Literature DB >> 25590188

Wing-shaped plastic stents vs. self-expandable metal stents for palliative drainage of malignant distal biliary obstruction: a randomized multicenter study.

Arthur Schmidt1, Bettina Riecken1, Susanne Rische1, Christoph Klinger1, Ralf Jakobs2, Matthias Bechtler2, Georg Kähler3, Arno Dormann4, Karel Caca1.   

Abstract

BACKGROUND AND STUDY AIMS: Previous studies have shown superior patency rates for self-expandable metal stents (SEMS) compared with plastic stents in patients with malignant biliary obstruction. The aim of this study was to compare stent patency, patient survival, and complication rates between a newly designed, wing-shaped, plastic stent and SEMSs in patients with unresectable, malignant, distal, biliary obstruction. PATIENTS AND METHODS: A randomized, multicenter trial was conducted at four tertiary care centers in Germany. A total of 37 patients underwent randomization between March 2010 and January 2013. Patients underwent endoscopic retrograde cholangiography with insertion of either a wing-shaped, plastic stent without lumen or an SEMS.
RESULTS: Stent failure occurred in 10/16 patients (62.5 %) in the winged-stent group vs. 4/18 patients (22.2 %) in the SEMS group (P = 0.034). The median time to stent failure was 51 days (range 2 - 92 days) for the winged stent and 80 days (range 28 - 266 days) for the SEMS (P = 0.002). Early stent failure (< 8 weeks after placement) occurred in 8 patients (50 %) vs. 2 patients (11.1 %), respectively (P = 0.022). After obtaining the results from this interim analysis, the study was discontinued because of safety concerns.
CONCLUSIONS: The frequency of stent failure was significantly higher in the winged-stent group compared with the SEMS group. A high incidence of early stent failure within 8 weeks was observed in the winged-stent group. Thus, the winged, plastic stent without central lumen may not be appropriate for mid or long term drainage of malignant biliary obstruction. Study registration ClinicalTrials.gov (NCT01063634). © Georg Thieme Verlag KG Stuttgart · New York.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25590188     DOI: 10.1055/s-0034-1391232

Source DB:  PubMed          Journal:  Endoscopy        ISSN: 0013-726X            Impact factor:   10.093


  6 in total

Review 1.  Plastic vs. Self-Expandable Metal Stents for Palliation in Malignant Biliary Obstruction: A Series of Meta-Analyses.

Authors:  Majid A Almadi; Alan Barkun; Myriam Martel
Journal:  Am J Gastroenterol       Date:  2016-11-15       Impact factor: 10.864

Review 2.  Recent Advances in Gastrointestinal Stent Development.

Authors:  Jin-Seok Park; Seok Jeong; Don Haeng Lee
Journal:  Clin Endosc       Date:  2015-05-29

3.  A randomized trial comparing winged versus conventional plastic stents for malignant bile duct strictures.

Authors:  Lachlan Ayres; Danny Cheriyan; Ryan Scott; Edward Kim; Terry Lee; Joel Singer; Fergal Donnellan; Michael Byrne; Alan Weiss; Robert Enns; Jack Amar; Eric Lam; Jennifer Telford
Journal:  Endosc Int Open       Date:  2017-07-06

4.  A multi-institution consensus on how to perform EUS-guided biliary drainage for malignant biliary obstruction.

Authors:  Jintao Guo; Marc Giovannini; Anand V Sahai; Adrian Saftoiu; Christoph F Dietrich; Erwin Santo; Pietro Fusaroli; Ali A Siddiqui; Manoop S Bhutani; Anthony Yuen Bun Teoh; Atsushi Irisawa; Brenda Lucia Arturo Arias; Chalapathi Rao Achanta; Christian Jenssen; Dong-Wan Seo; Douglas G Adler; Evangelos Kalaitzakis; Everson Artifon; Fumihide Itokawa; Jan Werner Poley; Girish Mishra; Khek Yu Ho; Hsiu-Po Wang; Hussein Hassan Okasha; Jesse Lachter; Juan J Vila; Julio Iglesias-Garcia; Kenji Yamao; Kenjiro Yasuda; Kensuke Kubota; Laurent Palazzo; Luis Carlos Sabbagh; Malay Sharma; Mitsuhiro Kida; Mohamed El-Nady; Nam Q Nguyen; Peter Vilmann; Pramod Kumar Garg; Praveer Rai; Shuntaro Mukai; Silvia Carrara; Sreeram Parupudi; Subbaramiah Sridhar; Sundeep Lakhtakia; Surinder S Rana; Takeshi Ogura; Todd H Baron; Vinay Dhir; Siyu Sun
Journal:  Endosc Ultrasound       Date:  2018 Nov-Dec       Impact factor: 5.628

5.  Survival, stent patency, and cost-effectiveness of plastic biliary stent versus metal biliary stent for palliation in malignant biliary obstruction in a developing country tertiary hospital.

Authors:  Muthia Farani; Siti R F Saldi; Hasan Maulahela; Murdani Abdullah; Ari F Syam; Dadang Makmum
Journal:  JGH Open       Date:  2021-07-20

6.  Early malfunction of a biliary self-expandable metal stent with an antireflux valve: A case report.

Authors:  Sang Hoon Kim; Chi Hyuk Oh; Jae Min Lee; Seong Ji Choi; Hyuk Soon Choi; Eun Sun Kim; Bora Keum; Yoon Tae Jeen; Hoon Jai Chun; Hong Sik Lee; Chang Duck Kim
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

  6 in total