Roos and NACP-02 ion chamber perturbations and water-air stopping-power ratios for clinical electron beams for energies from 4 to 22 MeV.

M Bailey1, D R Shipley, J W Manning.   

Abstract

Empirical fits are developed for depth-compensated wall- and cavity-replacement perturbations in the PTW Roos 34001 and IBA / Scanditronix NACP-02 parallel-plate ionisation chambers, for electron beam qualities from 4 to 22 MeV for depths up to approximately 1.1 × R₅₀,D. These are based on calculations using the Monte Carlo radiation transport code EGSnrc and its user codes with a full simulation of the linac treatment head modelled using BEAMnrc. These fits are used with calculated restricted stopping-power ratios between air and water to match measured depth-dose distributions in water from an Elekta Synergy clinical linear accelerator at the UK National Physical Laboratory. Results compare well with those from recent publications and from the IPEM 2003 electron beam radiotherapy Code of Practice.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25586026     DOI: 10.1088/0031-9155/60/3/1087

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  Beam Characterization of 10-MV Photon Beam from Medical Linear Accelerator without Flattening Filter.

Authors:  Tomohiro Shimozato; Yuichi Aoyama; Takuma Matsunaga; Katsuyoshi Tabushi
Journal:  J Med Phys       Date:  2017 Apr-Jun

2.  Dosimetric Evaluation of the QFix kVueTM Calypso Couch Top.

Authors:  Lingtong Hou; Huiqin Zhang; Xiaomei Sun; Qianqian Liu; Tingfeng Chen; Yong Liu; Xiaodong Jiang; Shengyu Yao
Journal:  Technol Cancer Res Treat       Date:  2021 Jan-Dec
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.