Dose calculation accuracy of different image value to density tables for cone-beam CT planning in head & neck and pelvic localizations.

Anaïs Barateau1, Christopher Garlopeau1, Audrey Cugny1, Bénédicte Henriques De Figueiredo1, Charles Dupin1, Jérôme Caron1, Mikaël Antoine2.   

Abstract

PURPOSE: We aimed to identify the most accurate combination of phantom and protocol for image value to density table (IVDT) on volume-modulated arc therapy (VMAT) dose calculation based on kV-Cone-beam CT imaging, for head and neck (H&N) and pelvic localizations.
METHODS: Three phantoms (Catphan(®)600, CIRS(®)062M (inner phantom for head and outer phantom for body), and TomoTherapy(®) "Cheese" phantom) were used to create IVDT curves of CBCT systems with two different CBCT protocols (Standard-dose Head and Standard Pelvis). Hounsfield Unit (HU) time stability and repeatability for a single On-Board-Imager (OBI) and compatibility of two distinct devices were assessed with Catphan(®)600. Images from the anthropomorphic phantom CIRS ATOM(®) for both CT and CBCT modalities were used for VMAT dose calculation from different IVDT curves. Dosimetric indices from CT and CBCT imaging were compared.
RESULTS: IVDT curves from CBCT images were highly different depending on phantom used (up to 1000 HU for high densities) and protocol applied (up to 200 HU for high densities). HU time stability was verified over seven weeks. A maximum difference of 3% on the dose calculation indices studied was found between CT and CBCT VMAT dose calculation across the two localizations using appropriate IVDT curves. One IVDT curve per localization can be established with a bi-monthly verification of IVDT-CBCT.
CONCLUSIONS: The IVDT-CBCTCIRS-Head phantom with the Standard-dose Head protocol was the most accurate combination for dose calculation on H&N CBCT images. For pelvic localizations, the IVDT-CBCTCheese established with the Standard Pelvis protocol provided the best accuracy.
Copyright © 2015 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.

Keywords:  Hounsfield units; IVDT curve; kV-CBCT

Mesh:

Year:  2015        PMID: 25595131     DOI: 10.1016/j.ejmp.2014.12.007

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  2 in total

1.  Trajectory log analysis and cone-beam CT-based daily dose calculation to investigate the dosimetric accuracy of intensity-modulated radiotherapy for gynecologic cancer.

Authors:  Yohei Utena; Jun Takatsu; Satoru Sugimoto; Keisuke Sasai
Journal:  J Appl Clin Med Phys       Date:  2021-01-10       Impact factor: 2.102

2.  Impact of Acquisition Protocols on Accuracy of Dose Calculation Based on XVI Cone Beam Computed Tomography.

Authors:  Slimani Souleyman; Khalal Dorea Maria; Tyeb Cheikh; Khalal-Kouache Karima
Journal:  J Med Phys       Date:  2021-08-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.