 6.2 months, range 0.2-33 months). Ten stents were redilated. Surgery in six patients (interval 1.9 to 10 months) showed patent endothelialised stents. No "late" type II or III stent fractures were seen.
CONCLUSIONS: The VALEO stent is useful in children. Low radial force is counterbalanced by high flexibility, allowing implantation in distal and tortuous lesions. Early fractures occur. Longer-term follow-up is needed.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25631368     DOI: 10.4244/EIJY15M01_09

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  2 in total

1.  Stent Therapy for Aortic Coarctation in Children <30 kg: Use of the Low Profile Valeo Stent.

Authors:  Sok-Leng Kang; Andrew Tometzki; Demetris Taliotis; Robin Martin
Journal:  Pediatr Cardiol       Date:  2017-08-05       Impact factor: 1.655

2.  Dilating and fracturing side struts of open cell stents frequently used in pediatric cardiac interventions-An in vitro study.

Authors:  Thomas Krasemann; Marco N Kruit; Arthur E van der Mark; Martijn Zeggelaar; Michiel Dalinghaus; Ingrid M van Beynum
Journal:  J Interv Cardiol       Date:  2018-08-08       Impact factor: 2.279