Literature DB >> 25612055

Impact of a Digital Power Line Filter in the 2-Global-Flash Multifocal Electroretinogram of Glaucoma Patients Compared to Controls.

Anna A Ledolter1,2, Margarita G Todorova1, Andreas Schoetzau1, Anja M Palmowski-Wolfe1.   

Abstract

PURPOSE: To assess the effect of a 50 Hz power line digital filter on the response to a 2-global-flash multifocal electroretinogram (mfERG) in primary open angle glaucoma (POAG) compared to control.
MATERIALS AND METHODS: A 2-global-flash mfERG (VERIS™) was recorded (23 control, 34 POAG). Eight recordings were noise contaminated: 4 control, 4 POAG. Response averages from the central 10° and 7 surrounding groups were analyzed with and without a 50 Hz digital filter for the following mfERG response epochs: direct component (15-45 ms), 1st (45-75 ms) and 2nd (75-105 ms) induced components.
RESULTS: A digital 50 Hz filter had little effect on uncontaminated with noise waveforms but, in noisy recordings, changed the waveform dramatically to resemble uncontaminated waveforms. In controls' 50 Hz-filtered uncontaminated with noise mfERGs differed significantly from unfiltered responses in induced components. Uncontaminated with noise recordings from glaucoma patients did not differ with or without the notch filter (p > 0.1 for all three epochs of mfERG). The mfERG response in the central 10°in glaucoma patients differed significantly from controls, whether the notch filter was used or not (p < 0.001).
CONCLUSIONS: A 50 Hz notch filter allows grossly contaminated waveforms to be analyzed in a meaningful manner. With a 50 Hz filter, glaucoma patients still differed significantly from normal.

Entities:  

Keywords:  50 Hz; digital power line filter; glaucoma; multifocal ERG; notch filter

Mesh:

Year:  2015        PMID: 25612055     DOI: 10.3109/02713683.2014.1002043

Source DB:  PubMed          Journal:  Curr Eye Res        ISSN: 0271-3683            Impact factor:   2.424


  1 in total

1.  Removing mains interference from the mfERG by applying a post-processing digital notch filter: for the good or the bad?

Authors:  Sven P Heinrich
Journal:  Doc Ophthalmol       Date:  2021-11-30       Impact factor: 2.379

  1 in total