Comparison of results of the manual and automated scoring of micronucleus frequencies in (60)Co-irradiated peripheral blood lymphocytes for triage dosimetry.

G Tamizh Selvan1, N K Chaudhury2, P Venkatachalam3.   

Abstract

Scoring micronuclei in the peripheral blood lymphocytes of individuals exposed to ionizing radiation is a rapid biodosimetry assay. Peripheral blood lymphocytes from five individuals were exposed in vitro to 0-5Gy of (60)Co γ-radiation at a dose rate of 0.76Gy/min. The blood cultures were initiated with RPMI-1640 (80%) supplemented with FBS (20%), stimulated with mitogen and incubated at 37°C for 44h. At the 44th hour, cytochalasin-B (6µg/mL) was added, and the cultures were incubated for 28h more. The cells were harvested with a pre-chilled hypotonic solution (0.075M) and fixed with a Carnoy's solution (methanol/acetic acid 5:1). Giemsa- and propidium-iodide-stained cells affixed to slides for microscopy were scored manually and automatically with the micronucleus scoring software from MetaSystems. The micronucleus frequencies determined in the Giemsa-stained cells by manual and automated scoring were 23.6% different (P<0.0001) with an efficiency of 24.9%. Slides stained with propidium iodide are a better choice for automated scoring than Giemsa-stained ones.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Automation; Biodosimetry; Blood; Micronuclei

Year:  2014        PMID: 25544665     DOI: 10.1016/j.apradiso.2014.12.018

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  Liquid Handling Optimization in High-Throughput Biodosimetry Tool.

Authors:  Dakai Bian; Jason C Tsui; Mikhail Repin; Guy Garty; Helen Turner; Y Lawrence Yao; David J Brenner
Journal:  J Med Device       Date:  2016-08-31       Impact factor: 0.582

2.  Optimized automated data analysis for the cytokinesis-block micronucleus assay using imaging flow cytometry for high throughput radiation biodosimetry.

Authors:  M A Rodrigues; C E Probst; L A Beaton-Green; R C Wilkins
Journal:  Cytometry A       Date:  2016-06-06       Impact factor: 4.355

3.  Rapid and automatic detection of micronuclei in binucleated lymphocytes image.

Authors:  Xiang Shen; Ying Chen; Chaowen Li; Fucheng Yang; Zhanbo Wen; Jinlin Zheng; Zhenggan Zhou
Journal:  Sci Rep       Date:  2022-03-10       Impact factor: 4.379

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.