Literature DB >> 25545655

Feasibility of closed-loop titration of propofol and remifentanil guided by the bispectral monitor in pediatric and adolescent patients: a prospective randomized study.

Gilles A Orliaguet1, Fatema Benabbes Lambert, Thierry Chazot, Pauline Glasman, Marc Fischler, Ngai Liu.   

Abstract

BACKGROUND: This study was designed to assess the feasibility of dual closed-loop titration of propofol and remifentanil guided solely by the Bispectral Index (BIS) monitor in pediatric and adolescent patients during anesthesia.
METHODS: Children undergoing elective surgery in this single-blind randomized study were allocated into the closed-loop (auto) or manual (manual) group. Primary outcome was the percentage of time with the BIS in the range 40 to 60 (BIS40-60). Secondary outcomes were the percentage of deep (BIS<40) anesthesia and drug consumption. Data are presented as median (interquartile range) or number (%).
RESULTS: Twenty-three patients (12 [10 to 14] yr) were assigned to the auto group and 19 (14 [7 to 14] yr) to the manual group. The closed-loop controller was able to provide induction and maintenance for all patients. The percentage of time with BIS40-60 was greater in the auto group (87% [75 to 96] vs. 72% [48 to 79]; P = 0.002), with a decrease in the percentage of BIS<40 (7% [2 to 17] vs. 21% [11 to 38]; P = 0.002). Propofol (2.4 [1.9 to 3.3] vs. 1.7 [1.2 to 2.8] mg/kg) and remifentanil (2.3 [2.0 to 3.0] vs. 2.5 [1.2 to 4.3] μg/kg) consumptions were similar in auto versus manual groups during induction, respectively. During maintenance, propofol consumption (8.2 [6.0 to 10.2] vs. 7.9 [7.2 to 9.1] mg kg h; P = 0.89) was similar between the two groups, but remifentanil consumption was greater in the auto group (0.39 [0.22 to 0.60] vs. 0.22 [0.17 to 0.32] μg kg min; P = 0.003). Perioperative adverse events and length of stay in the postanesthesia care unit were similar.
CONCLUSION: Intraoperative automated control of hypnosis and analgesia guided by the BIS is clinically feasible in pediatric and adolescent patients and outperformed skilled manual control.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25545655     DOI: 10.1097/ALN.0000000000000577

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  2 in total

1.  Effect of magnesium sulphate on bi-spectral index (BIS) values during general anesthesia in children.

Authors:  Mahmoud Mostafa Amer; Ahmed Abdelaal Ahmed Mahmoud; Marwa Khaled Abdelrahman Mohammed; Ahmed Mostafa Elsharawy; Doaa Abo-Elkasem Ahmed; Ehab Mohamed Farag
Journal:  BMC Anesthesiol       Date:  2015-09-22       Impact factor: 2.217

2.  Hybrid Intelligent System to Perform Fault Detection on BIS Sensor During Surgeries.

Authors:  José-Luis Casteleiro-Roca; José Luis Calvo-Rolle; Juan Albino Méndez Pérez; Nieves Roqueñí Gutiérrez; Francisco Javier de Cos Juez
Journal:  Sensors (Basel)       Date:  2017-01-18       Impact factor: 3.576

  2 in total