Literature DB >> 25530080

Resection of calcified aortic heart leaflets in vitro by Q-switched 2 µm microsecond laser radiation.

I Rohde1, J-M Masch, D Theisen-Kunde, M Marczynski-Bühlow, R Bombien Quaden, G Lutter, R Brinkmann.   

Abstract

BACKGROUND: Transcatheter aortic valve implantation (TAVI) can result in paravalvular leakage and stent deformation in the presence of severe calcification. This study was undertaken to determine the efficacy of laser-assisted resection of calcific aortic valve leaflets as a method to minimize the effects of calcium on perivalvular leakage during TAVI.
METHODS: A Q-switched Tm:YAG laser emitting at a wavelength of 2.01 μm was used to evaluate the cutting efficiency on highly calcified human aortic leaflets in vitro (N = 10). A pulse energy of 4.3 mJ, a pulse duration of 0.8-1 μs, and a repetition rate of 1 kHz were used. The radiation was transmitted via a 200 µm core diameter quartz fiber. Resection was performed in a fiber-tissue contact mode on water-covered samples in a dish. The remnant particles were analyzed with respect to quantity and size by light microscopy.
RESULTS: A resection rate of 40.4 ± 22.2 mg/min on highly calcified aortic leaflets was achieved. This corresponds to a cutting speed of approximately 1 cm/min; a laser dissection time of 3 min per leaflet is expected. The majority of the remnant particles (85.4%) were <6 μm in diameter, with only 0.1% exceeding 300 μm.
CONCLUSIONS: The Q-switched Tm:YAG laser system showed promising results in cutting calcified aortic valves, by transmitting sufficient energy through a small flexible fiber. Catheter-based removal of aortic valve calcification may help to improve TAVI technology.
© 2014 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 25530080     DOI: 10.1111/jocs.12481

Source DB:  PubMed          Journal:  J Card Surg        ISSN: 0886-0440            Impact factor:   1.620


  2 in total

1.  Effect of laser pulse shaping on photoacoustic dosimetry in retinal models.

Authors:  Robert B Brown; Suzie Dufour; Pascal Deladurantaye; Nolwenn Le Bouch; Pascal Gallant; Sébastien Méthot; Patrick J Rochette; Ozzy Mermut
Journal:  Biomed Opt Express       Date:  2020-10-21       Impact factor: 3.732

2.  Transcatheter aortic valve resection: new mechanical devices.

Authors:  Huangdong Dai; Georg Lutter; Derk Frank; Sandra Freitag-Wolf; Ayça Topal; Assad Haneya; Janarthanan Sathananthan; Thomas Puehler
Journal:  J Thorac Dis       Date:  2020-11       Impact factor: 2.895

  2 in total