Traction force during vacuum extraction: a prospective observational study.

K Pettersson1, J Ajne1, K Yousaf1, D Sturm2, M Westgren1, G Ajne1.   

Abstract

OBJECTIVE: To investigate the traction force employed during vacuum extractions.
DESIGN: Observational cross-sectional study.
SETTING: Obstetric Department, Karolinska University Hospital, Sweden, and the Swedish National Congress of Obstetrics and Gynaecology, 2013. POPULATION: Two hundred women with vacuum extraction at term and 130 obstetricians participating in a simulated setting.
METHODS: In a normal clinical setting, we used a specially adapted device to measure and record the force used to undertake vacuum extraction. In a subsequent part of the study, the force employed for vacuum extraction by a group of obstetricians in a fictive setting was estimated and objectively measured. MAIN OUTCOME MEASURES: Applied force during vacuum extraction in relation to the estimated level of difficulty in the delivery; perinatal diagnoses of asphyxia or head trauma; estimated force compared with objectively measured force employed in the fictive setting.
RESULTS: The median (minimum-maximum) peak forces for minimum, average and excessive vacuum extraction in the clinical setting were 176 N (5-360 N), 225 N (115-436 N), and 241 N (164-452 N), respectively. In 34% of cases a force in excess of 216 N was employed. There was no correlation between the umbilical arterial pH at delivery and the traction force employed during extraction. Four cases of mild hypoxic ischaemic encephalopathy were observed, three of which were associated with a delivery whereby excessive traction force was employed during the vacuum extraction. In the fictive setting, the actual exerted force was twice the quantitative estimation. The measured forces in the clinical setting were four times higher than that estimated in the fictive setting.
CONCLUSIONS: Higher than expected levels of traction force were used for vacuum extraction delivery. As obstetricians tend to underestimate the force applied during vacuum extraction, objective measurement with instantaneous feedback may be valuable in raising awareness.
© 2015 Royal College of Obstetricians and Gynaecologists.

Entities:  

Keywords:  Failed vacuum extraction; traction force; vacuum extraction

Mesh:

Year:  2015        PMID: 25558833     DOI: 10.1111/1471-0528.13222

Source DB:  PubMed          Journal:  BJOG        ISSN: 1470-0328            Impact factor:   6.531


  7 in total

1.  Long-term effects of vacuum extraction on pelvic floor function: a cohort study in primipara.

Authors:  Ida Nilsson; Sigvard Åkervall; Ian Milsom; Maria Gyhagen
Journal:  Int Urogynecol J       Date:  2015-12-29       Impact factor: 2.894

2.  Neck Injuries Caused by Delivery Modes? Late Questions to the Matshes et al.

Authors:  Niels Lynøe; Anders Eriksson
Journal:  Acad Forensic Pathol       Date:  2021-11-25

3.  Predictive factors for the success of McRoberts' manoeuvre and suprapubic pressure in relieving shoulder dystocia: a cross-sectional study.

Authors:  Zara Lin Zau Lok; Yvonne Kwun Yue Cheng; Tak Yeung Leung
Journal:  BMC Pregnancy Childbirth       Date:  2016-10-29       Impact factor: 3.007

4.  Decreased neonatal pain response after vaginal-operative delivery with Kiwi OmniCup versus metal ventouse.

Authors:  E A Huhn; E Visca; D R Vogt; S von Felten; E M Tinner Oehler; C Bührer; D Surbek; R Zimmermann; I Hoesli
Journal:  BMC Pregnancy Childbirth       Date:  2017-01-31       Impact factor: 3.007

5.  Predictive value of traction force measurement in vacuum extraction: Development of a multivariate prognostic model.

Authors:  Kristina Pettersson; Khurram Yousaf; Jonas Ranstam; Magnus Westgren; Gunilla Ajne
Journal:  PLoS One       Date:  2017-03-03       Impact factor: 3.240

6.  Perinatal outcome after vacuum assisted delivery with digital feedback on traction force; a randomised controlled study.

Authors:  Stefhanie Romero; Kristina Pettersson; Khurram Yousaf; Magnus Westgren; Gunilla Ajne
Journal:  BMC Pregnancy Childbirth       Date:  2021-02-26       Impact factor: 3.007

7.  Maternal and Neonatal Outcomes of Operative Vaginal Deliveries at a Single Tertiary Center.

Authors:  Nihal Al Riyami; Manar Al Salmiyah; Durdana Khan; Intisar Al Riyami
Journal:  Oman Med J       Date:  2021-05-31
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.