| Literature DB >> 25553292 |
Eun Youn Roh1, Ji Won In2, Sue Shin1, Jong Hyun Yoon1, Kyoung Un Park2, Eun Young Song3.
Abstract
Intermediate-resolution HLA-DQ typing has gained importance in organ transplantation recently. We evaluated the performance of the LIFECODES HLA-DQB1 typing kit (Immucor, USA) using sequence-specific oligonucleotide (SSO) probe and Luminex platform (Luminex Corp., USA) on 100 samples tested by sequence-based typing (SBT) using the AlleleSEQR HLA-DQB1 kit (Abbott Molecular, USA) in Korean individuals. No sample showed ambiguity in the assignment of 4-digit HLA-DQB1 allele with the LIFECODES HLA-DQB1 SSO typing kit, and the results were fully concordant with those of high-resolution typing of AlleleSEQR HLA-DQB1 SBT up to 4-digit level. Three samples required adjustment of false reactions (3/100, 3.0%): two samples with DQB1(*)03:03/(*)06:01 showed false-positive result in probe 253, and 1 sample with DQB1(*)04:02/(*)05:02 showed false-negative result in probe 217. We tested an additional sample with DQB1(*)03:03/(*)06:01, which showed same false-positivity in probe 253 and 2 samples with DQB1(*)04:02/(*)05:02, which showed no false reaction. The false reactions did not result in ambiguity or change in the HLA allele assignment. We could assign HLA-DQB1 alleles to 4 digit-level without ambiguity, with 100% concordance with the SBT results. Thus, LIFECODES HLA-DQB1 SSO typing kit showed good performance for intermediate-resolution HLA-DQB1 typing in clinical laboratory for organ transplantation in Koreans.Entities:
Keywords: HLA-DQB1; Korean; Luminex; Organ transplantation; PCR-SSO
Mesh:
Substances:
Year: 2014 PMID: 25553292 PMCID: PMC4272942 DOI: 10.3343/alm.2015.35.1.123
Source DB: PubMed Journal: Ann Lab Med ISSN: 2234-3806 Impact factor: 3.464
Allele frequencies (%) of HLA-DQB1 in Koreans and other ethnic groups
DQB1 alleles with allele frequency >0.1% in Koreans or >0.5% in other ethnic groups are shown.
†DQB1*02:01 allele was not differentiated from DQB1*02:02.
HLA-DQB1 assignment by using the LIFECODE HLA-DQB1 SSO typing kit in 100 Koreans
The assigned SSO results were 100% concordant with the AlleleSEQR HLA-DQB1 sequence-based typing (SBT) results.
†High resolution HLA-DQB1 typing was performed by using the AlleleSEQR HLA-DQB1 SBT kit. HLA-DQB1 allelic combinations with false-positive or false-negative probe by using the LIFECODE HLA-DQB1 SSO typing kit are shown in bold font.
Abbreviations: SSO, sequence-specific oligonucleotide; N, number; FP, false positive; FN, false negative.