| Literature DB >> 25562677 |
Johanna L Peeples1, Matthew H Stokely2, Michael C Poorman2, Gerald T Bida3, Bruce W Wieland2.
Abstract
A high power [F-18] fluoride target package for the RDS-112 cyclotron has been designed, tested, and commercially deployed. The upgrade includes the CF-1000 target, a 1.3kW water target with an established commercial history on RDS-111/Eclipse cyclotrons, and a redesigned collimator with improved heat rejection capabilities. Conjugate heat transfer analyses were employed to both evaluate the existing collimator capabilities and design a suitable high current replacement.Entities:
Keywords: Batch boiling target; Collimator; Cyclotron targetry; Fluorine-18; Positron emission tomography; RDS-112
Mesh:
Substances:
Year: 2014 PMID: 25562677 PMCID: PMC4329277 DOI: 10.1016/j.apradiso.2014.12.028
Source DB: PubMed Journal: Appl Radiat Isot ISSN: 0969-8043 Impact factor: 1.513