Literature DB >> 25611833

On-line haemodiafiltration with auto-substitution: assessment of blood flow changes on convective volume and efficiency.

Francisco Maduell, Raquel Ojeda, Lida Rodas, Nayra Rico, Néstor Fontseré, Marta Arias, Manel Vera, Elisabeth Massó, Mario Jiménez-Hernández, M Florencia Rossi, Giannina Bazán, Josep M Campistol.   

Abstract

INTRODUCTION: On-line haemodiafiltration (OL-HDF) is currently the most effective technique and several randomised studies and meta-analyses have seen a reduction in mortality and an association directly related with convective volume is observed. Blood flow (Qb) limits the infusion rate to 25-33 % and is the main limiting factor for reaching an optimum substitution volume. With the recent incorporation of monitors with auto-substitution systems, the aim of the study was to assess the effect of Qb variations on convective volume and purifying capacity.
MATERIAL AND METHODS: 23 patients, 17 men and 6 women, were included, with an average age of 65.5 ±10 years, time on dialysis 292.2 ± 15 minutes, which were in the OL-HDF programme with the 5008 Cordiax monitor with auto-substitution. Each patient was analysed over 5 sessions in which only the Qb was changed (250, 300, 350, 400 and 450 ml/min). In each session the substitution volume, total convective volume and parameters of dialysis were measured. The concentration of urea (60 Da), creatinine (113 Da), β2-microglobulin (11,800 Da), myoglobin (17,200 Da), prolactin (23,000 Da), α1-microglobulin (33,000 Da) and α1-acid glycoprotein (40,000 Da) in plasma was measured at the start and end of each session in order to calculate the percentage of reduction of these solutes.
RESULTS: The trans-membrane pressure was less, with Qb 250 ml/min. A significant increase in convective volume was observed with the increase in Qb, 23.7, 26.9, 30.2, 32.8 and 35.2 l/session to 250, 300, 350, 400 and 450 ml/min, respectively (P < 0.001), representing a percentage of total purified blood of 33.2, 31.2, 30.2, 28.7 and 27.3 % respectively. The percentages of reduction of urea and creatine progressively increased with Qb, slight differences were observed with β2-microglobulin and myoglobin, and no changes were observed in the larger molecules.
CONCLUSION: For each 50 ml/min increase in Qb, the convective volume increased by between 8 and 12 ml/min. The auto-substitution system strengthens the lowest Qbs in the percentage of convective volume with regards to total purified blood. Qb increases the purifying capacity of small molecules, favouring that of β2-microglobulin and myoglobin, and does not influence molecules of a greater molecular weight. 

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25611833     DOI: 10.3265/Nefrologia.pre2014.Sep.12726

Source DB:  PubMed          Journal:  Nefrologia        ISSN: 0211-6995            Impact factor:   2.033


  2 in total

1.  Consequences of increasing convection onto patient care and protein removal in hemodialysis.

Authors:  Nathalie Gayrard; Alain Ficheux; Flore Duranton; Caroline Guzman; Ilan Szwarc; Fernando Vetromile; Chantal Cazevieille; Philippe Brunet; Marie-Françoise Servel; Àngel Argilés; Moglie Le Quintrec
Journal:  PLoS One       Date:  2017-02-06       Impact factor: 3.240

2.  A reliable method to assess the water permeability of a dialysis system: the global ultrafiltration coefficient.

Authors:  A Ficheux; N Gayrard; F Duranton; C Guzman; I Szwarc; F Vetromile; P Brunet; M F Servel; A Argilés
Journal:  Nephrol Dial Transplant       Date:  2017-02-01       Impact factor: 5.992

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.