Fangfang Wang1, Jianhua Tong2, Yang Li3, Chao Bian4, Jizhou Sun5, Shanhong Xia6.
Abstract
This work describes the fabrication, characterization, and application of a gold microband array electrode (MAE) for the determination of phosphate in fresh water samples. The working principle of this MAE is based on the reduction of a molybdophosphate complex using the linear sweep voltammetric (LSV) method. The calibration of this microsensor was performed with standard phosphate solutions prepared with KH2PO4 and pH adjusted to 1.0. The microsensor consists of a platinum counter electrode, a gold MAE as working electrode, and an Ag/AgCl electrode as reference electrode. The microelectrode chips were fabricated by the Micro Electro-Mechanical System (MEMS) technique. To improve the sensitivity, gold nanoparticles (AuNPs) were electrodeposited on the working electrode. With a linear range from 0.02 to 0.50 mg P/L, the sensitivity of the unmodified microsensor is 2.40 µA per (mg P/L) (R2 = 0.99) and that of the AuNPs-modified microsensor is 7.66 µA per (mg P/L) (R2 = 0.99). The experimental results showed that AuNPs-modified microelectrode had better sensitivity and a larger current response than the unmodified microelectrode.Entities:
Year: 2014 PMID: 25536001 PMCID: PMC4299121 DOI: 10.3390/s141224472
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1.The fabrication process and a photograph of a gold MAE (coin diameter 19 mm).
Figure 2.Cyclic voltammetric responses of molybdophosphate complex with an unmodified microelectrode.
Figure 3.SEM images of bare (A) and AuNPs-modified MAE (B).
Figure 4.Cyclic voltammetric response in 0.01 M sulfuric acid. a: AuNPs-modified microelectrode; b: unmodified microelectrode.
Figure 5.LSV responses of different concentrations of phosphate.
Figure 6.Calibration curve of phosphate standard solution measurement.
The current response sensitivities of different gold microsensor for phosphate detection.
| Unmodified gold electrode in this paper | 0.5 | 4.80 | 12.5 |
| AuNPs-modified gold electrode in this paper | 0.5 | 15.32 | 41.0 |
| Jońca [ | 7.065 | 0.0787 | 3.72 |
| Bai [ | 0.45 | 0.0471 | 20.0 |
| Bai [ | 0.45 | 0.0638 | 3.72 |
The measurement results of a real sample.
|
|
| |||
|---|---|---|---|---|
| 0.105 | 0.110 | 104.76 | 0.131 | 124.76 |
| 0.115 | 109.52 | 0.124 | 118.09 | |
| 0.114 | 108.57 | 0.111 | 105.71 | |
|
| ||||
| average | 0.113 | 107.62 | 0.122 | 116.19 |