Kinematic analysis of a posterior-stabilized knee prosthesis.

Zhi-Xin Zhao, Liang Wen, Tie-Bing Qu1, Li-Li Hou, Dong Xiang, Jia Bin.   

Abstract

BACKGROUND: The goal of total knee arthroplasty (TKA) is to restore knee kinematics. Knee prosthesis design plays a very important role in successful restoration. Here, kinematics models of normal and prosthetic knees were created and validated using previously published data.
METHODS: Computed tomography and magnetic resonance imaging scans of a healthy, anticorrosive female cadaver were used to establish a model of the entire lower limbs, including the femur, tibia, patella, fibula, distal femur cartilage, and medial and lateral menisci, as well as the anterior cruciate, posterior cruciate, medial collateral, and lateral collateral ligaments. The data from the three-dimensional models of the normal knee joint and a posterior-stabilized (PS) knee prosthesis were imported into finite element analysis software to create the final kinematic model of the TKA prosthesis, which was then validated by comparison with a previous study. The displacement of the medial/lateral femur and the internal rotation angle of the tibia were analyzed during 0-135° flexion.
RESULTS: Both the output data trends and the measured values derived from the normal knee's kinematics model were very close to the results reported in a previous in vivo study, suggesting that this model can be used for further analyses. The PS knee prosthesis underwent an abnormal forward displacement compared with the normal knee and has insufficient, or insufficiently aggressive, "rollback" compared with the lateral femur of the normal knee. In addition, a certain degree of reverse rotation occurs during flexion of the PS knee prosthesis.
CONCLUSIONS: There were still several differences between the kinematics of the PS knee prosthesis and a normal knee, suggesting room for improving the design of the PS knee prosthesis. The abnormal kinematics during early flexion shows that the design of the articular surface played a vital role in improving the kinematics of the PS knee prosthesis.

Entities:  

Mesh:

Year:  2015        PMID: 25591565      PMCID: PMC4837841          DOI: 10.4103/0366-6999.149205

Source DB:  PubMed          Journal:  Chin Med J (Engl)        ISSN: 0366-6999            Impact factor:   2.628


INTRODUCTION

Osteoarthritis of the knee is a very common joint disease that primarily afflicts older people and leads to a series of symptoms, including pain, decreased range of motion (ROM), and loss of function, among others. Over the past 30 years, the total knee arthroplasty (TKA) procedure has proven to be one of the most effective surgeries for patients with degenerative knee disease, and TKA is accepted as the gold standard for the treatment of osteoarthritis.[12] According to one report, the number of annual TKA procedures in the United States is about 300,000–500,000. Based on the proportional population, an estimate of the number of annual TKA procedures in China is about 1,000,000–1,500,000. The clinical statistics over the past decade show that 90% of patients obtain good clinical results after receiving a TKA.[34] However, good clinical results do not necessarily indicate high postoperative satisfaction rates in the patients.[56] Abnormal knee kinematics is an important factor that influences postoperative satisfaction rates. For example, deficient rollback of the femur or abnormal rotation of the tibia will lead to a decreased range of flexion angles for patients after TKA.[789] Therefore, the in vivo kinematics after an operation is a key element for evaluating the design of knee prostheses. The posterior cruciate ligament (PCL) must be excised intraoperatively during TKA with a posterior-stabilized (PS) knee prosthesis. The PS knee prosthesis substitutes a cam and postinteraction for the stability normally offered by the PCL. This interaction can also help with the rollback of the femur and control the backward movement of the tibia, reducing instability during flexion. Several studies on in vivo knee kinematics using radiologic analysis have been published,[10111213] but studies that apply computer simulation to analyze and predict knee kinematics are limited. The in vivo knee kinematics studies using radiologic analysis show that with increasing knee flexion, the lateral femur rollback tends to exceed that of the medial femur and the internal rotation of the tibia. This is called “screw-home.” To improve the postoperative satisfaction rates of patients, knee prostheses should reproduce the “screw-home” effect. In the present study, a model of normal knee kinematics was created. We simulated the entire process of TKA using this model, and analyzed and predicted the in vivo knee kinematics of PS knee prosthesis. The results of the study may provide a useful kinematics reference for the design of knee prostheses.

METHODS

Establishing the three-dimensional model of normal knee kinematics

The sample was a healthy, anticorrosive female cadaver (age: 40 years; height: 164 cm; weight: 50 kg). Computed tomography (CT) scans (Siemens SOMATOM Emotion 16, Siemens Ltd., Munich, Germany) were taken from 5 cm above the tip of the femoral head to the ankle joint. The basic settings during the CT scan included: A scan interval of 3 mm, the apparent plane as the main plane, and a scanning resolution of 512 × 512 pixels. In addition, the bony structures 10 cm above and below the knee joint line were scanned using a magnetic resonance imaging (MRI) device (Siemens Avanto, Siemens Ltd., Munich, Germany). The interval was 0.5 mm, and the scanning resolution was 512 × 512 pixels. All the data obtained from the CT and MRI scans were saved as Digital Imaging and Communications in Medicine format files. After inputting these data, the medical modeling software Mimics 13.0 (Materialise Ltd., Leuven, Belgium) was used to establish the model of normal knee kinematics. The CT images were used to establish a model of the entire lower limbs, including the femur, tibia, patella, and fibula. The normal attenuation coefficient range of human skeletal bone is 226–1701 Hu; this threshold range was chosen to establish the mask in the setting of Mimics 13.0. The different colors indicate the masks of different bone models. Manual division and repair were applied to the scanned images for processing. First, a partial division was made for the scanned images of the connection structures among the femur, tibia, and fibula. Next, the “region growth” function of Mimics was used to further divide the selected image, and different bone structures were separated. Finally, the “three-dimensional (3D) calculation” function of Mimics was used to reconstruct each individual mask. After this process, the 3D bone structures were clearly visualized [Figure 1a].
Figure 1

(a) The three-dimensional (3D) bone model of lower limb. (b) The 3D model of cartilage and meniscus. (c) The 3D model of normal knee joint.

(a) The three-dimensional (3D) bone model of lower limb. (b) The 3D model of cartilage and meniscus. (c) The 3D model of normal knee joint. Magnetic resonance imaging scans were used to establish the model of the distal femur cartilage and medial and lateral menisci. The scanned MRI images were imported into Mimics 13.0. The same image processing described above was used to divide, repair, and reconstruct 3D models of the distal femur cartilage and menisci [Figure 1b]. Because the coordinate frame of the CT images was different than that of the MRI images, 3D rectification was performed to reconstruct the structure of the CT and MRI images. This helped the reconstructed 3D bone model based on the CT images match to the reconstructed model based on the MRI images. Our main idea was to perform a coordinate transformation on the 3D bone model based on the CT images using Geomagic software (Parametric Technology Corporation, Needham, MA, USA), and adapt the new coordinate frame to the frame of the MRI images. After splicing the CT and MRI models, the new data were imported into the 3D design software PRO/E (Parametric Technology Corp., Needham, MA, USA), finally establishing an intact 3D model of the knee joint model of knee joint was established [Figure 1c].

Simulation of the ligaments around the normal knee joint and muscle strength

The anterior cruciate ligament (ACL), PCL, medial collateral ligament (MCL), and lateral collateral ligament (LCL) were simulated in the present study. Springs with different stiffness coefficients (SCs) were used to simulate the force of each ligament. Based on the findings of Abdel-Rahman and Hefzy,[14] the ACL and PCL can be simulated using two branches: An anterior and a posterior branch. The MCL can be simulated using three branches: Anterior, deep, and oblique branches, while the LCL can be simulated using a single branch. The piecewise function used to define the force of ligaments was: In the above function, Fj represents the force of ligament; K1j and K2j are the SCs of the spring unit; L0j is the initial length of the ligament; Lj is the stretched length; ε1 is defined as 0.03; and εj is the deformation of the ligament. To precisely simulate the forces of the ligaments, a hypothesis was applied to the spring unit: The force of the ligaments is a tensile stress with no force during compression. The SC and compensation coefficient (CC) of different ligaments [Table 1] prevented the ligaments around the knee joint from experiencing compressive forces. The CC was further interpreted in a study by Blankevoort et al.[15] Each ligament was identified by the CC in each test, which helped to simulate the forces of the ligaments. To better simulate the physiological mechanisms within the knee joint, the patellar tendon and quadriceps tendon were also simulated. Both the patellar and quadriceps tendons consist of medial and lateral branches, which can also be simulated using spring units.[16] The SC of the patellar tendon is 1000 N/mm,[17] and that of the quadriceps tendon is 521 N/mm.[18] The attachment points of all the ligaments and the patellar and quadriceps tendons were identified by experienced orthopedic surgeons after repeated observations of the bone model and MRI data.
Table 1

The SC and CC of different ligaments

K1 (N/mm)K2 (N/mm)CC
ACL-Anterior22.4883.151
ACL-Posterior26.2783.151.051
PCL-Anterior31.26125.001.004
PCL-Posterior19.2960.001.05
MCL-Anterior10.0091.250.94
MCL-Oblique5.0027.861.031
MCL-Deep5.0021.071.049
LCL10.0072.221.05

SC: Stiffness coefficient; CC: Compensation coefficient; ACL: Anterior cruciate ligament; PCL: Posterior cruciate ligament; MCL: Medial collateral ligament; LCL: Lateral collateral ligament.

The SC and CC of different ligaments SC: Stiffness coefficient; CC: Compensation coefficient; ACL: Anterior cruciate ligament; PCL: Posterior cruciate ligament; MCL: Medial collateral ligament; LCL: Lateral collateral ligament.

Establishing the kinematics model of the normal knee

The data from the 3D model of the normal knee joint were imported into MD Adams R3 software (MSC Software, Newport Beach, CA, USA), and the properties of different bone structures were assigned. Next, the simulation unit was established to simulate the forces of the ligaments and muscle strength based on the settings described above. Because the MD Adams software could not simulate deformation, we designed the study to better simulate the real physiology. First, the medial and lateral menisci were divided into anterior and posterior parts based on the location of the centroid. These two parts of the menisci were then connected to the spring unit, and a damping coefficient of 0.5 N s/mm[19] was applied to the spring unit to ensure that the distal femur cartilage was in contact with the meniscus at the same time. The rotational axis during flexion of the femur (femur flexion center [FFC] axis) was based on the medial and lateral FFC. The rules of the coordinate frame were defined as the FFC axis as the X-axis, the mechanical axis of the lower limb as the Z-axis, and the Y-axis was identified using the “right-hand rule.” Together, the X, Y, and Z axes constitute a cartesian coordinate system in a certain space. As flexion increased, the displacements of the medial and lateral femur (forward and backward) were identical to the displacements of the medial and lateral femur along the Y-axis. The internal rotation of the tibia was reflected by the rotation around the Z-axis. Thus, the model of normal knee joint kinematics was established [Figure 2a]. With this model, the knee could be simulated within a ROM of 0–135° of flexion. The output of the model consisted of the displacements of the medial and lateral femur and the internal rotation angle of the tibia.
Figure 2

(a) The kinematics model of normal knee. (b) The kinematics model total knee arthroplasty prosthesis.

Establishing the kinematics model of the total knee arthroplasty prosthesis

A set of PS knee prostheses was selected as the study objects. A 3D model derived from the 3D scanner information was created, and this model was developed using Geomagic software. The entire TKA surgery procedure was simulated using the 3D model of the normal knee with the PS knee prosthesis implanted in it. The 3D knee model of the TKA prosthesis was then imported into MD Adams R3 software. The femur prosthesis was fixed on the femur; the tibia prosthesis was fixed on the tibia plateau, and the polyethylene inlay was well-seated on the tibia prosthesis. The femur was set in contact with the high molecular weight polyethylene inlay. In addition, the properties of the ligaments and muscle strength were set according to the methods described above, as well as the settings of the rotational axis and the coordinate system of measurement parameters. Given that the meniscus was already excised during the TKA, it was not necessary to simulate the parameters of the meniscus. Thus, the final kinematics model of the TKA prosthesis was established [Figure 2b], simulating a ROM (0–135° of flexion). The output of the model consisted of the displacements of the medial and lateral femur and the internal rotation angle of the tibia. (a) The kinematics model of normal knee. (b) The kinematics model total knee arthroplasty prosthesis.

RESULTS

Validation of the model of normal knees

Validation was performed to verify the accuracy of the target model by analyzing the trends of the output data and the values measured. There are two methods of validation: Experimental verification and literature comparison. Because it was difficult to find in vivo kinematics data, the method of literature comparison was used to validate the model in the present study. The displacements of the medial and lateral femur predicted by the normal knee model were compared with data from an in vivo study by Johal et al.[13] Because only the displacements of the medial and lateral femur were reported in Johal's study, we had to use the data for the model validation. The data comparison [Figure 3] between the simulated model of the normal knee and the results of the in vivo study shows that both the trends of the output data and the values measured, which were derived from the normal knee kinematics model, are very close to the results from Johal's in vivo study. Therefore, this model can be used for further analyses.
Figure 3

The data comparison between the simulated model of normal knee and in vivo study.

The data comparison between the simulated model of normal knee and in vivo study.

The kinematic characteristics of the posterior-stabilized knee prosthesis

Analysis was performed on the kinematic characteristics of the PS knee prosthesis based on the kinematics model of the TKA prosthesis and three other parameters: The displacement of the medial femur, the displacement of the lateral femur, and the internal rotation angle of the tibia. The kinematics characteristics of the normal knee and the PS knee prosthesis are compared below. The maximum displacement of the medial femur after implantation of the PS knee prosthesis was about 5.1 mm compared with 3.1 mm in a normal knee. This result indicates that the PS knee prosthesis underwent an abnormal forward displacement compared with the normal knee [Figure 4a]. The backward displacement of the lateral femur with the PS knee prosthesis was nearly 0 mm for flexion angles <90°, and the maximum backward displacement was about 10 mm for flexion angles more than 90°. The lateral femur of the normal knee experiences a continuous “rollback” movement, and the maximum backward displacement is about 21.1 mm. This finding indicates that the PS knee prosthesis has an insufficient, or insufficiently aggressive, “rollback” compared with the lateral femur of the normal knee [Figure 4b]. The internal rotation angle of the tibia in the PS knee prosthesis was <7° for flexion angles <105°. There was mild reverse rotation (about 1°) when the knee flexion angle was between 60° and 105°. This reverse rotation angle increased to 20.6° when the flexion angle was over 105°, while the internal rotation angle of the tibia in normal knees continuously increases until 22.3°. This finding indicates that the tibia, after implantation of a PS knee prosthesis, has insufficient internal rotation compared with the tibia in normal knees, and that there is a certain degree of reverse rotation during flexion [Figure 4c].
Figure 4

(a) The comparison on the displacement of medial femur between normal knee and posterior-stabilized knee prosthesis. (b) The comparison on the displacement of lateral femur between normal knee and posterior-stabilized knee prosthesis. (c) The comparison on the internal rotation angle of the tibia between normal knee and posterior-stabilized knee prosthesis.

(a) The comparison on the displacement of medial femur between normal knee and posterior-stabilized knee prosthesis. (b) The comparison on the displacement of lateral femur between normal knee and posterior-stabilized knee prosthesis. (c) The comparison on the internal rotation angle of the tibia between normal knee and posterior-stabilized knee prosthesis.

DISCUSSION

The design concept of the PS knee prosthesis includes the use of the cam and postinteraction as a substitute for the function of the PCL. The contact between the post and cam in most PS knee prostheses on the market occurs when the knee flexion angle is about 70–100°. Before the cam and postmechanism were created, the entire knee movement was controlled by the MCL, LCL, and the inlay articular surface. This suggests that the design of the inlay articular surface is critical during the early flexion of the knee joint, when the inlay acts as a substitute for the roles of the ACL and PCL. According to the results of several published in vivo studies on knee kinematics, the femur has an external rotation relative to the tibia with the medial femur being the center of rotation.[1013] The displacement of the medial femur was minimal, while the lateral femur had a backward displacement, which led to the internal rotation of the tibia. An in vivo kinematics study on patients after TKA[8] indicated that about 68.7% of patients after TKA with PS knee prostheses showed rotation with the center located on the medial side, while 22.7% of the patients showed the opposite trend with the center of rotation on the lateral side. A study by Banks and Hodge[20] indicated that weakly constrained prostheses tended to take the lateral side as the center of rotation, and the kinematics were closely related to the level of constraint of the knee prosthesis. Nakayama et al.[21] reported that fixed bearing PS knee prostheses perform excellently with regards to the internal rotation of the tibia but that the contact force of the cam and postmechanism was quite high. Nakayama et al.[21] believed that both the cam and post and the inlay articular surface were important design points to avoid impingement and edge loading. In addition, they thought that good designs should increase the security and stability of the joint during higher angle flexion. In the present study, a kinematics model of the normal knee was established that included the distal femur, cartilage of the distal femur, proximal tibia, medial and lateral menisci, proximal fibula, patella, and ligaments around knee. Analysis of the kinematic characteristics of the model TKA prosthesis was performed based on the normal knee model kinematics. Three parameters were used for analysis: The displacement of the medial femur, the displacement of the lateral femur, and the internal rotation angle of the tibia. The displacements of the medial and lateral femur predicted by the normal knee model were compared with the data from an in vivo study by Johal et al.[13] We found that both the trends of the output data and the measured values predicted by the normal knee kinematics model were quite similar to the results of the in vivo study by Johal et al.[13] Thus, the validation of the model was successful, suggesting that it could be used for further analyses. In the present study, the TKA surgical process was simulated based on the kinematics model of the normal knee and the established kinematics model of the TKA PS knee prosthesis. The analysis of the kinematics characteristics was performed after the TKA. The results indicate that several differences remain between the kinematics of the PS knee prosthesis and the normal knee. The key differences were that the PS knee prosthesis had an abnormal forward displacement, the “rollback” of lateral femoral condyle was insufficient, and that the tibia displayed an insufficient internal rotation during early flexion and a certain degree of reverse rotation. In particular, the backward displacement of the lateral femur in the PS knee prosthesis was nearly 0 mm when the flexion angle was <90°, which indicated a tendency for the joint to take the lateral side as the center of rotation. Together, these findings indicate that there remains room for improvement in the design of PS knee prostheses. The abnormal kinematics predicted during early flexion showed that the design of the articular surface plays a vital role in improving the kinematics of PS knee prostheses. The simulated kinematics model of the knee joint in the present study can be used as the basis for assessing further improvements on the design of cruciate-retaining (CR) and PS knee prostheses. Several limitations of the present study should be mentioned. The set of PS knee prostheses that were used here are not necessarily representative of all the different styles of commercial posterior substituting products. Similarly, a single sample female cadaver cannot be considered representative of the anatomical features of all patients who undergo knee prosthetic procedures. In addition, certain surrounding soft tissues, including the joint capsule, hamstrings, and gastrocnemius could not be involved in our model simulation. All of these problems should be addressed in follow-up studies. Our future work will focus on several points. We should make better use of the kinematics model established in this study. In addition, our future work should provide design proposals for the development of CR and PS knee prostheses aimed at restoring physiological kinematics after TKA.
  21 in total

1.  Three-dimensional dynamic simulation of total knee replacement motion during a step-up task.

Authors:  S J Piazza; S L Delp
Journal:  J Biomech Eng       Date:  2001-12       Impact factor: 2.097

2.  The effect of design variables of condylar total knees on the joint forces in step climbing based on a computer model.

Authors:  C H Yu; P S Walker; M E Dewar
Journal:  J Biomech       Date:  2001-08       Impact factor: 2.712

3.  Twenty-year evaluation of the New Jersey LCS Rotating Platform Knee Replacement.

Authors:  Frederick F Buechel; Frederick F Buechel; Michael J Pappas; Jerry Dalessio
Journal:  J Knee Surg       Date:  2002       Impact factor: 2.757

4.  Femoral rollback after cruciate-retaining and stabilizing total knee arthroplasty.

Authors:  Ephrat Most; Shay Zayontz; Guoan Li; Erik Otterberg; Kenneth Sabbag; Harry E Rubash
Journal:  Clin Orthop Relat Res       Date:  2003-05       Impact factor: 4.176

5.  A subject specific multibody model of the knee with menisci.

Authors:  Trent M Guess; Ganesh Thiagarajan; Mohammad Kia; Meenakshi Mishra
Journal:  Med Eng Phys       Date:  2010-03-31       Impact factor: 2.242

6.  Recruitment of knee joint ligaments.

Authors:  L Blankevoort; R Huiskes; A de Lange
Journal:  J Biomech Eng       Date:  1991-02       Impact factor: 2.097

7.  2003 Hap Paul Award Paper of the International Society for Technology in Arthroplasty. Design and activity dependence of kinematics in fixed and mobile-bearing knee arthroplasties.

Authors:  Scott A Banks; W Andrew Hodge
Journal:  J Arthroplasty       Date:  2004-10       Impact factor: 4.757

8.  In vivo fluoroscopic analysis of the normal human knee.

Authors:  Richard D Komistek; Douglas A Dennis; Mohamed Mahfouz
Journal:  Clin Orthop Relat Res       Date:  2003-05       Impact factor: 4.176

9.  Patient outcomes following tricompartmental total knee replacement. A meta-analysis.

Authors:  C M Callahan; B G Drake; D A Heck; R S Dittus
Journal:  JAMA       Date:  1994-05-04       Impact factor: 56.272

10.  Pain in the assessment of total knee replacement.

Authors:  D W Murray; S J Frost
Journal:  J Bone Joint Surg Br       Date:  1998-05
View more
  1 in total

1.  Preoperative Measurement of Tibial Resection in Total Knee Arthroplasty Improves Accuracy of Postoperative Limb Alignment Restoration.

Authors:  Pei-Hui Wu; Zhi-Qi Zhang; Shu-Ying Fang; Zi-Bo Yang; Yan Kang; Ming Fu; Wei-Ming Liao
Journal:  Chin Med J (Engl)       Date:  2016-11-05       Impact factor: 2.628

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.