Lumbar nerve rootlet entrapment by an iatrogenically spliced percutaneous intra-thecal lumbar cerebrospinal fluid catheter.

James J Yue1, Carlos A Castro1, David Scott1.   

Abstract

BACKGROUND: Complications associated with the use of percutaneous intra-thecal lumbar indwelling spinal catheters include infection, hematoma, neurologic dysfunction, and persistent undesired retention among others. A case of iatrogenic splicing associated with neurologic dysfunction with the use of a percutaneous intra-thecal indwelling spinal catheter is presented in this study.
METHOD: Single case study review.
RESULTS: Review of case materials indicate Y pattern splicing/fragmentation of an indwelling intra-thecal catheter causing neurologic dysfunction and resistance to removal during attempted removal. Pain and weakness were evident soon after insertion of the catheter and were amplified with attempted catheter removal. Computed tomography revealed a double dot sign on axial view and a Y appearance on sagittal view. Surgical findings revealed entrapment of nerve rootlets in the axilla of the spliced catheter.
CONCLUSIONS: Splicing/fragmentation causing neurologic dysfunction as well as catheter retention is described as a potential complication of intra-thecal indwelling cerebrospinal fluid catheters. A symptom of fragmentation of a catheter may include neurologic dysfunction including pain and weakness of a lumbar nerve root. If resistance is experienced upon attempted catheter removal, with or without associated neurologic dysfunction, further attempts at removal should not be attempted. In those cases in which pain and/or lumbar weakness are evident post catheter placement and/or following attempted removal, computed tomography should be performed. If fragmentation of a catheter is evident on CT scan, spinal surgical consultation should be obtained. Recommended spinal surgical intervention includes an open durotomy and visualization of catheter fragments and nerve rootlets and removal of catheter fragments.
Copyright © 2015 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  Complication; Double dot sign; Fragmentation; Intra-thecal spinal catheter; Retention; Splicing

Year:  2015        PMID: 25600724      PMCID: PMC4336429          DOI: 10.1016/j.ijscr.2015.01.012

Source DB:  PubMed          Journal:  Int J Surg Case Rep        ISSN: 2210-2612


Introduction

Intra-thecal cerebrospinal catheters are often used for anesthetic as well as for cerebrospinal decompressive purposes. Complications such as intra-pleural migration, local hematoma, intra-cranial subdural hematoma, cranial nerve dysfunction, placement of catheter in pulmonary vessels, infection, neurologic changes, cerebrospinal fluid leak and associated headaches, pseudo meningocele formation, tip granulomas, complete fracturing, catheter migration and others have been described [1-16]. To the best of our knowledge, the complication of intra-thecal splicing and nerve root entrapment with subsequent neurologic dysfunction and catheter retention has not been described.

Case presentation

V.M. is an 82 year old female who presented to the emergency department for the worsening of chronic lower back pain over a 3 day period. On the day of admission, her pain was unrelieved with pain medications and the patient remained prostrate. Subsequent imaging and workup revealed an aneurysm of the aorta from her proximal descending thoracic aortic to her iliac bifurcation with aneurysm of the abdominal aorta (AAA) portion from 8.5 cm to 10 cm. Patient was hemodynamically stable and was admitted to the vascular surgery service. The patient declined open repair of her aneurysm defects. Instead, she elected endovascular repair with initial repair of the descending thoracic aortic aneurysm (TAA) portion and staged extension with a fenestrated abdominal aortic stent graft. After admission, the patient was brought to the operating room for endovascular repair followed by placement of a lumbar drain (EDM Lumbar Catheter, Closed Tip Medtronic Neurosurgery, Goleta, CA, USA) at the L3–4 interspace by the anesthesia service to decrease the potential complication of paraplegia increasing the perfusion pressure to the spinal cord [17]. On post-op day 1, the patient noted pain in the right anterior thigh as well as right hip flexor weakness. Neurology was consulted with suspicion for right plexopathy. Her motor exam was notable for 4/5 weakness of right hip flexor (psoas L1–2), 4/5 right quadriceps (femoral nerve L3–4) and 4/5 right hamstring strength (sciatic nerve S1). No clear weakness of adductor of hip (obturator L2–3) or abductor of hip (gluteus medius S1). No weakness of right ankle dorsiflexors, inverters, everters, plantar flexors or right EHL. No weakness of the left leg lower extremity was noted. Pulses were dopplerable bilaterally. Sensation was normal and symmetric to light touch of both legs L2-S1. Subjective decreased vibration on the right patella. Symmetric vibratory sensation was present on the great toes, medial and lateral ankle malleoli and both shins (proximal and distal). Computed tomography examination at that time showed no evidence of retroperitoneal or psoas hematoma. On post-op day 2, the patient’s right leg weakness slightly improved but on the morning of post-op day 3, an attempt at catheter removal was unsuccessful due to resistance. During the attempted removal, the patient experienced severe burning down the anterior lateral aspect of her right thigh and lower leg. The right hip flexor weakness also worsened. No further attempts of catheter removal were attempted. Subsequently, the patient began having episodes of bowel and bladder incontinence. No long tract findings were evident. The spine surgery service was consulted and obtained an MRI that showed no discernible significant epidural hematoma from L3 down to S1 (Fig 1). However from L3 proximal, it was difficult to appreciate due to metal artifact.
Fig. 1

Sagittal MRI of lumbar spine.

A CT scan was ordered and showed a catheter that appeared spliced and coiled within the thecal sac (Fig. 2A and B-3). Axial view of the CT scan revealed a double dot appearance representing a section through a coiled catheter (Fig. 3). On POD 4, the patient was brought to the operating room for removal of the retained lumbar spinal catheter (Fig 4).
Fig. 2

Sagittal view of lumbar spine.

Note Y slice appearance in Fig. 2A.

Fig. 3

Axial CT scan demonstrating double dot finding due to coiled catheter.

Fig. 4

View of catheter prior to removal.

An incision approximately 6 cm in length was made proximally and distally centered over the L3–L4 spinous processes, careful dissection down to the intervertebral interlaminar space was performed. The distal aspect of L3 and the proximal aspect of L4 of the spinous process were removed. The ligamentum flavum and lamina were also removed. The catheter was followed down to the dura and a 4 mm durotomy was made in a distal direction below the catheter. A nerve hook placed within the thecal sac and control of the spliced portion of the catheter was obtained. A small nerve rootlet was identified in the Y portion of the spliced catheter. Once the spliced portion of the catheter was released from the rootlet, the remaining Y portion and the contiguous distal aspect of the catheter was removed without difficulty (Fig 5). Lastly, the dura was repaired in watertight sealed manner.
Fig. 5

Gross examination of catheter after removal from lumbar spine.

Patient was discharged to rehab on 4 days after catheter removed with slight anterior thigh pain and without significant recovery of her right thigh strength. The patient's thigh strength was noted to be normal by 3 months after index surgery with no CSF complications. Patient recovered full motor, sensory as well as bowel and bladder control after removal of the catheter. Patient has minimal residual pain.

Discussion

Complications associated with the use intra-thecal catheters are infreqent [2,16]. A catheter which has been compromised and entrapped within the thecal sac has not been described to the best of our knowledge. We hypothesize that the catheter was inadvertently spliced as a result of either a pulling motion of the catheter over the tip of the insertion needle or advancement of the needle into the catheter. In either potential mechanism, splicing of the catheter may have been obviated if the bevel of the needle was positioned open toward the dorsal aspect of the spine. Patients who experience radicular pain and/or other neurologic findings should be evaluated for catheter fragmentation/splicing using a CT scan. Attempted removal should not be performed prior to CT assessment in those patients who are experiencing post catheter insertion or pain during attempted removal. Successful removal of catheter components can be achieved with open laminotomy, controlled durotomy, catheter removal and dural repair.

Conflict of interest

No conflicts.

Funding

No funding.

Ethical approval

None.

Author contribution

James Yue – Main author/surgeon on case. C Castro – Contributing author. D Scott – Contributinig author.

Consent

No confidential information disclosed in paper.

Gurantor

James J. Yue.
  17 in total

1.  Cranial nerve palsy and intracranial subdural hematoma following implantation of intrathecal drug delivery device.

Authors:  C A Velarde; R E Zuniga; R F Leon; S E Abram
Journal:  Reg Anesth Pain Med       Date:  2000 Jan-Feb       Impact factor: 6.288

2.  Rising complication rates after intrathecal catheter and pump placement in the pediatric population: analysis of national data between 1997 and 2006.

Authors:  Narayana C Varhabhatla; Zhiyi Zuo
Journal:  Pain Physician       Date:  2012 Jan-Feb       Impact factor: 4.965

3.  Cerebrospinal fluid drainage to reverse paraplegia after endovascular thoracic aortic aneurysm repair.

Authors:  K Tiesenhausen; W Amann; G Koch; K A Hausegger; P Oberwalder; B Rigler
Journal:  J Endovasc Ther       Date:  2000-04       Impact factor: 3.487

4.  Case report: inadvertent intrathecal placement of a pulmonary artery catheter introducer.

Authors:  Marcel Schepers; Marcel Vercauteren; Dina De Bock; Inez Rodrigus; David Vanderplanken; Michael Camerlinck
Journal:  Anesth Analg       Date:  2012-08-10       Impact factor: 5.108

5.  Catheter migration after implantation of an intrathecal baclofen infusion pump for severe spasticity: a case report.

Authors:  Tung-Chou Li; Mei-Hsiu Chen; Jing-Shan Huang; Jun-Yeen Chan; Yaun-Kai Liu; Ming-Hong Chen
Journal:  Kaohsiung J Med Sci       Date:  2008-09       Impact factor: 2.744

6.  Cranial migration of a baclofen pump catheter associated with subarachnoid hemorrhage: case report.

Authors:  Peter Nakaji; Giac D Consiglieri; Mark P Garrett; Nicholas C Bambakidis; Andrew G Shetter
Journal:  Neurosurgery       Date:  2009-12       Impact factor: 4.654

Review 7.  Serious complications associated with external intrathecal catheters used in cancer pain patients: a systematic review and meta-analysis.

Authors:  Daniel Aprili; Oliver Bandschapp; Christoph Rochlitz; Albert Urwyler; Wilhelm Ruppen
Journal:  Anesthesiology       Date:  2009-12       Impact factor: 7.892

8.  Is postoperative intrathecal catheter use associated with central nervous system infection?

Authors:  B K Bevacqua; A V Slucky; W F Cleary
Journal:  Anesthesiology       Date:  1994-06       Impact factor: 7.892

9.  Retraction of an intrathecal baclofen infusion catheter following suprapubic cystotomy: a case report.

Authors:  Frank M J Martens; Diederik M Somford; Dirk H J M van Kuppevelt; Monica J M van den Burg; John P F A Heesakkers
Journal:  J Rehabil Med       Date:  2009-01       Impact factor: 2.912

10.  Distal baclofen pump catheter migration associated with far-lateral paraspinal surgical placement.

Authors:  Andrew J Fabiano; Carolyn Doyle; Robert J Plunkett
Journal:  Neuromodulation       Date:  2009-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.