Evaluation of the maintenance of the apical limit during instrumentation with hybrid equipment in rotary and reciprocating modes.

Bruno Carvalho de Vasconcelos1, Luciana Maria Arcanjo Frota2, Tamara de Abreu Souza2, Ricardo Affonso Bernardes3, Marco Antonio Hungaro Duarte4.   

Abstract

INTRODUCTION: The objective of the present study was to evaluate ex vivo the maintenance of the apical limit during instrumentation with the Root ZX (J Morita, Tokyo, Japan) and VDW Gold (VDW GbmH, Munich, Germany) hybrid devices in rotary and reciprocating modes and determine the possible function variations of different preparation levels (0.0 mm and -1.0 mm).
METHODS: Seventy-two human uniradicular mandibular premolars, which had their coronal access performed and their apical foramen (AF) standardized (200 μm) were used. After their randomized division, the root canals were chemically mechanically prepared with #F4 ProTaper (Dentsply Maillefer, Ballaigues, Switzerland) and #R40 Reciproc (VDW GbmH) instruments. At the end of the instrumentation, the last instrument was fixed, and 4.0 mm of the apical portion was worn, permitting the measurement of the distance between the instrument tips and the AF.
RESULTS: The determined precision values of 0.0 and -1.0 mm were 100% and 33.4% (Root), 100% and 41.6% (VDW Gold with rotary movement) and 100% and 33.3% (VDW Gold with reciprocating movement), respectively, with a margin of ±0.5 mm. No statistical differences were found with respect to the comparisons of every level of preparation. However, significant differences were found when comparing the groups with rotary function at the different levels, with a higher precision for the groups that reached the AF.
CONCLUSIONS: It was concluded that both devices, independent of the function used, were more efficient in the maintenance of the apical limit when used until the AF; however, such reliability was compromised when -1.0 mm from the apical limit was established.
Copyright © 2015 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Electronic foramen locator; root canal preparation; working length determination

Mesh:

Year:  2015        PMID: 25595465     DOI: 10.1016/j.joen.2014.11.029

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  2 in total

1.  An ex vivo comparison of working length determination by three electronic root canal length measurement devices integrated into endodontic rotary motors.

Authors:  Manal Maree Ali; Ronald Wigler; Shaul Lin; Arieh Y Kaufman
Journal:  Clin Oral Investig       Date:  2016-07-09       Impact factor: 3.573

2.  Multidimensional Analysis of Curved Root Canal Preparation Using Continuous or Reciprocating Nickel-titanium Instruments.

Authors:  Iussif Mamede-Neto; Álvaro Henrique Borges; Ana Helena Gonçalves Alencar; Marco Antonio Hungaro Duarte; Manoel Damião Sousa Neto; Carlos Estrela
Journal:  Open Dent J       Date:  2018-01-29
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.