Literature DB >> 25545684

Cardiovascular perforation during placement of an Avalon Elite® Bicaval dual lumen ECMO cannula in a newborn.

Matias E Czerwonko1, Maria V Fraga, David J Goldberg, Holly L Hedrick, Pablo Laje.   

Abstract

Proper functioning of the Avalon Elite® bicaval dual lumen ECMO cannula (Maquet Cardiovascular, Wayne, NJ, USA) requires precise placement of the distal draining port within the lumen of the inferior vena cava (IVC). In order to advance the cannula to the correct position, a 0.038" guidewire is placed into the IVC under echocardiographic or fluoroscopic guidance. We report a case of perforation of the intrapericardiac section of the IVC by the guidewire in a neonate, not detected at the time of placement, resulting in cardiac tamponade and death. We recommend routine echocardiographic surveillance after placement of Avalon Elite® cannulas to rule out hemopericardium.
© 2014 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 25545684     DOI: 10.1111/jocs.12507

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  2 in total

Review 1.  The Role of Echocardiography in Neonates and Pediatric Patients on Extracorporeal Membrane Oxygenation.

Authors:  Carles Bautista-Rodriguez; Joan Sanchez-de-Toledo; Eduardo M Da Cruz
Journal:  Front Pediatr       Date:  2018-10-26       Impact factor: 3.418

2.  Right Atrial Perforation Leading to Cardiac Tamponade Following Veno-Venous Extracorporeal Membrane Oxygenation Cannulation.

Authors:  Jasmine Sandhu; Ryan K Dean; David Landsberg
Journal:  Cureus       Date:  2021-02-05
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.