Study of the feasible size of a bone conduction implant transducer in the temporal bone.

Sabine Reinfeldt1, Per Östli, Bo Håkansson, Hamidreza Taghavi, Måns Eeg-Olofsson, Joacim Stalfors.   

Abstract

HYPOTHESIS: The aim was to assess the temporal bone volume to determine the suitable size and position of a bone conduction implant (BCI) transducer.
BACKGROUND: A BCI transducer needs to be sufficiently small to fit in the mastoid portion of the temporal bone for a majority of patients. The anatomical geometry limits both the dimension of an implanted transducer and its positions in the temporal bone to provide a safe and simple surgery.
METHODS: Computed tomography (CT) scans of temporal bones from 22 subjects were virtually reconstructed. With an algorithm in MATLAB, the maximum transducer diameter as function of the maximum transducer depth in the temporal bone, and the most suitable position were calculated in all subjects.
RESULTS: An implanted transducer diameter of 16 mm inserted at a depth of 4 mm statistically fitted 95% of the subjects. If changing the transducer diameter to 12 mm, a depth of 6 mm would fit in 95% of the subjects. The most suitable position was found to be around 20 mm behind the ear canal.
CONCLUSION: The present BCI transducer casing, used in ongoing clinical trials, was designed from the results in this study, demonstrating that the present BCI transducer casing (largest diameter [diagonal]: 15.5 mm, height: 6.4 mm) will statistically fit more than 95% of the subjects. Hence, the present BCI transducer is concluded to be sufficiently small to fit most normal-sized temporal bones and should be placed approximately 20 mm behind the ear canal.

Entities:  

Mesh:

Year:  2015        PMID: 25569368     DOI: 10.1097/MAO.0000000000000682

Source DB:  PubMed          Journal:  Otol Neurotol        ISSN: 1531-7129            Impact factor:   2.311


  3 in total

1.  The bone conduction implant: Clinical results of the first six patients.

Authors:  Sabine Reinfeldt; Bo Håkansson; Hamidreza Taghavi; Karl-Johan Fredén Jansson; Måns Eeg-Olofsson
Journal:  Int J Audiol       Date:  2015-02-23       Impact factor: 2.117

Review 2.  New developments in bone-conduction hearing implants: a review.

Authors:  Sabine Reinfeldt; Bo Håkansson; Hamidreza Taghavi; Måns Eeg-Olofsson
Journal:  Med Devices (Auckl)       Date:  2015-01-16

Review 3.  Implantation of a new active bone conduction hearing device with optimized geometry.

Authors:  S K Plontke; G Götze; C Wenzel; T Rahne; R Mlynski
Journal:  HNO       Date:  2020-08       Impact factor: 1.284

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.