How to secure the connection between thoracostomy tube and drainage system?

Ka Ki Pat Li1, Kit Shing John Wong1, Yau Hang Henry Wong2, Ka Lok Cheng3, Fung Ling So1, Chu Leung Lau1, Chak Wah Kam1.   

Abstract

BACKGROUND: Thoracostomy tube insertion is one of the common bedside procedures in emergency medicine and many acute specialties. Dislodgement of thoracostomy tube from the connection tube of chest drainage system is an important problem with potential complications such as contamination, infection and pneumothorax. Besides, mere loosening can also lead to malfunction. It is a common practice to tape the connection of the system. This study aimed to evaluate the materials and methods of connection of chest drain system to minimize drainage dislodgement.
METHODS: We conducted an experimental study to assess the tightness of the connection with various taping materials and methods. We selected three commonly used adhesive materials (3M™ Transpore™ Medical tape, 3M™ Micropore™ Medical tape, 3M™ Soft Cloth Tape on Liner) and three different methods (cross method, straight method, nylon band) to secure the junction between the thoracostomy tube and the bi-conical adaptor in the drainage system. The measured outcome was the weight causing visible loosening of the junction between thoracotomy tube and the adaptor.
RESULTS: For each taping material and taping method, 10 trials were performed. The median weight required to disconnect the junction is 26.22 lb for Transpore™, 31.29 lb for Micropore™ and 32.44 lb for Soft Cloth Tape on Liner. A smaller force was required to disconnect if Transpore™ is used (P<0.001). There was no statistical significant difference between Micropore™ and Soft Cloth Tape on Liner (P=0.98). The median disconnecting force is 32.44 lb for straight taping method, 40.55 lb for cross taping method and 21.15 lb for plastic band. The cross-taping method was the more secure method (P<0.0001 when compared with plastic band) (P=0.033 when compared with straight method).
CONCLUSION: Cross-taping is the most secure method among the tested varieties in connecting the thoracostomy tube to the chest drainage system. Transpore™ is not a recommended material for thoracostomy tube taping.

Entities:  

Keywords:  Cross-taping; Drainage system; Thoracostomy tube

Year:  2014        PMID: 25548598      PMCID: PMC4272928          DOI: 10.5847/wjem.j.issn.1920-8642.2014.04.003

Source DB:  PubMed          Journal:  World J Emerg Med        ISSN: 1920-8642


  2 in total

1.  BTS guidelines for the insertion of a chest drain.

Authors:  D Laws; E Neville; J Duffy
Journal:  Thorax       Date:  2003-05       Impact factor: 9.139

2.  Tube thoracostomy: complications and its management.

Authors:  Emeka B Kesieme; Andrew Dongo; Ndubueze Ezemba; Eshiobo Irekpita; Nze Jebbin; Chinenye Kesieme
Journal:  Pulm Med       Date:  2011-10-16