| Literature DB >> 25559781 |
Sören Weyer1, Tobias Menden, Lennart Leicht, Steffen Leonhardt, Tobias Wartzek.
Abstract
Cardiovascular diseases as well as pulmonary oedema can be early diagnosed using vital signs and thoracic bio-impedance. By recording the electrocardiogram (ECG) and the impedance cardiogram (ICG), vital parameters are captured continuously. The aim of this study is the continuous monitoring of ECG and multi-frequency ICG by a mobile system. A mobile measuring system, based on 'low-power' ECG, ICG and an included radio transmission is described. Due to the high component integration, a board size of only 6.5 cm×5 cm could be realized. The measured data can be transmitted via Bluetooth and visualized on a portable monitor. By using energy-efficient hardware, the system can operate for up to 18 hs with a 3 V battery, continuously sending data via Bluetooth. Longer operating times can be realized by decreased transfer rates. The relative error of the impedance measurement was less than 1%. The ECG and ICG measurements allow an approximate calculation of the heart stroke volume. The ECG and the measured impedance showed a high correlation to commercial devices (r=0.83, p<0.05). In addition to commercial devices, the developed system allows a multi-frequency measurement of the thoracic impedance between 5-150 kHz.Entities:
Keywords: Impedance cardiography; multi-frequency; personal healthcare; portable
Mesh:
Year: 2015 PMID: 25559781 DOI: 10.3109/03091902.2014.990161
Source DB: PubMed Journal: J Med Eng Technol ISSN: 0309-1902