Literature DB >> 25553546

Analysis of drugs of abuse by online SPE-LC high resolution mass spectrometry: communal assessment of consumption.

Nubia V Heuett1, Cesar E Ramirez2, Adolfo Fernandez3, Piero R Gardinali4.   

Abstract

An online SPE-LC-HRMS method was developed to monitor the consumption of 18 drugs of abuse (DOAs) including amphetamines, opioids, cocainics, cannabinoids, lysergics, and their corresponding metabolites in a well characterized college campus setting via wastewater analysis. Filtered and diluted (10×) sewage water samples (5 mL inj.) were automatically pre-concentrated and analyzed in 15 min using a Thermo EQuan MAX online SPE system equipped with a HyperSep™ Retain PEP (20×2.1 mm×12 μm) SPE column and a Hypersil Gold™ aQ (150×2.1 mm×3 μm) analytical column. A Q Exactive™ Hybrid Quadrupole-Orbitrap HRMS was used in full scan mode (R=140,000) for positive identification, and quantitation of target compounds. Method detection limits for all analytes ranged between 0.6 and 1.7 ng/L in sewage. A total of 14 DOAs were detected from two different locations (dorms and main college campus) within a one-year period. Most frequently detected drugs throughout the entire study were amphetamine (>96%) and THC's metabolite 11-nor-9-carboxy-Δ-9-THC (>100%) with maximum concentrations of 5956 and 2413 ng/L respectively. Daily doses per 1000 people were determined in order to assess consumption of THC, amphetamine, heroin and cocaine, in both dorms and main campus.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  College campus; High resolution mass spectrometry; Illicit drugs; Online-SPE; Sewage influents

Mesh:

Substances:

Year:  2014        PMID: 25553546     DOI: 10.1016/j.scitotenv.2014.12.043

Source DB:  PubMed          Journal:  Sci Total Environ        ISSN: 0048-9697            Impact factor:   7.963


  6 in total

Review 1.  Quantitative mass spectrometry methods for pharmaceutical analysis.

Authors:  Glenn Loos; Ann Van Schepdael; Deirdre Cabooter
Journal:  Philos Trans A Math Phys Eng Sci       Date:  2016-10-28       Impact factor: 4.226

2.  Automated method to determine pharmaceutical compounds in wastewater using on-line solid-phase extraction coupled to LC-MS/MS.

Authors:  César Augusto Marasco Júnior; Bianca Ferreira da Silva; Rafaela Silva Lamarca; Paulo Clairmont Feitosa de Lima Gomes
Journal:  Anal Bioanal Chem       Date:  2021-06-26       Impact factor: 4.142

3.  Long-term tracking of opioid consumption in two United States cities using wastewater-based epidemiology approach.

Authors:  Adam J Gushgari; Arjun K Venkatesan; Jing Chen; Joshua C Steele; Rolf U Halden
Journal:  Water Res       Date:  2019-06-03       Impact factor: 11.236

4.  Automatic analytical approach for the determination of 12 illicit drugs and nicotine metabolites in wastewater using on-line SPE-UHPLC-MS/MS.

Authors:  Jingyuan Wang; Likai Qi; Chenzhi Hou; Tingting Zhang; Mengyi Chen; Haitao Meng; Mengxiang Su; Hui Xu; Zhendong Hua; Youmei Wang; Bin Di
Journal:  J Pharm Anal       Date:  2021-01-22

Review 5.  Mass spectrometric strategies for the investigation of biomarkers of illicit drug use in wastewater.

Authors:  Félix Hernández; Sara Castiglioni; Adrian Covaci; Pim de Voogt; Erik Emke; Barbara Kasprzyk-Hordern; Christoph Ort; Malcolm Reid; Juan V Sancho; Kevin V Thomas; Alexander L N van Nuijs; Ettore Zuccato; Lubertus Bijlsma
Journal:  Mass Spectrom Rev       Date:  2016-10-17       Impact factor: 10.946

6.  Quantitative analysis of opioids and cannabinoids in wastewater samples.

Authors:  Alethea Jacox; Jillian Wetzel; Shu-Yuan Cheng; Marta Concheiro
Journal:  Forensic Sci Res       Date:  2017-01-30
  6 in total