| Literature DB >> 25545895 |
Sándor Beniczky1, Isa Conradsen, Mihai Moldovan, Poul Jennum, Martin Fabricius, Krisztina Benedek, Noémi Andersen, Helle Hjalgrim, Peter Wolf.
Abstract
Our objective was the clinical validation of an automated algorithm based on surface electromyography (EMG) for differentiation between convulsive epileptic and psychogenic nonepileptic seizures (PNESs). Forty-four consecutive episodes with convulsive events were automatically analyzed with the algorithm: 25 generalized tonic-clonic seizures (GTCSs) from 11 patients, and 19 episodes of convulsive PNES from 13 patients. The gold standard was the interpretation of the video-electroencephalographic recordings by experts blinded to the EMG results. The algorithm correctly classified 24 GTCSs (96%) and 18 PNESs (95%). The overall diagnostic accuracy was 95%. This algorithm is useful for distinguishing between epileptic and psychogenic convulsive seizures.Entities:
Mesh:
Year: 2015 PMID: 25545895 DOI: 10.1002/ana.24338