Literature DB >> 25567717

[Design and fabrication of the custom-made titanium condyle by selective laser melting technology].

Jianyu Chen1, Chongdai Luo, Chunyu Zhang, Gong Zhang, Weiqian Qiu, Zhiguang Zhang2.   

Abstract

OBJECTIVE: To design and fabricate the custom-made titanium mandibular condyle by the reverse engineering technology combined with selective laser melting (SLM) technology and to explore the mechanical properties of the SLM-processed samples and the application of the custom-made condyle in the temporomandibular joint (TMJ) reconstruction.
METHODS: The three-dimensional model of the mandibular condyle was obtained from a series of CT databases. The custom-made condyle model was designed by the reverse engineering software. The mandibular condyle was made of titanium powder with a particle size of 20-65 µm as the basic material and the processing was carried out in an argon atmosphere by the SLM machine. The yield strength, ultimate strength, bending strength, hardness, surface morphology and roughness were tested and analyzed. The finite element analysis (FEA) was used to analyze the stress distribution.
RESULTS: The complex geometry and the surface of the custom-made condyle can be reproduced precisely by the SLM. The mechanical results showed that the yield strength, ultimate strength, bending strength and hardness were (559±14) MPa, (659±32) MPa, (1 067±42) MPa, and (212±4)HV, respectively. The surface roughness was reduced by sandblast treatment.
CONCLUSIONS: The custom-made titanium condyle can be fabricated by SLM technology which is time-saving and highly digitized. The mechanical properties of the SLM sample can meet the requirements of surgical implant material in the clinic. The possibility of fabricating custom-made titanium mandibular condyle combined with the FEA opens new interesting perspectives for TMJ reconstruction.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25567717

Source DB:  PubMed          Journal:  Zhonghua Kou Qiang Yi Xue Za Zhi        ISSN: 1002-0098


  1 in total

Review 1.  A review on computer-aided design and manufacturing of patient-specific maxillofacial implants.

Authors:  Afaque Rafique Memon; Enpeng Wang; Junlei Hu; Jan Egger; Xiaojun Chen
Journal:  Expert Rev Med Devices       Date:  2020-03-12       Impact factor: 3.166