Ryan T Hill1, Klaudia M Kozek1, Angus Hucknall1, David R Smith1, Ashutosh Chilkoti1. 1. Department of Biomedical Engineering, Department of Electrical and Computer Engineering, Center for Metamaterials and Integrated Plasmonics, and Center for Biologically Inspired Materials and Material Systems, Duke University , Durham, North Carolina 27708, United States.
Abstract
The widespread use of plasmonic nanorulers (PNRs) in sensing platforms has been plagued by technical challenges associated with the development of methods to fabricate precisely controlled nanostructures with high yield and characterize them with high throughput. We have previously shown that creating PNRs in a nanoparticle-film (NP-film) format enables the fabrication of an extremely large population of uniform PNRs with 100% yield using a self-assembly approach, which facilitates high-throughput PNR characterization using ensemble spectroscopic measurements and eliminates the need for expensive microscopy systems required by many other PNR platforms. We expand upon this prior work herein, showing that the NP-film PNR can be made compatible with aqueous sensing studies by adapting it for use in a transmission localized surface plasmon resonance spectroscopy format, where the coupled NP-film resonance responsible for the PNR signal is directly probed using an extinction measurement from a standard spectrophotometer. We designed slide holders that fit inside standard spectrophotometer cuvettes and position NP-film samples so that the coupled NP-film resonance can be detected in a collinear optical configuration. Once the NP-film PNR samples are cuvette-compatible, it is straightforward to calibrate the PNR in aqueous solution and use it to characterize dynamic, angstrom-scale distance changes resulting from pH-induced swelling of polyelectrolyte (PE) spacer layers as thin as 1 PE layer and also of a self-assembled monolayer of an amine-terminated alkanethiol. This development is an important step toward making PNR sensors more user-friendly and encouraging their widespread use in various sensing schemes.
The widespread use of plasmonic nanorulers (PNRs) in sensing platforms has been plagued by technical challenges associated with the development of methods to fabricate precisely controlled nanostructures with high yield and characterize them with high throughput. We have previously shown that creating PNRs in a nanoparticle-film (NP-film) format enables the fabrication of an extremely large population of uniform PNRs with 100% yield using a self-assembly approach, which facilitates high-throughput PNR characterization using ensemble spectroscopic measurements and eliminates the need for expensive microscopy systems required by many other PNR platforms. We expand upon this prior work herein, showing that the NP-film PNR can be made compatible with aqueous sensing studies by adapting it for use in a transmission localized surface plasmon resonance spectroscopy format, where the coupled NP-film resonance responsible for the PNR signal is directly probed using an extinction measurement from a standard spectrophotometer. We designed slide holders that fit inside standard spectrophotometer cuvettes and position NP-film samples so that the coupled NP-film resonance can be detected in a collinear optical configuration. Once the NP-film PNR samples are cuvette-compatible, it is straightforward to calibrate the PNR in aqueous solution and use it to characterize dynamic, angstrom-scale distance changes resulting from pH-induced swelling of polyelectrolyte (PE) spacer layers as thin as 1 PE layer and also of a self-assembled monolayer of an amine-terminated alkanethiol. This development is an important step toward making PNR sensors more user-friendly and encouraging their widespread use in various sensing schemes.
Plasmonic
nanorulers (PNRs)—rulers
that transduce distance via a distance-dependent spectral shift of
a coupled plasmon resonance—have been in development for at
least a decade.[1−7] However, their widespread incorporation into robust sensing platforms
has been impeded by technical challenges associated with precisely
defining, modulating, and detecting nanoscale separation distances
between two surface plasmon-supporting structures. Nevertheless, the
development of PNRs for sensing remains attractive due to their exquisite,
angstrom-scale sensitivity to distance.[8] This distance sensitivity can be exploited in a manner similar to
Förster resonance energy transfer (FRET)-based sensors, but
in the case of PNRs the distance sensitivity extends over a longer
range[2] and signal transduction is more
stable since it is not susceptible to photobleaching.[9,10]The use of PNRs in sensing has been hindered by the lack of
a viable
strategy to create a PNR platform that is both efficient—in
fabrication and characterization—and robust. PNR sensors in
the form of colloidal nanoparticle (NP) assemblies[11,12] are efficient from the standpoint of high-throughput signal transduction,
because many PNRs can be interrogated with a single spectroscopic
measurement, and also from the standpoint of fabrication since the
process relies on self-assembly. However, the main disadvantage of
colloidal PNRs is that they are not robust. The stability of colloids
in suspension is extremely sensitive to salt concentration and other
solutes such as proteins and DNA, and so it is challenging to use
colloidal PNR sensors in direct contact with biological fluids such
as blood or serum.A more robust method to create PNR sensors
is to immobilize them
onto a substrate[2,3,7,13−15] to minimize the risk
of colloidal flocculation during a sensing experiment. Indeed, this
strategy has enabled PNRs to sensitively detect biomolecules, even
in complex media such as serum.[14,15] However, a major drawback
that prevents surface-based NP dimer PNR sensors from being widely
used is the low efficiency of their synthesis. As a result, their
characterization is often cumbersome and time-consuming, as it must
be done one PNR at a time using a microscope to identify single PNRs
among a large background population of plasmon resonant NPs that are
not PNRs.We recently demonstrated a surface-based version of
the PNR where
distances can be measured using plasmon coupling between gold NPs
and a thin gold film (NP–film).[8,16−18] Here, NP–film PNRs are created with 100% yield by simply
immobilizing NPs to the gold film with an intervening molecular spacer
that is typically deposited on the gold film before NP deposition.
To understand the optical properties of film-coupled NPs, the gold
film (analogous to a reflecting mirror) can be conceptually replaced
by an image dipole; light scattered from a film-coupled NP thus resembles
that scattered from two “real” metal NPs that are physically
separated in space.[18] Because the NP–film
PNR requires only chemically synthesized gold NPs to be tethered to
a bulk gold film by a molecular linker, and because these NPs are
easy to reproducibly synthesize with controlled sizes and shapes,
the NP–film PNR allows one to efficiently create, identify,
and interrogate millions of identical molecular rulers. Since the
NP–film PNR fabrication process results in such a uniform population
of PNRs, it is possible to simultaneously interrogate an ensemble
of PNRs with a single optical measurement, as opposed to having to
identify and characterize the rulers one-by-one, as is the case for
immobilized NP dimer PNRs. The significance of ensemble measurements
is that they dramatically increase the signal-to-noise ratio, analytical
precision and accuracy, and throughput of the PNR sensor.[8] The high efficiency of the NP–film format
combined with the robust nature of the surface-bound PNR platforms
suggests that the NP–film PNR is well suited for incorporation
into more user-friendly PNR sensing schemes.The NP–film
PNR still faces a significant challenge to be
useful for biosensing, which is the development of a strategy to make
high-throughput PNR measurements in an aqueous environment. This is
a nontrivial challenge because a reflective, metallic film is an integral
part of the NP–film PNR sensing platform. Initial configuration
of the platform[18] assumed the PNR to be
in air (dry state) and interrogated using reflected dark field microscopy,
which does not easily accommodate hydrated NP–film samples
due to the fact that, at least for the case of spherical NPs, NP–film
PNRs must be illuminated at an oblique angle so as excite the coupled
NP–film resonance. In reflected dark field imaging the numerical
aperture (NA) of the objective determines the illumination angle,
and so high-NA objectives provide better NP–film PNR excitation.
However, imaging through a water–air or glass–air interface
with high-NA reflected dark field objectives, which are not available
as immersion objectives, usually creates high background scattering
that degrades the NP–film PNR signal. To combat this problem,
we explored the use of totally internally reflected (TIR) white light
illumination coupled into the back side of a glass/gold film substrate
using a prism, which facilitates the use of liquid cells on top of
the gold film surface.[17,18] This approach was successful
to some degree, but it is spectrally complicated due to the strong
presence of the gold film SPR, which is also excited in the TIR configuration.The focus of this paper is to demonstrate a new and experimentally
convenient method to make high-throughput NP–film PNR measurements
in water using more standard laboratory equipment. We show that when
semitransparent gold films (30 nm thick) are used to create the NP–film
PNRs, they can be characterized in a transmitted light configuration,
where the ensemble NP–film extinction is measured using a standard
UV–vis spectrophotometer from an NP–film PNR sample
inserted into a cuvette. This measurement configuration is similar
to the previously demonstrated transmission localized surface plasmon
resonance spectroscopy (T-LSPR, also called T-SPR) configuration,[19−24] where the extinction of a population of plasmon resonant NPs immobilized
onto a transparent substrate is measured by a spectrophotometer. The
advantages of this technique are that (1) the instrumentation required
is standard equipment, as spectrophotometers are relatively inexpensive
and widely used; (2) the spectroscopic measurement is of an ensemble
of sensors, which increases accuracy, precision, and throughput; and
(3) the configuration is amenable to hydrated sensing experiments
given the use of a standard spectrophotometry cuvette as a sample
container and a transmitted light optical configuration. We present
herein a characterization and distance calibration of the hydrated
NP–film PNR in T-LSPR mode and then demonstrate that the NP–film
T-LSPR PNR can be used to detect dynamic, angstrom-scale thickness
changes in molecular spacer layers in response to changes in solution
pH. This demonstration should prove to be an important step toward
the incorporation of PNRs into useful sensing platforms that require
aqueous environments or into those where a collinear optical path
is favorable over a reflected geometry.Excitation of the coupled
nanoparticle–film (NP–film)
resonance depends on the illuminating beam having p-polarization or
a polarization dipole component that aligns along the axis defined
by the NPs and film. In a collinear transmitted light configuration
using unpolarized light, the illumination of a NP–film sample
at normal incidence results effectively in s-polarized illumination
and is incapable of exciting the coupled NP–film resonance
because all polarization dipoles in the propagating beam are orthogonal
to the NP–film resonant dipole axis (A). If the NP–film
sample is placed at an angle relative to the illumination path, then
the NP–film resonant dipole can be excited because some fraction
of the propagating beam contributes to p-polarization (B).
Results and Discussion
Excitation
of the coupled NP–film resonance depends on the
illuminating beam having a polarization component that aligns along
the axis normal to the surface of the gold film. We and others have
shown this previously by demonstrating that NP–film reflectivity
and surface-enhanced Raman measurements reveal strong interaction
with p-polarized incident light and minimal interaction with s-polarized
light.[8,17,25−27] Thus, the key to exciting the coupled NP–film resonance using
a transmitted, collinear optical extinction measurement is to place
the sample at an angle relative to the light path so that the propagating
beam has a polarization component that contributes to p-polarization
and is capable of exciting the NP–film dipole (Figure 1). If the NP–film axis, which is normal to
the film surface, is parallel to the axis of the propagating beam
(Figure 1A), then the illumination is effectively
s-polarized and is incapable of exciting the NP–film resonance
because the axes of the film-coupled NP dipoles are all orthogonal
to the electric field component of the beam. As soon as the NP–film
sample is rotated such that the NP–film axis is not parallel
to the axis of the propagating beam (Figure 1B), then the electric field component of the beam can contribute
to p-polarized illumination of the NP–film sample, hence exciting
the NP–film resonant dipoles.
Figure 1
Excitation of the coupled
nanoparticle–film (NP–film)
resonance depends on the illuminating beam having p-polarization or
a polarization dipole component that aligns along the axis defined
by the NPs and film. In a collinear transmitted light configuration
using unpolarized light, the illumination of a NP–film sample
at normal incidence results effectively in s-polarized illumination
and is incapable of exciting the coupled NP–film resonance
because all polarization dipoles in the propagating beam are orthogonal
to the NP–film resonant dipole axis (A). If the NP–film
sample is placed at an angle relative to the illumination path, then
the NP–film resonant dipole can be excited because some fraction
of the propagating beam contributes to p-polarization (B).
We demonstrate the effect
of illumination angle on the NP–film
LSPR measurement in both reflectivity and transmission measurements.
Reflectivity was measured using a variable-angle spectroscopic ellipsometer
(VASE), and transmission measurements were made using a UV–vis
spectrophotometer and custom slide holders (Figure 2A–C) to support NP–film samples inside standard,
1 cm spectrophotometer cuvettes such that the angular orientation
of the sample could be controlled relative to the incident light.
With these mounts, we were able to make NP–film LSPR extinction
measurements at a series of angles to compare the PNR response seen
from reflectivity and transmission illumination modes.
Figure 2
To demonstrate the effect of illumination angle on transmission
localized surface plasmon resonance (T-LSPR) spectroscopy of a coupled
nanoparticle–film (NP–film) sample, we designed slide
holders (A, B) that fit into standard 1 cm spectrophotometer cuvettes
(C). (D) Reflectivity spectra at various p-polarized illumination
angles relative to the normal to the gold film of a dry sample composed
of 60 nm gold NPs separated from 30 nm thick gold film (supported
by a glass microscope slide) by 1 polyelectrolyte (PE) spacer layer,
which positioned the NPs <1 nm away from the gold film. The coupled
NP–film resonance increases in intensity as the illumination
angle increases. The same sample is characterized in water using T-LSPR
spectroscopy (unpolarized illumination) and the slide holders in A
and B. (E) Similar to the reflectivity measurements in D, the coupled
NP–film resonance increases with increasing illumination angle
in the transmitted configuration. The coupled NP–film resonance
is not apparent in the spectrum when the illumination angle is 0°.
The spectra in E are baseline corrected so that they all have the
same average absorbance value in the noisy region between 850 and
900 nm that defines the base of the long-wavelength side of the NP–film
resonance peak. Spectra in D and E are normalized/baseline corrected
to a blank spectrum taken of a gold film sample containing the PE
spacer layer and no immobilized NPs. Insets in D and E depict the
respective detection modes (S: source, D: detector).
We created
an NP–film sample with a 30 nm gold film, a single
polyelectrolyte (PE) spacer layer (poly(allylamine) hydrochloride,
PAH), and 60 nm gold NPs at a surface coverage of ∼5 NPs per
μm2, which produced an NP–film separation
distance of <1 nm so as to obtain a red-shifted, coupled NP–film
LSPR. A reflected dark field image and corresponding scanning electron
micrograph showing NP surface coverage of such a sample can be seen
in the Supporting Information and shows
that the surface coverage of the NPs is high enough that single NPs
are not individually resolved in the optical image, but still low
enough that lateral plasmonic coupling between NPs is minimal.[17]To demonstrate the effect of illumination angle on transmission
localized surface plasmon resonance (T-LSPR) spectroscopy of a coupled
nanoparticle–film (NP–film) sample, we designed slide
holders (A, B) that fit into standard 1 cm spectrophotometer cuvettes
(C). (D) Reflectivity spectra at various p-polarized illumination
angles relative to the normal to the gold film of a dry sample composed
of 60 nm gold NPs separated from 30 nm thick gold film (supported
by a glass microscope slide) by 1 polyelectrolyte (PE) spacer layer,
which positioned the NPs <1 nm away from the gold film. The coupled
NP–film resonance increases in intensity as the illumination
angle increases. The same sample is characterized in water using T-LSPR
spectroscopy (unpolarized illumination) and the slide holders in A
and B. (E) Similar to the reflectivity measurements in D, the coupled
NP–film resonance increases with increasing illumination angle
in the transmitted configuration. The coupled NP–film resonance
is not apparent in the spectrum when the illumination angle is 0°.
The spectra in E are baseline corrected so that they all have the
same average absorbance value in the noisy region between 850 and
900 nm that defines the base of the long-wavelength side of the NP–film
resonance peak. Spectra in D and E are normalized/baseline corrected
to a blank spectrum taken of a gold film sample containing the PE
spacer layer and no immobilized NPs. Insets in D and E depict the
respective detection modes (S: source, D: detector).Figure 2D shows the extinction
of the dry
NP–film sample measured in reflectivity mode using p-polarized
light and with illumination angle varying from 45° to 70°
relative to the normal to the gold film surface. In each case, the
sample spectrum is normalized to a blank spectrum of a gold film containing
1 PE layer and no immobilized NPs. The general trend in this data
is that the NP–film extinction peak intensifies as the angle
is increased. There are two reasons for this trend in the data. The
first reason is that as the angle of illumination increases, the number
of film-coupled NPs interrogated by the beam increases due to the
elongation of the illuminated spot on the substrate. The second and
more significant reason for the trend in the data is that as the illumination
angle increases, the polarization of the excitatory electric field
in the illumination beam is better aligned with the NP–film
resonant dipole, which is aligned along the axis normal to the surface
of the gold film.Using the same NP–film sample with
a 1-PE spacer layer,
we collected NP–film extinction spectra in T-LSPR mode using
a spectrophotometer with the samples inserted into the custom slide
holders contained inside of water-filled cuvettes, which positioned
the sample at various illumination angles (Figure 2E). Here again, the sample spectra were baseline corrected
to blank spectra collected at each angle from samples containing 30
nm gold film, 1 PE layer, and no immobilized NPs. We used unpolarized
light from the spectrophotometer in this case so as to make the T-LSPR
measurement in the simplest possible optical configuration, though
we note that a polarizer can be easily added to the optical path.
When the sample was positioned such that the illumination angle was
at 0° relative to the gold film surface normal (i.e., such that the NP–film resonant dipoles are parallel to
the propagating beam), an illumination angle that was not possible
in reflectivity mode due to instrument limitations, the red-shifted
NP–film resonance was not apparent at all in the spectrum.
There was, however, a weak resonance peak located at ∼545 nm,
which we attribute to a slight excitation of the uncoupled 60 nm NP
LSPR. The absence of the coupled NP–film resonance is due to
the illumination in this case being effectively s-polarized and incapable
of exciting the coupled NP–film resonance. It is not until
the illumination angle increases that the red-shifted NP–film
coupled resonance appears at ∼740 nm. The NP–film resonance
in this case is red-shifted relative to the reflectivity data because
the T-LSPR data were taken in water. We note that in the T-LSPR spectra
we collected throughout these experiments, it was not uncommon for
the overall baseline signal levels to vary slightly from sample to
sample. We attribute this to a combination of instrument baseline
drift and the slight variation in sample placement that is caused
by the tolerance of the 3D printed slide holders. Thus, the increasing
NP–film extinction peak with increasing illumination angle
presented in Figure 2E becomes more apparent
and similar to that seen in the reflectivity data (Figure 2D), after the spectra are baseline adjusted to a
feature common to all the spectra, which we chose to be the average
extinction value of the noisy region between 850 and 900 nm that defines
the base of the long-wavelength side of the NP–film extinction
peak.The data presented in Figure 2 show
that
the ability to excite the coupled NP–film resonance in transmission
mode is crucially dependent on the illumination angle being such that
the axis of the coupled NP–film dipole is not parallel to the
propagating illumination beam, so that the illumination contributes
to p-polarized excitation. It is interesting to note that similar
T-LSPR measurements have been made on NP–film samples,[28−32] and in some cases a distinct, coupled NP–film resonance was
not observed. In some instances,[28,31] the films
were extremely thin (4 nm) and described as gold nanoislands, which
were being coupled to a dense, close-packed network of very small
NPs (12–13 nm diameter) entrapped within polymer brushes to
make sensitive plasmonic pH sensors. Since the film in this case was
likely not continuous and the NPs likely penetrate to some degree
into the polymer brush films to form a complex plasmonically coupled
3D network,[33] the coupled resonant dipoles—both
NP–NP coupling and NP–nanoisland coupling—responsible
for the sensor signal were likely varying dramatically in orientation.
We presume that the apparent shift in the native LSPR of the Au nanoislands
with the addition of nearby NPs into the polymer brushes was measured
with some degree of NP–NP and NP–nanoisland LSPR coupling
from resonant dipoles that could be excited by s-polarized illumination
at normal incidence. Polymerswelling was detected by a plamson shift
from the ensemble, networked NP assembly as the polymer brushes changed
configuration.In systems where the coupled LSPR dipoles vary
in orientation,
it is difficult to correlate the plasmon shifts observed to precise
changes in the coupled LSPR separation distance, which complicates
calibration of the PNR and hinders the ability to perform PNR measurements
where distance changes are measured. These systems work well for detecting
polymerswelling, but they depend on the ability to form close-packed
NP networks, which is aided by the use of NP multilayers or 3D branched
polymer brushes containing many NP binding sites on each brush. They
would likely suffer from lack of signal in cases where a molecular
layer binding NPs to film does not promote the formation of an immobilized
high-density, close-packed NP network (e.g., a monolayer of biomolecules
designed to bind 1 NP per molecule) due to the lack of coupled resonant
dipoles excitable by s-polarized illumination at a 0° illumination
angle.These previous NP–film T-LSPR studies are drastically
different
from the NP–film system presented in this paper, where much
larger, 60 nm NPs are used at much lower film surface coverage (∼5
NPs per μm2), and all of the coupled NP–film
resonant dipoles have the same orientation (aligned normal to the
gold film surface), excitable by p-polarized illumination. The results
we see in our NP–film T-LSPR measurements are most similar
to those observed by Okamoto and Yamaguchi[32] in an early characterization of the plasmonic coupling between gold
NPs of various sizes and 25 nm thick gold film, where they observed
distinct, coupled NP–film resonances in addition to native
NP LSPRs with increasing T-LSPR illumination angle and p-polarized
illumination. In such a system where the film-coupled NPs are all
separated from the underlying continuous film by a uniform separation
distance, it is straightforward to correlate shifts in the coupled
NP–film LSPR to changes in the NP–film separation distance
using a similar approach to our previous reports on NP–film
PNRs.[8,16−18] In this NP–film
system, the coupled resonance responsible for PNR signal transduction
arises from the coupling of single NPs to the underlying film, with
minimal contribution from lateral NP–NP coupling,[17] and hence it will not suffer from the use of
molecular spacer layers that do not promote dense, close-packed NP
immobilization.We next used the NP–film PNR in the T-LSPR
configuration
to interrogate small, dynamic molecular conformational changes by
measuring the actuation of molecular spacer layers in response to
solution pH. It has been shown previously that PE multilayers containing
at least one weak PE, such as PAH, will swell and deswell in response
to solution pH changes[34−37] because of protonation and deprotonation as a function of solution
pH. While this swelling behavior has been shown previously using PE
multilayers consisting of many bilayers, we show here how the NP–film
PNR can be used to probe this actuation using only 1–3 PE layers
(i.e., 0.5–1.5 bilayers), which highlights the unique ability
of the NP–film PNR to sense distance changes from very thin
molecular layers.(A) A nanoparticle–film (NP–film) plasmon
nanoruler
(PNR) is used in transmission localized surface plasmon resonance
spectroscopy (T-LSPR, 60° illumination angle) mode to characterize
pH-induced swelling of a molecular spacer layer composed of 3 polyelectrolyte
(PE) layers. (B) Spectra show the coupled NP–film LSPR blue-shifted
by incubation at pH 2, suggesting that the 60 nm gold NPs have been
pushed away from the underlying 30 nm gold film because the PE layer
is in a swollen state. The NP–film LSPR red shifts when the
sample is immersed in pH 12 solution, suggesting PE layer deswelling.
(C) The NP–film LSPR peak centroid is plotted over the course
of 8 pH switches, indicating semireversible pH-induced layer swelling.
The data suggest that the PE layer experiences a net contraction after
8 pH switches, each of which consisted of a 30 min incubation time
at the respective pH.Figure 3 shows the NP–film
PNR T-LSPR
response from a pH swelling experiment in which we observed the actuation
of a 3-PE layer (3L) sample over eight pH switches. We used layer-by-layer
(LBL) deposition of oppositely charged PEs,[38,39] PAH, and polystyrenesulfonate (PSS), with PAH being the initial
and terminal layers and also the weak PE that is susceptible to protonation
with a change in pH. For this initial experiment, we deposited all
three layers (Figure 3A) from solutions containing
3 mM of the PE (with respect to the monomer molecular weight) and
1 M NaCl dissolved in ultrapure water and made no attempt to adjust
the pH of the deposition solutions.
Figure 3
(A) A nanoparticle–film (NP–film) plasmon
nanoruler
(PNR) is used in transmission localized surface plasmon resonance
spectroscopy (T-LSPR, 60° illumination angle) mode to characterize
pH-induced swelling of a molecular spacer layer composed of 3 polyelectrolyte
(PE) layers. (B) Spectra show the coupled NP–film LSPR blue-shifted
by incubation at pH 2, suggesting that the 60 nm gold NPs have been
pushed away from the underlying 30 nm gold film because the PE layer
is in a swollen state. The NP–film LSPR red shifts when the
sample is immersed in pH 12 solution, suggesting PE layer deswelling.
(C) The NP–film LSPR peak centroid is plotted over the course
of 8 pH switches, indicating semireversible pH-induced layer swelling.
The data suggest that the PE layer experiences a net contraction after
8 pH switches, each of which consisted of a 30 min incubation time
at the respective pH.
Figure 3B shows representative NP–film
T-LSPR data at each pH value. The NP–film coupled resonance,
which is the strong peak between 650 and 700 nm, red shifts when the
pH is switched from 2 to 12. This implies that at pH 2 the NPs on
top of the 3L sample are further away from the film, because the PE
multilayer is swollen, and that at pH 12 the NPs are closer to the
film because the PE multilayer has contracted. This swelling trend
is consistent with previous studies using multiple PE bilayers.[34,35] Once the NP–film LSPR peak location is plotted versus pH
(Figure 3C), we see that there is a semireversible
trend in the peak shift throughout the eight pH cycles. In the initial
cycle, the switch from pH 2 to 12 results in a larger red shift than
the blue shift that occurs when the pH is switched back from 12 to
2. This trend is consistent throughout the remainder of the cycles
and implies that the 3L multilayer experienced a net contraction over
the eight successive pH cycles.To correlate the NP–film
LSPR peak shifts we observed during
the pH switches with changes in the molecular spacer layer thickness,
we conducted an NP–film PNR distance calibration study similar
to those we have demonstrated previously[8,17] with the exception
that this study was done in aqueous solution using both reflectivity
(Supporting Information) and T-LSPR NP–film
extinction measurements. In the past we have only been able to perform
this distance calibration of the NP–film PNR in the dry state,
using either reflected dark field scattering or reflectivity extinction
measurements. We created a series of NP–film samples with PE
spacer layers of increasing thickness using LBL deposition of PEs.
We used in situ ellipsometry to estimate the thicknesses
of the hydrated PE spacer layers (Supporting Information), which are plotted along with the hydrated NP–film T-LSPR
data to create a calibration curve.Nanoparticle–film (NP–film)
plasmon nanoruler (PNR)
transmission localized surface plasmon (T-LSPR) spectroscopy response
using 60 nm Au NPs and 30 nm Au film was calibrated in water by creating
NP–film samples using polyelectrolyte (PE) spacer layers with
increasing thickness. (A) T-LSPR spectra (60° illumination angle,
p-polarized) show the NP–film extinction spectra using PE spacer
layers varying in thickness from 1 to 15 layers. The coupled NP–film
LSPR blue shifts with increasing number of spacer layers. (B) The
NP–film LSPR peak position is plotted versus NP–film
separation distance, which is set by the PE spacer layer thickness,
and shows the expected nonlinear blue shift of the NP–film
LSPR with increasing NP–film separation distance. These data
were fitted to a power law function (line plotted in B) and serve
as a calibration curve for the NP–film PNR interrogated by
T-LSPR spectroscopy.The NP–film T-LSPR data from the calibration study
done
in water is shown in Figure 4. Similar data
from both hydrated and dry PE layers in reflectivity and transmission
modes are shown in the Supporting Information. We were able to identify a distinct NP–film coupled resonance
in the T-LSPR extinction spectra (Figure 4A)
for the NP–film samples with spacer layers extending out to
15 PE layers, which corresponded to a 34 nm separation distance between
the NPs and film. Spectra from the 7L, 9L, and 11L (“nL” where “n” is the
number of layers) NP–film samples produced oddly shaped NP–film
extinction peaks, which we attribute to a spectral mixing effect between
the uncoupled 60 nm NP LSPR and the coupled NP–film LSPR. Interestingly,
this spectral mixing effect was not as apparent in the data taken
in reflectivity mode or in the dry T-LSPR data (see Supporting Information for more discussion). We were able
to minimize the effect of these features by inserting an inexpensive
linear polarizer sheet into the spectrophotomer light path so as to
deliver only p-polarized illumination to the NP–film samples,
which preferentially excites the coupled NP–film resonance
and, hence, produces more accurate NP–film LSPR peak positions
to use for the distance calibration. We did not use the polarizer
for the studies presented in Figure 3 or 5, because the use of the polarizer does not significantly
alter the peak location for thin NP–film spacer layers such
as the ones used in these studies (Supporting
Information).
Figure 4
Nanoparticle–film (NP–film)
plasmon nanoruler (PNR)
transmission localized surface plasmon (T-LSPR) spectroscopy response
using 60 nm Au NPs and 30 nm Au film was calibrated in water by creating
NP–film samples using polyelectrolyte (PE) spacer layers with
increasing thickness. (A) T-LSPR spectra (60° illumination angle,
p-polarized) show the NP–film extinction spectra using PE spacer
layers varying in thickness from 1 to 15 layers. The coupled NP–film
LSPR blue shifts with increasing number of spacer layers. (B) The
NP–film LSPR peak position is plotted versus NP–film
separation distance, which is set by the PE spacer layer thickness,
and shows the expected nonlinear blue shift of the NP–film
LSPR with increasing NP–film separation distance. These data
were fitted to a power law function (line plotted in B) and serve
as a calibration curve for the NP–film PNR interrogated by
T-LSPR spectroscopy.
Figure 5
A nanoparticle–film (NP–film) plasmon nanoruler (PNR)
was used to investigate the pH-induced swelling behavior of different
molecular spacer layers. The parameter “PNR measurement”,
which is determined by converting the NP−film LSPR peak positions
to NP−film separation distance using the calibration curve
determined in Figure 4B, is plotted against
pH for each spacer layer. Swelling of polyelectrolyte (PE) spacer
layers fabricated in the presence of 1 M NaCl (dark green squares)
and also without NaCl and with pH adjusted to pH 10 (designated as
“NS-pH10”, light green open circles) was compared. The
3 PE layer samples (“3L”, C) show differing initial
thicknesses and similar semireversible pH-induced swelling behavior
(A). Both 1L (D) samples have similar initial thickness values and
showed qualitatively similar swelling to the 3L samples (B). The NS-pH10
samples on average show a higher degree of swelling and deswelling
relative to the same layers fabricated from solutions containing 1
M NaCl. A self-assembled monolayer of an 11-carbon amine-terminated
thiol (C11 amine thiol, E) also demonstrated pH-induced swelling behavior
(B, blue triangles).
Figure 4B shows
the coupled NP–film
LSPR peak position, as determined by calculating the centroid of the
top 50% of the peaks, plotted against the NP–film separation
distance, which shows the expected nonlinear blue-shifting LSPR trend
with distance. We fitted these data to a power law function[8,17,40] by performing a linear regression
(R2 = 0.972 49) of the data plotted
on a log–log scale (Supporting Information). These regression data for the T-LSPR data were used as a calibration
curve to estimate dynamic changes in thickness from hydrated molecular
layers in T-LSPR mode. The calibration curve determined from the fitting
procedure was y = 768.635x–0.08692 and is shown as a solid line in Figure 4B.With the NP–film PNR T-LSPR calibration curve as reference
to convert wavelength shifts in the NP–film LSPR to changes
in distance, we used the NP–film PNR to compare the pH-induced
swelling behavior of molecular spacer layers that differ in thickness
and fabrication conditions. It has been suggested previously that
the swelling behavior and thickness of PE multilayers are affected
by the pH and ionic strength of the deposition solution.[35−37,41] We used the NP–film PNR
to investigate the pH swelling behavior of very thin PE layers, consisting
of 3 PE layers (Figure 5C) and only 1 PE layer
(Figure 5D). For both PE layer thicknesses,
we also compared the swelling behavior of layers deposited from PE
solutions containing 1 M NaCl in ultrapure water with those deposited
from solutions that contained no additional NaCl and were adjusted
to be pH 10, which is above the pKa of
PAH (pKa ∼9;[37] these layers are designated “NS-pH10”, where
“NS” stands for “no salt”). Each of the
layers produced distinct dry ellipsometric thicknesses and corresponding
NP–film LSPR peak positions after NP deposition: 24.7 Å
and 612.9 nm for 3L, 20.1 Å and 623.9 nm for 3L NS-pH10, 3.8
Å and 654.3 nm for 1L, and 5.0 Å and 654.8 nm for 1L NS-pH10.
The 3L samples showed a difference in thickness between the two deposition
conditions, as indicated by the ellipsometric thickness and by the
NP–film LSPR position. This observation is supported by previous
studies showing that PE layers generally form thicker layers when
deposited from solutions containing higher salt concentration.[41] Though the 1L samples showed a slight thickness
variation (1.2 Å) in the ellipsometry data, the dry NP–film
PNR LSPR peak positions differ only by 0.5 nm (corresponding roughly
to 0.1 Å in thickness using the dry NP–film PNR sensitivity
determined in Hill et al.[8]), which suggests
that these 1L layers are quite similar in thickness. This discrepancy
between the ellipsometry and PNR data is not surprising, as we have
found that our ellipsometric thickness values below ∼10 Å
are not as accurate due to inherent limitations of ellipsometry.[8]A summary of the pH-induced swelling of
the multiple layers characterized
by the NP–film PNR in T-LSPR mode is shown in Figure 5A and B. The PNR measurement was determined by tracking
the coupled NP–film LSPR peak position for each layer type
at each pH and then using the calibration curve generated by our calibration
study (Figure 4B) to convert the peak positions
to NP–film separation distances, which is designated as “PNR
measurement”. These data are from hydrated NP–film samples,
and even so, the thickness trends we observed from the dry layers
discussed above are also observed for the hydrated layers. The 3L
sample fabricated in the non-pH-adjusted solutions containing salt
was significantly thicker than the 3L NS-pH10 sample, both initially
when immersed in pH 2 solution and throughout the pH cycles. Both
1L samples also had similar thicknesses initially at the first pH
2 measurement. Also, similar to the data from Figure 3, all layers demonstrated semireversible swelling such that
after the 8 pH incubations all layers appeared to have experienced
a net contraction relative to their initial thickness as measured
at pH 2. We note that we have cycled a 1L sample through 16 pH switches
over the course of 2 days and observed that the net layer contraction
over successive pH cycles seems to approach an asymptotic limit after
many pH cycles (Supporting Information).A striking feature that is consistent throughout all of the PE
spacer layer pH data in Figure 5 is that the
NS-pH10 layers showed increased deswelling and swelling throughout
the pH cycles relative to the layers that were deposited from solutions
containing salt and ultrapure water that was not pH adjusted. The
average deswelling and swelling between each pH switch for the 3L
sample in terms of layer thickness were 4.4 Å (deswelling) and
2.3 Å (swelling), respectively, whereas the same figures for
the 3L NS-pH10 sample were 5.0 Å (deswelling) and 3.4 Å
(swelling). For the 1L samples, the averages were 1.8 Å (deswelling)
and 0.5 Å (swelling) for the 1L sample and 2.5 Å (deswelling)
and 1.3 Å (swelling) for the 1L NS-pH10 sample. The 3L NS-pH10
samples experienced 14% more deswelling and 48% more swelling relative
to the 3L samples, and the 1L NS-pH10 samples experienced 41% more
deswelling and 152% more swelling relative to the 1L samples.Further study is required to firmly establish the trends in PE
layer swelling observed from these NP–film PNR experiments
and also to elucidate the effect, if any, of the NP stabilizing ligand
(citrate) on the PNR response to pH changes. Nevertheless, the trends
presented here qualitatively support those seen from previous studies
showing pH-induced swelling of PE layers.[34,35] It has been shown before that PE multilayers containing PAH and
PSS show exaggerated pH-induced swelling when the multilayers are
assembled from alkaline solutions near the pKa of PAH.[35−37] This is likely due to the increased incorporation
of free amines into the PE multilayers, which creates a large amount
of electrostatic repulsion when the pH is reduced to the point that
all amines become protonated and positively charged. However, in contrast
to previous reports,[35−37] our data suggest that the PE layers fabricated from
relatively acidic solutions that were not pH adjusted to 10 are also
capable of pH-induced swelling. We note here that the relatively acidic
fabrication solutions used in this study also contained 1 M NaCl,
whereas the pH < 10 fabrication solutions in the previous studies
did not.A nanoparticle–film (NP–film) plasmon nanoruler (PNR)
was used to investigate the pH-induced swelling behavior of different
molecular spacer layers. The parameter “PNR measurement”,
which is determined by converting the NP−film LSPR peak positions
to NP−film separation distance using the calibration curve
determined in Figure 4B, is plotted against
pH for each spacer layer. Swelling of polyelectrolyte (PE) spacer
layers fabricated in the presence of 1 M NaCl (dark green squares)
and also without NaCl and with pH adjusted to pH 10 (designated as
“NS-pH10”, light green open circles) was compared. The
3 PE layer samples (“3L”, C) show differing initial
thicknesses and similar semireversible pH-induced swelling behavior
(A). Both 1L (D) samples have similar initial thickness values and
showed qualitatively similar swelling to the 3L samples (B). The NS-pH10
samples on average show a higher degree of swelling and deswelling
relative to the same layers fabricated from solutions containing 1
M NaCl. A self-assembled monolayer of an 11-carbon amine-terminated
thiol (C11 amine thiol, E) also demonstrated pH-induced swelling behavior
(B, blue triangles).We also tested the susceptibility of an amine-terminated
11-carbon
alkanethiol (C11 amine thiol) self-assembled monolayer (SAM) to pH-induced
changes in NP–film separation distance using the NP–film
PNR in T-LSPR mode (Figure 5E). The PNR data
report a surprising degree of actuation (Figure 5B, blue triangles), as we expected the amine thiol SAM spacer layer
to be more well-ordered and less prone to electrostatic reorganization
relative to the disorderedPE spacer layers. However, the data we
obtained from the NP–film PNR suggest that the amine thiol
layer changes similarly to the PE layers, increasing the NP–film
separation distance at pH 2 and decreasing it at pH 12.We propose
two mechanisms that may contribute to the pH-induced
modulation of the NP–film separation distance observed for
the amine thiol SAM spacer layer. The first is that the PNR may be
transducing differences in the relative affinity of the amine group
toward the gold NPs at the different pH values. We assume here that
the gold NPs associate directly with the amine groups of the SAM due
to the known affinity of amine groups toward gold[42,43] and the propensity of the citrate anions on the gold NPs to be easily
displaced by amine groups.[44] However, it
is believed that protonated amine groups do not interact as strongly
with gold surfaces due to their loss of the free electron pair upon
protonation.[42,45] Thus, we hypothesize that at
extremely low pH values the amines become fully protonated and start
to lose their affinity toward the gold NPs, which may encourage the
NPs to move away from the SAM and underlying gold film. As the pH
increases, the amines become fully deprotonated, which causes a tighter
association with the gold NPs and pulls the NPs toward the SAM and
gold film. The second mechanism we propose for the observed pH-induced
swelling of the amine thiol SAM, which seems more likely than the
first proposed mechanism, is that the SAM itself reorganizes in response
to pH changes due to protonation and deprotonation of the terminal
amine groups. It is commonly accepted that when SAMs form, the thiol
molecules assume some degree of tilt relative to the gold surface,[46,47] which represents a molecular arrangement with a minimum free energy.
In the case of amine thiolSAMs, it is reasonable to expect that the
packing and orientation of the molecules within the SAM depend on
the charge state of the amine groups and that the SAM can reorganize
slightly with changes in the charge state. The electrostatic reorganization
of SAMs with charged terminal functional groups has been investigated
in previous reports.[48−50] We suggest that when the amine thiol SAM is immersed
in pH 2 solution, the SAM reorganizes slightly and effectively increases
in thickness due to charge repulsion of the protonated, positively
charged amine groups, and at pH 12, the SAM assumes a more compact
configuration when the amines are fully deprotonated and uncharged.
Conclusions
The results presented herein are an important step forward in making
the PNR technology more useful and accessible to researchers by adapting
it for cuvette-based sensing studies that can be performed with standard
spectrophotometers using T-LSPR spectroscopy. The distinctive feature
of the platform we describe here relative to previous reports of T-LSPR
measurements involving plasmonic coupling is that this NP–film
system—instead of relying on the ability to form close-packed
NP networks for adequate LSPR coupling response—simply uses
single 60 nm NPs plasmonically coupled to a 30 nm film, which produces
uniformly aligned resonant dipoles that shift in wavelength very sensitively
with subtle changes in NP–film separation distance. The coupled
NP–film LSPR, and hence the precise NP–film separation
distance, can be probed directly by T-LSPR measurements as long as
the sample is placed at an angle relative to the illumination beam
such that the NP–film axis is not parallel to the illumination
beam and some degree of p-polarization is achieved. To enable high-throughput
NP–film PNR studies that can be conducted in aqueous solutions,
we used cuvette slide holders to place the NP–film PNR samples
at an optimal angle relative to the collinear optical path defined
by the spectrophotometer. This allows NP–film PNR samples to
be immersed in various solutions in the cuvette and enables measurement
of the response from millions of NP–film PNRs simultaneously
by a single spectroscopic measurement from a standard UV–vis
spectrophotometer. We calibrated the distance sensitivity of the T-LSPR
mode NP–film PNR in water, which we used to suggest pH-induced
angstrom-scale swelling and deswelling of PE spacer layers as thin
as 1 PE layer and also from a thin amine-terminated SAM. The ability
to perform very sensitive, high-throughput PNR measurements in aqueous
solution on dynamically changing molecular spacer layers using a UV–vis
spectrophotometer is a first step in the quest to design technologically
simple sensors that take advantage of the exquisite distance sensitivity
of the NP–film PNR.
Methods
Molecular Spacer Layer
Preparation
Gold films and the
molecular spacer layers with attached gold nanoparticles were prepared
and characterized similarly to our previously published articles.[8,17] Thin gold films of 30 nm were deposited in a class 100 clean room
onto Nexterion Glass B slides (“clean room cleaned”,
Schott North America, Inc.) by an electron beam evaporator (CHA Industries)
at 2 Å/s using a 5 nm chromium adhesion layer (deposited at 1
Å/s). Gold films were stored in 200 proof ethanol at 4 °C
until use in experiments, as we have found previously that using more
rigorous cleaning methods to completely remove all organic contaminants
can have detrimental effects on our ability to produce good NP–film
samples.[8] We use 30 nm gold films in this
study as a compromise between our requirements of (1) maximizing the
light transmission through the film and (2) producing a strong, coupled
NP–film LSPR peak. We have found previously that 15 nm gold
films produce a relatively weak and broadened coupled NP–film
LSPR relative to 30 nm gold films. We avoid using films thicker than
30 nm because the 30 nm films produce an acceptable coupled NP–film
response and adding thickness to the film would decrease the amount
of light that can pass through the film, which would hinder T-LSPR
NP–film measurements.Polyelectrolyte spacer layers were
prepared by layer-by-layer deposition[38,39] of poly(allylamine)
hydrochloride (MW = 70 kDa, Aldrich) and polystyrenesulfonate (MW
= 70 kDa, Aldrich). For each deposition step, the gold-coated glass
slides were immersed in 0.003 mol-of-monomer/L PE and 1 M NaCl for
5 min, rinsed thoroughly with a gentle stream of ultrapure water (18
MΩ, used throughout), and immersed in fresh ultrapure water
for 1 min, after which the substrates were either immersed in 1 M
NaCl for 30 s before repeating the same steps for deposition of the
oppositely charged PE or dried with a stream of high-purity nitrogen
for analysis. No pH adjustments were made to the fabrication or rinse
solutions. Solutions of PSS and PAH with 1 M NaCl were found to be
pH 5.6 and 5.7, respectively. In the case of the NS-pH10 samples,
the deposition procedure differed in that the solutions of PEs did
not contain any additional NaCl and were adjusted to pH 10 using concentrated
NaOH and sulfuric acid, and the brief NaCl rinse was replaced with
a brief rinse in pH 10 water. All LBL depositions were initiated and
terminated with the cationic PAH layer to facilitate both the attachment
of the first PE layer to the gold film through amine–gold interactions[42,43] and the electrostatic immobilization of gold NPs to the PE spacer
layer.Self-assembled monolayers were fabricated on gold films
using 11-amino-1-undecanethiol
hydrochloride (referred to in this article as “C11 amine thiol”,
Sigma 674397, used as received). The SAMs were fabricated by incubating
a gold slide in a clean glass vial containing a ∼2 mM thiol
solution in 200 proof ethanol for 18 h. Following incubation, the
vials containing the gold slides and thiol solutions were sonicated
in a water bath at low power (power “4” out of “10”
using a Crest Ultrasonics model 230D sonicator) for 2 min and then
overflow rinsed with five reaction volumes of 200 proof ethanol. This
sonication and rinsing step was performed a total of four times for
each slide before removing the slide from the ethanol solution and
drying it with a stream of high-purity nitrogen.For most studies
in this article, gold film samples were cut to
size prior to deposition of molecular spacer layers and NPs. However,
for the LBL calibration study, PE layers were deposited onto full
gold film coated slides and then later cut into the pieces necessary
to make dry and hydrated ellipsometry, spectroscopic reflectivity,
and absorption measurements.
NP Deposition
Gold NPs (BBI) of
60 nm were electrostatically
immobilized on each molecular spacer layer. Deposition of the gold
NPs onto the PE layers was done by applying drops of the undiluted
stock solution of gold NPs to each functionalized gold film for an
incubation time of 30 min and then rinsing with ultrapure water and
drying with a stream of high-purity nitrogen. This NP deposition results
in 5–6 scatterers per μm2, with 98% of them
being single NPs on gold film (Supporting Information).
Ellipsometry
Molecular spacer layer thicknesses were
characterized using a J.A. Woollam Co., Inc., M-88 variable-angle
spectroscopic ellipsometer and WVASE32 software (version 3.460). Spectroscopic
scans (277.5–763 nm) of each spacer layer were performed in
three distinct regions of the functionalized gold films that did not
contain immobilized NPs at 65°, 70°, and 75° relative
to the normal of the surface of the slide. Ellipsometry data were
analyzed using a two-layer model[51] composed
of a bulk gold layer underneath an organic layer, which was used to
represent the molecular layer. The thickness of each spacer layer
was fitted using the Cauchy expression for normal dispersion[52] provided by the WVASE32 software using the default
values for all parameters, including 1.45 for the parameter “A” representing the long wavelength asymptotic refractive
index of the organic layer, such that the mean standard error of the
fit was minimized. The nominal thickness of each spacer layer was
determined to be the average of three independent thickness measurements
of the spacer layer. The optical constants of the bare gold film were
determined immediately prior to LBL deposition by spectroscopic scans
of the bare gold films at 65°, 70°, and 75° and fitting n (refractive index) and k to the known
values of bulk gold, which were provided by the WVASE32 software,
to account for any shifts in the optical constants due to variation
in the thickness of the gold films. These fitted optical constants
for each gold slide were saved and used later, respectively, when
fitting for thickness of the molecular layers deposited onto the gold
slides.
In Situ Ellipsometry
Hydrated thicknesses
of the PE spacer layers used in the LBL calibration study were measured
using the same ellipsometry system described above in the Ellipsometry section. Inspired by liquid cell
designs from Richter’s group,[53,54] we fabricated
a liquid cell by first creating a 3D printed mold (printed using a
Stratasys Dimension 1200ES) to hold glass slides in place so as to
create optical windows oriented to accept 70° illumination from
the ellipsometer. The glass slides were then sealed to form a chamber
with an open top using aquarium safe silicone sealant (see the Supporting Information for more detail). The
WVASE32 software was used along with a silicon standard to calculate
and account for the optical effects of the liquid cell on the ellipsometry
measurements. With the cell in place, containing a sample and filled
with water, thickness measurements were made similar to the dry state,
but using only a 70° illumination angle. The thicknesses were
fitted using the same model described above, but included a top ambient
layer of water in the model.
NP–Film Plasmonic Characterization
Plasmonic
properties of the NP–film samples were characterized by spectroscopic
reflectivity and transmitted extinction measurements on dry and hydrated
samples. Reflectivity spectra were acquired using the VASE instrument
described above. Aside from the angle study presented in Figure 2D, all reflectivity spectra were collected using
an illumination beam directed at 70° relative to the gold film
surface normal. Integration time was controlled by the “Revs/Meas”
variable in the WVASE32 software, which was kept constant at 400 for
all reflectivity spectra. The illumination beam was provided by the
xenon lamp included with the instrument and had a beam diameter of
∼4 mm. Dry samples were simply placed on the ellipsometer stage
for measurement, and hydrated samples were placed inside of the water-filled
ellipsometry liquid cell described above. The spectral range of this
VASE instrument is limited, which is why the spectra appear to be
cropped at 775 nm. This is simply an instrument limitation that we
accepted due to the ease with which reflectivity spectra at various
angles can be acquired using a VASE. Reflectivity spectra containing
a wider wavelength range from similar NP–film samples than
those present here can be seen in Mock et al.[17]Transmitted extinction spectra were collected using a Cary
300 (Varian, Inc.) spectrophotometer equipped with a temperature controller.
All spectra were collected at 25 °C using a 0.2 s/nm integration
time. NP–film samples were inserted into standard size (1 cm
path length) disposable plastic cuvettes containing custom slide holders
that positioned the NP–film samples at various angles relative
to the beam path of the instrument (Figure 2A–C). The NP–film samples were positioned so that the
illumination beam first passed through the gold film (with immobilized
NPs) and then through the supporting glass slide. The beam size at
the sample was set using the spectral bandwidth software parameter
of 2, which produced a beam ∼3 mm wide and ∼7 mm tall.
The slide holders were designed using Autocad and 3D printed from
ABSplus thermoplastic using a Stratasys uPrint SE Plus. After 3D printing
and support removal, the slide holders were sonicated in ultrapure
water containing sodium dodecyl sulfate for 1 h, soaked overnight
in the same solution, and then thoroughly rinsed with ultrapure water
before use. Aside from the angle study presented in Figure 2E, all extinction spectra were obtained using the
60° slide holders. Even though the NP–film LSPR peak is
stronger in T-LSPR spectra acquired at greater illumination angles
(Figure 2E), we used the 60° holder to
ensure that the transmitted beam was not clipped by the width of the
NP–film samples, which effectively narrows with respect to
the collinear optical path as the illumination angle increases. Dry
NP–film T-LSPR extinction spectra were obtained with samples
mounted in the slide holders within cuvettes containing no water.
Hydrated extinction spectra were acquired by simply adding water to
the cuvette containing the slide holder and NP–film sample.All NP–film sample spectra were normalized/baseline corrected
to that of a gold film containing a corresponding molecular spacer
layer and no immobilized NPs. This was done automatically by either
the WVASE32 or the spectrophometer software through the acquisition
of a blank spectrum. In the case of the spectrophotometer, the instrument
was operated in double beam mode with the reference cell empty during
sample acquisition.Plasmon resonance peak positions were calculated
by taking the
centroid of top (or bottom in the case of reflectivity spectra where
peaks are “dips”) 50% of the resonances observed in
either the transmitted extinction or the reflectivity curves. All
of the NP–film spectra presented here represent the signal
from millions of individual PNRs from each sample, and thus it was
not necessary to take multiple spectra for each sample, as one would
do if interrogating single PNRs on a sample one-by-one. The standard
deviation from the population of PNRs on each sample is represented
by the peak width of the NP–film LSPR and manifests itself
in the calculation of the peak position through our use of the peak
centroid to determine the peak position. For this reason, the data
plots that show the NP–film LSPR or PNR measurement versus
NP–film separation distance or pH do not have error bars.
pH-Induced Swelling Experiments
pH-induced swelling
experiments were done using four separate cuvettes and 60° slide
holders: two cuvettes for the sample and blank with pH 2 water and
two cuvettes for the sample and blank with pH 12 water. We did this
to minimize cross contamination during the pH cycling, as the 3D printed
slide holders are slightly porous. The low-pH solution was ultrapure
water with concentrated sulfuric acid added dropwise until the pH
of the solution reach 2.1, which we call “pH 2” throughout
the article. The high-pH solution was ultrapure water with concentrated
NaOH added dropwise until the pH of the solution reached 11.8, which
we call “pH 12” here. We measured refractive indices
(Bellingham & Stanley model 340 refractometer) of the pH 2 and
pH 12 solutions to be 1.332 76 and 1.332 73, respectively,
which indicates that there was a negligible difference in refractive
index between the solutions. Each pH swelling experiment was initiated
by placing an NP–film sample and corresponding blank sample
into the pH 2 cuvettes containing 3D printed slide holders and incubated
for 30 min, after which the extinction spectrum was obtained using
the spectrophotometer. Then the NP–film samples were removed
from the cuvettes containing the pH 2 solutions and added to the cuvettes
containing the pH 12 solutions. These cuvettes containing the samples
were then rinsed three times with fresh pH 12 solution before starting
the 30 min incubation at pH 12 and then acquiring the pH 12 sample
spectra. This process was repeated for each sample until eight pH
switches were obtained.
Authors: Nicola E Cant; Hao-Li Zhang; Kevin Critchley; Tetyana A Mykhalyk; Geoffrey R Davies; Stephen D Evans Journal: J Phys Chem B Date: 2003-12-11 Impact factor: 2.991
Authors: Fu Min Huang; Dean Wilding; Jonathon D Speed; Andrea E Russell; Philip N Bartlett; Jeremy J Baumberg Journal: Nano Lett Date: 2011-02-01 Impact factor: 11.189
Authors: Ryan T Hill; Jack J Mock; Angus Hucknall; Scott D Wolter; Nan M Jokerst; David R Smith; Ashutosh Chilkoti Journal: ACS Nano Date: 2012-09-21 Impact factor: 15.881
Authors: Jacob T Heggestad; Cassio M Fontes; Daniel Y Joh; Angus M Hucknall; Ashutosh Chilkoti Journal: Adv Mater Date: 2019-11-29 Impact factor: 30.849
Authors: Sander van den Driesche; Frieder Lucklum; Frank Bunge; Michael J Vellekoop Journal: Micromachines (Basel) Date: 2018-02-06 Impact factor: 2.891