Youcef Nedjadi1, Claude Bailat2, Yvan Caffari2, Philippe Cassette3, François Bochud2.
Abstract
A triple-to-double coincidence ratio (TDCR) counter was recently constructed at IRA-METAS for liquid scintillation based primary activity standardisations. A description of its optical chamber, efficiency change tools, photomultiplier tubes (PMTs) and electronics is given. This TDCR system was validated by measuring several standard solutions of beta emitters including (45)Ca, (14)C, (63)Ni and (3)H. The activity concentrations, obtained from these measurements and efficiencies computed with a FORTRAN code we developed for symmetric and asymmetric PMTs, agree with the certified values within uncertainties.Entities:
Keywords: (14)C; (3)H; (45)Ca; (63)Ni; Activity standardisation; Liquid scintillation; TDCR
Year: 2014 PMID: 25569200 DOI: 10.1016/j.apradiso.2014.12.023
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513