High compressive pre-strains reduce the bending fatigue life of nitinol wire.

Shikha Gupta1, Alan R Pelton2, Jason D Weaver1, Xiao-Yan Gong3, Srinidhi Nagaraja4.   

Abstract

Prior to implantation, Nitinol-based transcatheter endovascular devices are subject to a complex thermo-mechanical pre-strain associated with constraint onto a delivery catheter, device sterilization, and final deployment. Though such large thermo-mechanical excursions are known to impact the microstructural and mechanical properties of Nitinol, their effect on fatigue properties is still not well understood. The present study investigated the effects of large thermo-mechanical pre-strains on the fatigue of pseudoelastic Nitinol wire using fully reversed rotary bend fatigue (RBF) experiments. Electropolished Nitinol wires were subjected to a 0%, 8% or 10% bending pre-strain and RBF testing at 0.3-1.5% strain amplitudes for up to 10(8) cycles. The imposition of 8% or 10% bending pre-strain resulted in residual set in the wire. Large pre-strains also significantly reduced the fatigue life of Nitinol wires below 0.8% strain amplitude. While 0% and 8% pre-strain wires exhibited distinct low-cycle and high-cycle fatigue regions, reaching run out at 10(8) cycles at 0.6% and 0.4% strain amplitude, respectively, 10% pre-strain wires continued to fracture at less than 10(5) cycles, even at 0.3% strain amplitude. Furthermore, over 70% fatigue cracks were found to initiate on the compressive pre-strain surface in pre-strained wires. In light of the texture-dependent tension-compression asymmetry in Nitinol, this reduction in fatigue life and preferential crack initiation in pre-strained wires is thought to be attributed to compressive pre-strain-induced plasticity and tensile residual stresses as well as the formation of martensite variants. Despite differences in fatigue life, SEM revealed that the size, shape and morphology of the fatigue fracture surfaces were comparable across the pre-strain levels. Further, the mechanisms underlying fatigue were found to be similar; despite large differences in cycles to failure across strain amplitudes and pre-strain levels, cracks initiated from surface inclusions in nearly all wires. Compressive pre-strain-induced damage may accelerate such crack initiation, thereby reducing fatigue life. The results of the present study indicate that large compressive pre-strains are detrimental to the fatigue properties of Nitinol, and, taken together, the findings underscore the importance of accounting for thermo-mechanical history in the design and testing of wire-based percutaneous implants. Published by Elsevier Ltd.

Entities:  

Keywords:  Bending; Crimp; Endovascular; Fatigue; Nitinol; Pre-strain; Pseudoelastic; Shape memory

Mesh:

Substances:

Year:  2014        PMID: 25625888      PMCID: PMC6553620          DOI: 10.1016/j.jmbbm.2014.12.007

Source DB:  PubMed          Journal:  J Mech Behav Biomed Mater        ISSN: 1878-0180


  2 in total

1.  Comparing Rotary Bend Wire Fatigue Test Methods at Different Test Speeds.

Authors:  Jason D Weaver; Erick J Gutierrez
Journal:  J Mater Eng Perform       Date:  2015-12       Impact factor: 1.819

2.  Designing Better Cardiovascular Stent Materials - A Learning Curve.

Authors:  Irsalan Cockerill; Carmine Wang See; Marcus L Young; Yadong Wang; Donghui Zhu
Journal:  Adv Funct Mater       Date:  2020-11-04       Impact factor: 18.808

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.