Literature DB >> 25615919

Effect of implant divergence on the accuracy of definitive casts created from traditional and digital implant-level impressions: an in vitro comparative study.

Wei-Shao Lin, Bryan T Harris, Eiad N Elathamna, Tamer Abdel-Azim, Dean Morton.   

Abstract

PURPOSE: The purpose of this research was to compare the accuracy of definitive casts created with digital and conventional methods for implants with internal-octagon connections placed parallel or at different angles (15, 30, or 45 degrees).
MATERIALS AND METHODS: Four customized epoxy resin master casts were fabricated with two-implant analogs placed in the posterior mandible with different degrees of divergence. For the conventional (control) group, 10 traditional impressions were taken on each master cast with custom trays, open-tray impression copings, and polyvinyl siloxane; definitive stone casts were poured with type IV dental stone. For the digital group, 10 digital impressions were taken on each master cast with two-piece scannable impression copings and an intraoral digital scanner; definitive milled polyurethane casts were fabricated by the manufacturer. All four master casts and 80 control and test casts were scanned and digitized, and the data sets were compared. Any deviations in measurements between the definitive and corresponding master casts were analyzed statistically.
RESULTS: The amount of divergence between implants did not affect the accuracy of the stone casts created conventionally; however, it significantly affected the accuracy of the milled casts created digitally. A decreasing linear trend in deviations for both distance and angle measurements suggested that the digital technique was more accurate when the implants diverged more. At 0 and 15 degrees of divergence, the digital method resulted in highly significantly less accurate definitive casts. At 30 and 45 degrees of divergence, however, the milled casts showed either no difference or marginal differences with casts created conventionally.
CONCLUSION: The digital pathway produced less accurate definitive casts than the conventional pathway with the tested two-implant scenarios. To ensure passive fit of definitive prostheses, verification devices and casts may be used when materials are produced digitally.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25615919     DOI: 10.11607/jomi.3592

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  12 in total

Review 1.  Precision and practical usefulness of intraoral scanners in implant dentistry: A systematic literature review.

Authors:  Ignacio García-Gil; Jorge Cortés-Bretón-Brinkmann; Jaime Jiménez-García; Jesus Peláez-Rico; María-Jesús Suárez-García
Journal:  J Clin Exp Dent       Date:  2020-08-01

2.  In Vitro Comparison of Three Intraoral Scanners for Implant-Supported Dental Prostheses.

Authors:  Vitória Costa; António Sérgio Silva; Rosana Costa; Pedro Barreiros; Joana Mendes; José Manuel Mendes
Journal:  Dent J (Basel)       Date:  2022-06-15

3.  Comparison of Digital and Conventional Impression Methods by Preclinical Students: Efficiency and Future Expectations.

Authors:  Halenur Bilir; Ceren Ayguzen
Journal:  J Int Soc Prev Community Dent       Date:  2020-08-06

Review 4.  Intraoral scanners in dentistry: a review of the current literature.

Authors:  Francesco Mangano; Andrea Gandolfi; Giuseppe Luongo; Silvia Logozzo
Journal:  BMC Oral Health       Date:  2017-12-12       Impact factor: 2.757

5.  Three-Dimensional Accuracy of Digital Impression versus Conventional Method: Effect of Implant Angulation and Connection Type.

Authors:  Marzieh Alikhasi; Hakime Siadat; Alireza Nasirpour; Mahya Hasanzade
Journal:  Int J Dent       Date:  2018-06-04

Review 6.  The direct digital workflow in fixed implant prosthodontics: a narrative review.

Authors:  George Michelinakis; Dimitrios Apostolakis; Phophi Kamposiora; George Papavasiliou; Mutlu Özcan
Journal:  BMC Oral Health       Date:  2021-01-21       Impact factor: 2.757

Review 7.  Accuracy of Digital Dental Implants Impression Taking with Intraoral Scanners Compared with Conventional Impression Techniques: A Systematic Review of In Vitro Studies.

Authors:  María Isabel Albanchez-González; Jorge Cortés-Bretón Brinkmann; Jesús Peláez-Rico; Carlos López-Suárez; Verónica Rodríguez-Alonso; María Jesús Suárez-García
Journal:  Int J Environ Res Public Health       Date:  2022-02-11       Impact factor: 3.390

8.  Trueness of digital implant impressions based on implant angulation and scan body materials.

Authors:  Jae-Hyun Lee; Jae-Hwi Bae; Su Young Lee
Journal:  Sci Rep       Date:  2021-11-08       Impact factor: 4.379

Review 9.  Trueness and precision of digital implant impressions by intraoral scanners: a literature review.

Authors:  Minoru Sanda; Keita Miyoshi; Kazuyoshi Baba
Journal:  Int J Implant Dent       Date:  2021-07-27

10.  Accuracy of digital impressions versus conventional impressions for 2 implants: an in vitro study evaluating the effect of implant angulation.

Authors:  Jaafar Abduo; Joseph E A Palamara
Journal:  Int J Implant Dent       Date:  2021-07-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.