Are tubes containing sodium fluoride still needed for the measurement of blood glucose in hospital laboratory practice?

Amira Al-Kharusi1, Nuha Al-Lawati1, Manal Al-Kindi1, Waad-Allah Mula-Abed1.   

Abstract

OBJECTIVES: To compare glucose values obtained using sodium flouride (NaF) tubes and serum separation tubes (SST) in a tertiary care hospital laboratory setting.
METHODS: This study was conducted at the Clinical Biochemistry Laboratory, Royal Hospital, Oman. During the study period (1 September-30 November 2013), 50 pairs (one NaF tube and one SST) of patient's blood specimens were randomly collected. Following separation of plasma (NaF tubes) and serum (SST), glucose concentrations were measured by hexokinase assay using the Architect c8000. Fifteen pairs of these tubes were kept in the refrigerator at 4°C and plasma/serum glucose concentrations were measured daily up to seven days after collection.
RESULTS: Comparing plasma (NaF) and serum (SST) results of glucose values (n=50) showed an average difference of 0.00mmol/L (range -0.60 to +0.60mmol/L). Bland Altman analysis gave a non-significant constant bias of 0.10 ±0.195mmol/L (bias ±SD). Pearson correlation between plasma (NaF) and serum (SST) glucose concentrations revealed a significant correlation approaching unity with r(2)= 0.9991. No significant differences in glucose values were noted for both plasma and serum in 15 pairs of NaF and SST tubes when analyzed seven days following refrigeration. Hemolysis was observed in five (10%) NaF tubes compared with two (4%) SST.
CONCLUSION: There is no difference in glucose values collected from plasma NaF tubes or serum SST, and so SST can be used in hospital laboratory settings as there are practical advantages, including cost-effectiveness and reduction in blood volume drawn when utilizing these tubes for glucose and other tests from a single blood collection tube.

Entities:  

Keywords:  Blood glucose; Diabetes mellitus; Serum separator tubes (SST); Sodium fluoride

Year:  2014        PMID: 25584156      PMCID: PMC4289486          DOI: 10.5001/omj.2014.109

Source DB:  PubMed          Journal:  Oman Med J        ISSN: 1999-768X


  17 in total

1.  Glucose: a simple molecule that is not simple to quantify.

Authors:  Raymond Gambino
Journal:  Clin Chem       Date:  2007-12       Impact factor: 8.327

2.  A comparison of glucose concentration in paired specimens collected in serum separator and fluoride/potassium oxalate blood collection tubes under survey 'field' conditions.

Authors:  Lois Fernandez; Penny Jee; Mari-Jill Klein; Peter Fischer; Sherry L Perkins; Stephen P J Brooks
Journal:  Clin Biochem       Date:  2012-12-05       Impact factor: 3.281

3.  Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus.

Authors:  David B Sacks; Mark Arnold; George L Bakris; David E Bruns; Andrea Rita Horvath; M Sue Kirkman; Ake Lernmark; Boyd E Metzger; David M Nathan
Journal:  Clin Chem       Date:  2011-05-26       Impact factor: 8.327

4.  Effectiveness of sodium fluoride as a preservative of glucose in blood.

Authors:  A Y Chan; R Swaminathan; C S Cockram
Journal:  Clin Chem       Date:  1989-02       Impact factor: 8.327

5.  Glycolysis inhibitors negatively bias blood glucose measurements: potential impact on the reported prevalence of diabetes mellitus.

Authors:  W S Waring; L E Evans; C T Kirkpatrick
Journal:  J Clin Pathol       Date:  2007-07       Impact factor: 3.411

6.  Rapid blood separation is superior to fluoride for preventing in vitro reductions in measured blood glucose concentration.

Authors:  R Z Shi; E S Seeley; R Bowen; J D Faix
Journal:  J Clin Pathol       Date:  2009-08       Impact factor: 3.411

7.  Sodium fluoride: an ineffective inhibitor of glycolysis.

Authors:  Raymond Gambino
Journal:  Ann Clin Biochem       Date:  2012-11-05       Impact factor: 2.057

8.  Familial Clustering of Type 2 Diabetes among Omanis.

Authors:  Sawsan Al-Sinani; Mohammed Al-Shafaee; Ali Al-Mamari; Nicholas Woodhouse; Omaima Al-Shafie; Mohammed Hassan; Said Al-Yahyaee; Sulayma Albarwani; Deepali Jaju; Khamis Al-Hashmi; Mohammed Al-Abri; Syed Rizvi; Riad Bayoumi
Journal:  Oman Med J       Date:  2014-01

Review 9.  National, regional, and global trends in fasting plasma glucose and diabetes prevalence since 1980: systematic analysis of health examination surveys and epidemiological studies with 370 country-years and 2·7 million participants.

Authors:  Goodarz Danaei; Mariel M Finucane; Yuan Lu; Gitanjali M Singh; Melanie J Cowan; Christopher J Paciorek; John K Lin; Farshad Farzadfar; Young-Ho Khang; Gretchen A Stevens; Mayuree Rao; Mohammed K Ali; Leanne M Riley; Carolyn A Robinson; Majid Ezzati
Journal:  Lancet       Date:  2011-06-24       Impact factor: 79.321

10.  Impact of blood sample collection and processing methods on glucose levels in community outreach studies.

Authors:  Michael Turchiano; Cuong Nguyen; Arthur Fierman; Mark Lifshitz; Antonio Convit
Journal:  J Environ Public Health       Date:  2013-01-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.