Standardized perfusion value of the esophageal carcinoma and its correlation with quantitative CT perfusion parameter values.

A Djuric-Stefanovic1, Dj Saranovic2, D Sobic-Saranovic3, D Masulovic4, V Artiko5.   

Abstract

PURPOSE: Standardized perfusion value (SPV) is a universal indicator of tissue perfusion, normalized to the whole-body perfusion, which was proposed to simplify, unify and allow the interchangeability among the perfusion measurements and comparison between the tumor perfusion and metabolism. The aims of our study were to assess the standardized perfusion value (SPV) of the esophageal carcinoma, and its correlation with quantitative CT perfusion measurements: blood flow (BF), blood volume (BV), mean transit time (MTT) and permeability surface area product (PS) of the same tumor volume samples, which were obtained by deconvolution-based CT perfusion analysis.
METHODS: Forty CT perfusion studies of the esophageal cancer were analyzed, using the commercial deconvolution-based CT perfusion software (Perfusion 3.0, GE Healthcare). The SPV of the esophageal tumor and neighboring skeletal muscle were correlated with the corresponding mean tumor and muscle quantitative CT perfusion parameter values, using Spearman's rank correlation coefficient (rS).
RESULTS: Median SPV of the esophageal carcinoma (7.1; range: 2.8-13.4) significantly differed from the SPV of the skeletal muscle (median: 1.0; range: 0.4-2.4), (Z=-5.511, p<0.001). The cut-off value of the SPV of 2.5 enabled discrimination of esophageal cancer from the skeletal muscle with sensitivity and specificity of 100%. SPV of the esophageal carcinoma significantly correlated with corresponding tumor BF (rS=0.484, p=0.002), BV (rS=0.637, p<0.001) and PS (rS=0.432, p=0.005), and SPV of the skeletal muscle significantly correlated with corresponding muscle BF (rS=0.573, p<0.001), BV (rS=0.849, p<0.001) and PS (rS=0.761, p<0.001).
CONCLUSIONS: We presented a database of the SPV for the esophageal cancer and proved that SPV of the esophageal neoplasm significantly differs from the SPV of the skeletal muscle, which represented a sample of healthy tissue. The SPV was validated against quantitative CT perfusion measurements and statistically significant correlation was proved.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.

Entities:  

Keywords:  CT perfusion; Deconvolution; Esophageal carcinoma; Standardized perfusion value

Mesh:

Year:  2014        PMID: 25579474     DOI: 10.1016/j.ejrad.2014.12.004

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  1 in total

1.  Use of patient outcome endpoints to identify the best functional CT imaging parameters in metastatic renal cell carcinoma patients.

Authors:  Jill Rachel Mains; Frede Donskov; Erik Morre Pedersen; Hans Henrik Torp Madsen; Jesper Thygesen; Kennet Thorup; Finn Rasmussen
Journal:  Br J Radiol       Date:  2018-01-02       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.