Soo Min Jang, Katie E Cardone, Thomas D Nolin, Darius L Mason, Darren W Grabe1.
Abstract
BACKGROUND: The purpose of this study was to evaluate the feasibility of utilizing an in-vitro, closed loop hemodialysis system as a method to assess drug clearance. Secondarily, this study tested the influence of variables (blood flow rate, dialysate flow rate, and type of filter) in the hemodialysis procedure on the clearance of vancomycin and gentamicin.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25526750 PMCID: PMC4289559 DOI: 10.1186/1471-2369-15-204
Source DB: PubMed Journal: BMC Nephrol ISSN: 1471-2369 Impact factor: 2.388
Dialysis conditions tested
| Conditions | Blood flow rate (mL/min) | Dialysate flow rate (mL/min) | Filter | Filter characteristics |
|---|---|---|---|---|
| C1 | 400 | 600 | Polyflux (Gambro) | Surface area 1.7 m2; Wall thickness 50 microns; Inner diameter 215 microns; Ultrafiltration coefficient 70* |
| C2 | 500 | 800 | Polyflux (Gambro) | |
| C3 | 400 | 600 | Optiflux F180 (Fresenius) | Surface area 1.8 m2; Wall thickness 40 microns; Inner diameter 200 microns; Ultrafiltration coefficient 60+ |
| C4 | 500 | 800 | Optiflux F180 (Fresenius) |
*Measured with bovine blood, Hematocrit = 32%, Protein 60 g/L, at 37°C.
+Measured with in vitro bovine, Hematocrit = 32%.
The concentration of vancomycin and gentamicin at different time interval in different conditions
| Condition | Filter | Sol | Time (min) | BFR (mL/min) | DFR (mL/min) | Vanco (ug/mL) | TG (ug/mL) |
|---|---|---|---|---|---|---|---|
| C1 | Poly | SBS | 0 | 400 | 600 | 20.04 ± 0.86 | 26.65 ± 1.69 |
| Poly | SBS | 180 | 400 | 600 | BLQ | 0.51 ± 0 | |
| Poly | Dial | 0 | 400 | 600 | BLQ | BLQ | |
| Poly | Dial | 180 | 400 | 600 | 1.35 ± 0.41 | 1.31 ± 0.18 | |
| C2 | Poly | SBS | 0 | 500 | 800 | 31.91 ± 2.5 | 35.93 ± 0.13 |
| Poly | SBS | 180 | 500 | 800 | BLQ | 0.53 ± 0.02 | |
| Poly | Dial | 0 | 500 | 800 | BLQ | BLQ | |
| Poly | Dial | 180 | 500 | 800 | 1.06 ± 0.05 | 0.99 ± 0 | |
| C3 | Opt* | SBS | 0 | 400* | 600* | 23.74 ± 5.27 | 25.41 ± 3.47 |
| Opt | SBS | 180 | 400 | 600 | BLQ | 0.52 ± 0.01 | |
| Opt | Dial | 0 | 400 | 600 | BLQ | BLQ | |
| Opt | Dial | 180 | 400 | 600 | 1.20 ± 1 | 1.14 ± 0.93 | |
| C4 | Opt | SBS | 0 | 500 | 800 | 30.75 ± 0.43 | 36.97 ± 0.18 |
| Opt | SBS | 180 | 500 | 800 | BLQ | 0.51 ± 0 | |
| Opt | Dial | 0 | 500 | 800 | BLQ | BLQ | |
| Opt | Dial | 180 | 500 | 800 | 0.91 ± 0.08 | 0.94 ± 0.15 |
UFR is constant at rate of 300 ml/hr except for the one condition. *UFR was 250 mL/hr. Sol: Solution; Poly: Polyflux filter; Opt: Optiflux filter; Dial: Dialysate; Vanco: Mean vancomycin concentration; TG: Mean total gentamicin concentration; BLQ: Below the lower limit of quantification.
Summary of the quantity of vancomycin and gentamicin using the mass balance equation and dialytic clearance
| Condition | Filter | Sol | BFR (mL/min) | DFR (mL/min) | Vanco (mg) | Gent (mg) | Cl Vanc(mL/min) | Cl Gent(mL/min) |
|---|---|---|---|---|---|---|---|---|
| C1 | Poly | Dial | 400 | 600 | 147.0 ± 44.7 | 142.7 ± 20 | 82.1 | 58.2 |
| C2 | Poly | Dial | 500 | 800 | 152.9 ± 7.2 | 142.7 ± 1.02 | 53.3 | 43.5 |
| C3 | Opt* | Dial | 400* | 600* | 131.0 ± 108.4* | 124.0 ± 101.5* | 65.3* | 56.6* |
| C4 | Opt | Dial | 500 | 800 | 131.9 ± 12.3 | 137.0 ± 21.5 | 47.7 | 40.6 |
UFR is constant at rate of 300 ml/hr except for the one condition. *UFR was 250 ml/hr. Sol: Solution; Poly: Polyflux filter; Opt: Optiflux filter; Dial: Dialysate; Vanco: Vancomycin concentration; Gent: Total Gentamicin concentration; BLQ: Below the lower limit of quantification; ClVanc: Mean Vancomycin Clearance; ClGent: Mean Gentamicin clearance.