Determination of scattering in intraocular lenses by spectrophotometric measurements.

José M Artigas1, Adelina Felipe1, Amparo Navea2, M Carmen García-Domene1, Álvaro Pons3, Jorge Mataix2.   

Abstract

This study presents a method for measuring scattering in explanted intraocular lenses (IOLs). Currently, determining scattering in IOLs is usually performed by Scheimpflug cameras and the results are expressed in the units used by this apparatus. The method we propose uses a spectrophotometer and this makes it possible to measure the total transmission of the IOL by using an integrating sphere; the direct transmission is determined by the double-beam mode. The difference between these two transmissions gives a value of the scattering in percentage values of light lost. In addition, by obtaining the spectral transmission curve, information about the most scattered wavelengths is also obtained. The IOL power introduces errors when directly measured, particularly with high powers. This problem can be overcome if a tailor-made cuvette is used that shortens the distance between the IOL and the condensing lens of the spectrophotometer when the IOL powers are below 24 diopters. We checked the effectiveness of this method by measuring the scattering of three explanted IOLs from cornea donors. This method, however, does not make it possible to ascertain whether the scattering measured is caused by surface light scattering or internal light scattering.

Mesh:

Year:  2014        PMID: 25539187     DOI: 10.1117/1.JBO.19.12.127006

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  1 in total

1.  Scanning Light Scattering Profiler (SLPS) Based Methodology to Quantitatively Evaluate Forward and Backward Light Scattering from Intraocular Lenses.

Authors:  Bennett N Walker; Robert H James; Don Calogero; Ilko K Ilev
Journal:  J Vis Exp       Date:  2017-06-06       Impact factor: 1.355

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.