Wireless simultaneous stimulation-and-recording device to train cortical circuits in somatosensory cortex.

John T Ramshur, Amy L de Jongh Curry, Robert S Waters.   

Abstract

We describe for the first time the design, implementation, and testing of a telemetry controlled simultaneous stimulation and recording device (SRD) to deliver chronic intercortical microstimulation (ICMS) to physiologically identified sites in rat somatosensory cortex (SI) and test hypotheses that chronic ICMS strengthens interhemispheric pathways and leads to functional reorganization in the enhanced cortex. The SRD is a custom embedded device that uses the Cypress Semiconductor's programmable system on a chip (PSoC) that is remotely controlled via Bluetooth. The SRC can record single or multiunit responses from any two of 12 available inputs at 1-15 ksps per channel and simultaneously deliver stimulus pulses (0-255 μA; 10 V compliance) to two user selectable electrodes using monophasic, biphasic, or pseudophasic stimulation waveforms (duration: 0-5 ms, inter-phase interval: 0-5 ms, frequency: 0.1-5 s, delay: 0-10 ms). The SRD was bench tested and validated in vivo in a rat animal model.

Entities:  

Mesh:

Year:  2014        PMID: 25569987     DOI: 10.1109/EMBC.2014.6943619

Source DB:  PubMed          Journal:  Conf Proc IEEE Eng Med Biol Soc        ISSN: 1557-170X