Currents induced by fast movements inside the MRI room may cause inhibition in an implanted pacemaker.

Eugenio Mattei, Federica Censi, Matteo Mancini, Antonio Napolitano, Elisabetta Genovese, Vittorio Cannatà, Giancarlo Burriesci, Rosaria Falsaperla, Giovanni Calcagnini.   

Abstract

The static magnetic field generated by MRI systems is highly non-homogenous and rapidly decreases when moving away from the bore of the scanner. Consequently, the movement around the MRI scanner is equivalent to an exposure to a time-varying magnetic field at very low frequency (few Hz). If people with an implanted pacemaker (PM) enter the MRI room, fast movements may thus induce voltages on the loop formed by the PM lead, with the potential to modify the correct behavior of the stimulator. In this study, we performed in-vitro measurements on a human-shaped phantom, equipped with an implantable PM and with a current sensor, able to monitor the activity of the PM while moving the phantom in the MRI room. Fast rotational movements in close proximity of the bore of the scanner caused the inappropriate inhibition of the PM, programmed in VVI modality, maximum sensitivity, unipolar sensing and pacing. The inhibition occurred for a variation of the magnetic field of about 3 T/s. These findings demonstrate that great care must be paid when extending PM MRI compatibility from patients to healthcare personnel, since the safety procedures and the MRI-conditional PM programming (e.g. asynchronous stimulation or bipolar sensing) used for patients cannot be applied.

Entities:  

Mesh:

Year:  2014        PMID: 25570102     DOI: 10.1109/EMBC.2014.6943734

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X


  1 in total

1.  Assessment of Electromagnetic Interference with Active Cardiovascular Implantable Electronic Devices (CIEDs) Caused by the Qi A13 Design Wireless Charging Board.

Authors:  Tobias Seckler; Kai Jagielski; Dominik Stunder
Journal:  Int J Environ Res Public Health       Date:  2015-05-27       Impact factor: 3.390

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.