| Literature DB >> 25569759 |
Li Wang1, Huiren Xu2, Yilin Song3, Jinping Luo4, Shengwei Xu5, Song Zhang6, Juntao Liu4, Xinxia Cai7.
Abstract
Vesicular exocytosis is ubiquitous, but it is difficult to detect within the cells' communication mechanism. For this purpose, a 2 µm ultramicrodic carbon fiber electrode was fabricated in this work based on electrodeposition with over-oxidized polypyrrole nanoparticle (PPyox-CFE), which was applied successfully for real-time monitoring of quantal exocytosis from individual pheochromocytoma (PC12) cells. PPyox-CFE was evaluated by dopamine (DA) solutions through cyclic voltammetry and amperometry electrochemical methods, and results revealed that PPyox-CFE improved the detection limit of DA. In particular, the sensitivity of DA was improved to 24.55 µA·µM(-1)·µm(-2) using the PPyox-CFE. The ultramicrodic electrode combined with the patch-clamp system was used to detect vesicular exocytosis of DA from individual PC12 cells with 60 mM K+ stimulation. A total of 287 spikes released from 7 PC12 cells were statistically analyzed. The current amplitude (Imax) and the released charge (Q) of the amperometric spikes from the DA release by a stimulated PC12 cell is 45.1 ± 12.5 pA and 0.18 ± 0.04 pC, respectively. Furthermore, on average ~562,000 molecules were released in each vesicular exocytosis. PPyox-CFE, with its capability of detecting vesicular exocytosis, has potential application in neuron communication research.Entities:
Mesh:
Substances:
Year: 2015 PMID: 25569759 PMCID: PMC4327054 DOI: 10.3390/s150100868
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.(A) The fabricated CFE with diameter of 2 μm; the inset shows interface between carbon fiber and the glass insulator; (B) The SEM at the cycle of 8.
Figure 2.Signal-to-noise (S/N) statistics of current response to 2 μM DA.
Figure 3.Cyclic voltammetry response of bare CFE was compared to those CFE modified with PPyox the electrochemical in 2 μM DA solution.
Figure 4.(A) Current-time response of the CFE electrodeposited with PPyox film at different concentrations at 149 mV vs. Ag|AgCl; (B) Detection limit of DA using PPyox-CFE; (C) Plot of current versus concentration of DA in the range of 0.5 μM to 12 μM.
Figure 5.(A) The traces are recorded by amperometric method continuously for 25 s at the distance of 1 μm (i), 500 μm (ii) from cell with 60 mM K+ stimulation solution, respectively; (B) The diagram of the single PC12 cell release monitoring with the PPyox-CFE; (C) The obtained minimum amplitude of the spike.
Figure 6.Stability of PPyox-CFE, expressed as percentage of sensitivity to DA in PBS (pH 7.4) and the detection limit of DA.