Pre-clinical evaluation of an adult extracorporeal carbon dioxide removal system with active mixing for pediatric respiratory support.

R Garrett Jeffries1, Yerbol Mussin, Denis S Bulanin, Laura W Lund, Ergin Kocyildirim, Zhaksybay Zh Zhumadilov, Farkhad S Olzhayev, William J Federspiel, Peter D Wearden.   

Abstract

The objective of this work was to conduct pre-clinical feasibility studies to determine if a highly efficient, active-mixing, adult extracorporeal carbon dioxide removal (ECCO2R) system can safely be translated to the pediatric population. The Hemolung Respiratory Assist System (RAS) was tested in vitro and in vivo to evaluate its performance for pediatric veno-venous applications. The Hemolung RAS operates at blood flows of 350-550 ml/min and utilizes an integrated pump-gas exchange cartridge with a membrane surface area of 0.59 m² as the only component of the extracorporeal circuit. Both acute and seven-day chronic in vivo tests were conducted in healthy juvenile sheep using a veno-venous cannulation strategy adapted to the in vivo model. The Hemolung RAS was found to have gas exchange and pumping capabilities relevant to patients weighing 3-25 kg. Seven-day animal studies in juvenile sheep demonstrated that veno-venous extracorporeal support could be used safely and effectively with no significant adverse reactions related to device operation.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25588763      PMCID: PMC5410869          DOI: 10.5301/ijao.5000372

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  14 in total

1.  The Aachen MiniHLM--a miniaturized heart-lung machine for neonates with an integrated rotary blood pump.

Authors:  Jutta Arens; Heike Schnoering; Michael Pfennig; Ilona Mager; Jaime F Vázquez-Jiménez; Thomas Schmitz-Rode; Ulrich Steinseifer
Journal:  Artif Organs       Date:  2010-09       Impact factor: 3.094

Review 2.  ECMO for neonatal respiratory failure.

Authors:  K Rais Bahrami; Krisa P Van Meurs
Journal:  Semin Perinatol       Date:  2005-02       Impact factor: 3.300

3.  In vitro performance testing of a pediatric oxygenator with an integrated pulsatile pump.

Authors:  Ralf Borchardt; Peter Schlanstein; Ilona Mager; Jutta Arens; Thomas Schmitz-Rode; Ulrich Steinseifer
Journal:  ASAIO J       Date:  2012 Jul-Aug       Impact factor: 2.872

4.  Red blood cell damage by shear stress.

Authors:  L B Leverett; J D Hellums; C P Alfrey; E C Lynch
Journal:  Biophys J       Date:  1972-03       Impact factor: 4.033

5.  Extracorporeal membrane oxygenation and neonatal respiratory failure: experience from the extracorporeal life support organization.

Authors:  C J Stolar; S M Snedecor; R H Bartlett
Journal:  J Pediatr Surg       Date:  1991-05       Impact factor: 2.545

6.  A simple empirical formula for calculating approximate surface area in children.

Authors:  H Costeff
Journal:  Arch Dis Child       Date:  1966-12       Impact factor: 3.791

7.  Functional and biocompatibility performances of an integrated Maglev pump-oxygenator.

Authors:  Tao Zhang; Guangming Cheng; Andrew Koert; Juntao Zhang; Barry Gellman; G Kwame Yankey; Aditee Satpute; Kurt A Dasse; Richard J Gilbert; Bartley P Griffith; Zhongjun J Wu
Journal:  Artif Organs       Date:  2009-01       Impact factor: 3.094

8.  Respiratory dialysis with an active-mixing extracorporeal carbon dioxide removal system in a chronic sheep study.

Authors:  Peter D Wearden; William J Federspiel; Scott W Morley; Meir Rosenberg; Paul D Bieniek; Laura W Lund; Burt D Ochs
Journal:  Intensive Care Med       Date:  2012-08-25       Impact factor: 17.440

9.  Surface area in children: a simple formula.

Authors:  Muhammad Furqan; Anwarul Haque
Journal:  Indian Pediatr       Date:  2009-04-01       Impact factor: 1.411

10.  Computational Fluid Dynamics and Experimental Characterization of the Pediatric Pump-Lung.

Authors:  Zhongjun J Wu; Barry Gellman; Tao Zhang; M Ertan Taskin; Kurt A Dasse; Bartley P Griffith
Journal:  Cardiovasc Eng Technol       Date:  2011-12-01       Impact factor: 2.495

View more
  4 in total

1.  Dual Carbon Dioxide Capture to Achieve Highly Efficient Ultra-Low Blood Flow Extracorporeal Carbon Dioxide Removal.

Authors:  Brian Y Chang; Steven P Keller
Journal:  Ann Biomed Eng       Date:  2020-02-18       Impact factor: 3.934

2.  An extracorporeal carbon dioxide removal (ECCO2R) device operating at hemodialysis blood flow rates.

Authors:  R Garrett Jeffries; Laura Lund; Brian Frankowski; William J Federspiel
Journal:  Intensive Care Med Exp       Date:  2017-09-06

Review 3.  Paracorporeal Lung Devices: Thinking Outside the Box.

Authors:  Timothy M Maul; Jennifer S Nelson; Peter D Wearden
Journal:  Front Pediatr       Date:  2018-09-05       Impact factor: 3.418

4.  Alkaline Liquid Ventilation of the Membrane Lung for Extracorporeal Carbon Dioxide Removal (ECCO2R): In Vitro Study.

Authors:  Luigi Vivona; Michele Battistin; Eleonora Carlesso; Thomas Langer; Carlo Valsecchi; Sebastiano Maria Colombo; Serena Todaro; Stefano Gatti; Gaetano Florio; Antonio Pesenti; Giacomo Grasselli; Alberto Zanella
Journal:  Membranes (Basel)       Date:  2021-06-22
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.