Evaluation of different diameter arterial tubing and arterial cannulae in simulated neonatal/pediatric cardiopulmonary bypass circuits.

Shigang Wang1, Tami Rosenthal, Allen R Kunselman, Akif Ündar.   

Abstract

The objective of this study is to evaluate three different diameters of arterial tubing and three diameters of arterial cannulae in terms of pressure drop, and hemodynamic energy delivery in simulated neonatal/pediatric cardiopulmonary bypass (CPB) circuits. The CPB circuit consisted of a Terumo Capiox Baby FX05 oxygenator (Terumo Corporation, Tokyo, Japan), arterial tubing (1/4 in, 3/16 in, or 1/8 in × 150 cm), and a Medtronic Bio-Medicus arterial cannula (8, 10, or 12 Fr; Medtronic, Inc., Minneapolis, MN, USA). The pseudo patient's pressure was maintained at 50 mm Hg. The circuit was primed using lactated Ringer's solution and heparinized packed human red blood cells (hematocrit 30%). Trials were conducted at different flow rates and temperatures (35 and 28°C). Flow and pressure data were collected using a custom-based data acquisition system. Using 8 Fr arterial cannula at 500 mL/min, small diameter arterial tubing generated higher circuit pressure (294.6 ± 0.1 mm Hg [1/8 in], 213.5 ± 0.5 mm Hg [3/16 in], 208.4 ± 0.4 mm Hg [1/4 in] at 35°C) and arterial line pressure drop (158.3 ± 0.1 mm Hg [1/8 in], 79.6 ± 0.1 mm Hg [3/16 in], 62.1 ± 0.1 mm Hg [1/4 in] at 35°C). Using 10 Fr arterial cannula at 1000 mL/min, pre-oxygenator pressures were 266.8 ± 0.2 mm Hg (3/16 in) and 248.0 ± 0.3 mm Hg (1/4 in); arterial line pressure drops were 111.6 ± 0.0 mm Hg (3/16 in) and 74.0 ± 0.1 mm Hg (1/4 in) at 35°C. When using 12 Fr arterial cannula at 1500 mL/min, preoxygenator pressures reached 324.4 ± 0.3 mm Hg (3/16 in) and 302.5 ± 0.4 mm Hg (1/4 in); arterial line pressure drops were 154.0 ± 0.1 mm Hg (3/16 in) and 92.0 ± 0.2 mm Hg (1/4 in) at 35°C. Pressure drops across arterial line tubing were main CPB circuit pressure drops. High flow rate, hypothermia, small diameter arterial tubing. and arterial cannula created more hemodynamic energy at the preoxygenator site, but energy loss across CPB circuit also increased. Although small diameter (<1/4 in ID) arterial tubing may decrease total CPB priming volume, it also led to significantly higher circuit pressure, higher pressure drop, and more hemodynamic energy loss across CPB circuit. Larger diameter arterial cannula had less pressure drop and allowed more hemodynamic energy delivery to the patient.
Copyright © 2015 International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.

Entities:  

Keywords:  Arterial cannula; Arterial tubing; Cardiopulmonary bypass; Neonates; Pediatrics

Mesh:

Year:  2015        PMID: 25626579     DOI: 10.1111/aor.12446

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  2 in total

1.  Functional Performance of Different Venous Limb Options in Simulated Neonatal/Pediatric Cardiopulmonary Bypass Circuits.

Authors:  Luiz Fernando Caneo; Gregory S Matte; Daniel Peres Guimarães; Guilherme Viotto; Marcelo Mazzeto; Idagene Cestari; Rodolfo A Neirotti; Marcelo B Jatene; Shigang Wang; Akif Ündar; João Chang Junior; Fabio B Jatene
Journal:  Braz J Cardiovasc Surg       Date:  2018 May-Jun

2.  Effects of cardiopulmonary bypass with low-priming volume on clinical outcomes in children undergoing congenital heart disease surgery.

Authors:  Lei Wang; Qiang Chen; Yong Qiang Qiu; Jian Xi Ye; Jian Zhi Du; Xiao Chai Lv; Yan Ting Hou; Liang Wan Chen
Journal:  J Cardiothorac Surg       Date:  2020-05-27       Impact factor: 1.637

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.