Donald Schwartz1, John Samples2, Olga Korosteleva3. 1. Long Beach Eye Care Associates, 2650 Elm Avenue #108, Long Beach, CA 90806, USA ; USC Eye Institute, Los Angeles, CA, USA ; UC Irvine Gavin Herbert Eye Institute, Irvine, CA, USA. 2. Rocky Vista University, Parker, CO, USA. 3. Department of Mathematics and Statistics, California State University, Long Beach, CA, USA.
Abstract
BACKGROUND: This is a first-in-human study to determine the efficacy and tolerability of a new method of treating glaucoma using a low-power, low-frequency, focused therapeutic ultrasound for glaucoma (TUG) device designed to trigger an inflammatory reaction in the anterior chamber angle and trabecular meshwork to enhance outflow. The use of the device is anticipated for mild or moderate open-angle glaucoma as an enhancement to outflow. METHODS: In a two-branch clinical trial, a total of 26 primary open-angle glaucoma patients underwent a procedure consisting of the external application of the TUG device. In branch 1, nine of these patients were naïve to pharmaceutical treatment or had been off of medication for over 6 months. In branch 2, 17 patients were treated after a medication washout period. All patients in the study were followed for 12 months. RESULTS: In branch 1, there was a decrease in intraocular pressure averaging over 20% lasting at least a year in 74% of the eyes with non-normotensive open-angle glaucoma. In branch 2, an average of two visits while on medication provided the comparison intraocular pressure (IOP) to the effect of the TUG treatment after washout. It was seen that the intraocular pressure over the year post-treatment was equal to or better than the pharmaceutical control in close to 80% of measurements. CONCLUSION: A novel device for lowering intraocular pressure is described with a potential for adding to our armamentarium for treating glaucoma. This is a small cohort study which indicates beneficial trends. TRIAL REGISTRATION NUMBER: The study was a registered clinical trial, #ISRCTN50904302.
BACKGROUND: This is a first-in-human study to determine the efficacy and tolerability of a new method of treating glaucoma using a low-power, low-frequency, focused therapeutic ultrasound for glaucoma (TUG) device designed to trigger an inflammatory reaction in the anterior chamber angle and trabecular meshwork to enhance outflow. The use of the device is anticipated for mild or moderate open-angle glaucoma as an enhancement to outflow. METHODS: In a two-branch clinical trial, a total of 26 primary open-angle glaucomapatients underwent a procedure consisting of the external application of the TUG device. In branch 1, nine of these patients were naïve to pharmaceutical treatment or had been off of medication for over 6 months. In branch 2, 17 patients were treated after a medication washout period. All patients in the study were followed for 12 months. RESULTS: In branch 1, there was a decrease in intraocular pressure averaging over 20% lasting at least a year in 74% of the eyes with non-normotensive open-angle glaucoma. In branch 2, an average of two visits while on medication provided the comparison intraocular pressure (IOP) to the effect of the TUG treatment after washout. It was seen that the intraocular pressure over the year post-treatment was equal to or better than the pharmaceutical control in close to 80% of measurements. CONCLUSION: A novel device for lowering intraocular pressure is described with a potential for adding to our armamentarium for treating glaucoma. This is a small cohort study which indicates beneficial trends. TRIAL REGISTRATION NUMBER: The study was a registered clinical trial, #ISRCTN50904302.
Open-angle glaucoma is a worldwide problem for which newer, portable, low-cost, and
effective treatments are needed. Glaucoma affects approximately 3 million people in
the United States (source: preventblindness.org—Prevent Blindness America) and
70 million worldwide (source: glaucoma.org—Glaucoma Research Foundation). It
is expected that the increasing age of our population will significantly increase
the number of people with this blinding disease. The present methods used to treat
glaucoma have significant drawbacks. Pharmaceutical agents must depend on
compliance, often have side effects, and may interfere with other medications
required by the patient. The use of lasers such as argon or selective laser
trabeculoplasty offers a useful alternative to many of the problems inherent with
pharmaceutical agents. However, these instruments require a slit lamp biomicroscope
apparatus for viewing the trabecular meshwork and a contact lens system for
application of the energy for such treatment.Cataract surgery lowers intraocular pressure in patients with coexisting glaucoma [1-6]. The basis for this effect may be due to both anatomic and biochemical
changes in the area of the trabecular meshwork since observations have suggested
that the use of ultrasound in the eye may result in a decrease in pressure [7]. The association of the decrease in pressure with cataract surgery became
more evident when the use of phacoemulsification (ultrasound) became prevalent. The
finding that there is a decrease in the intraocular pressure (IOP) after
phacoemulsification has often been attributed to an increase in the opening of the
angle. This is certainly true with narrow-angle glaucoma, but recent studies have
revealed no correlation between the change in chamber depth and the IOP decrease
with open-angle glaucoma whereas the decrease is correlated with the pre-treatment
IOP [8-11]. In addition, reports of cataract surgery performed with an intracapsular
lens extraction did not indicate a decrease in IOP. Intracapsular cataract surgery
did not involve the use of ultrasound. Radius et al. reported that after the
intracapsular cataract surgery, there was a slight increase in the IOP [12].A prototype instrument hand piece (see Figure 1) was
designed and built to produce low-frequency ultrasound of 40 kHz. This
frequency is the same as that of a typical cataract surgery ultrasound. Although the
frequency is the same, the energy from the bubbles created by phacoemulsification
can create cavitation with a temperature of over 7,000°C; the therapeutic
ultrasound for glaucoma (TUG) treatment power of less than 2 W/cm2
only allows a temperature within the focal area to reach 45°C [13,14]. The instrument was developed for the external application of ultrasound
with the purpose of decreasing the intraocular pressure. We hypothesized that the
application of low-power and low-frequency focused ultrasound energy to the
trabecular meshwork could result in the lowering of intraocular pressure by
triggering a similar cytokine cascade to that triggered by SLT laser. To verify the
hypothesis, we designed a prospective study.
Figure 1
A prototype TUG instrument hand piece.
A prototype TUG instrument hand piece.
Methods
The author confirms that (1) the research followed the tenets of the Declaration of
Helsinki, (2) informed consent was obtained, (3) the research was approved by the
institutional review board to allow clinical studies (on the basis of previous
animal studies supporting a non-significant risk classification), and (4) power
analysis was performed to justify the number of patients enrolled in the study.
Device development
Prior to the use of the instrument, a power determination was performed with
in vitro and in vivo animal studies based on a temperature
elevation of approximately 6°C. This temperature elevation was just below
that which causes pain and cell necrosis [15-17]. It was found that with external application of the TUG device of
approximately 3 W of power/cm2 to the hand piece, the
temperature increased and then stabilized at approximately 43°C, when
measured by a “K type” microthermocouple positioned 0.5 mm
below the surface of a pig eye at the limbus using an Omega HH508 digital
thermometer. The instrument was developed and tested with this microthermocouple
to determine the power required to raise the temperature by 7° from
baseline. Testing was then repeated with pig eyes raised to a basal temperature
of 36.5°C by water bath to determine the power to raise the temperature to
43°C to 44°C and maintain this steady-state increased temperature.In vivo work on the animal model at power settings above
4 W/cm2 resulted in a large corneal inflammatory reaction.
Further in vivo studies therefore used only power settings below this
level. A point was then chosen on the sclera side with 0.5 mm of clearance
from the limbus for this study.
Study description
In early clinical work, the first two series TUG.1 and TUG.2 (not published) were
conducted to determine the tolerability (TUG.1) and to refine the parameters
(TUG.2) for treatment. The presently reported study (TUG.3) was a prospective
controlled study using the ultrasound instrument in the manner which optimized
the intraocular pressure-lowering effect.This study was funded by EyeSonix Incorporated and followed a protocol approved
by Western Institutional Review Board (WIRB). The study was conducted at one
site in the Long Beach Eye Care Associates offices. Health Insurance Portability
and Accountability Act (HIPPA)-compliant informed consent was approved by WIRB.
Each patient who participated in the study (or a designated family member) gave
written consent. The study was a registered clinical trial, #ISRCTN50904302.
Patient selection
Recruitment was conducted among glaucomapatients seen at the study site.The eligibility criteria were twofold:1. Patients with open-angle glaucoma and without medical or laser
treatment for at least 6 months, or2. Patients presently on pharmaceutical treatment for glaucoma.The exclusion criteria included both previous invasive glaucoma surgery and an
inability to comply with follow-up visits. Treatment-naïve patients were
offered the option of pharmaceutical agents or laser (always in this order) and
also the option of being part of the study.
Randomization and treatment protocol
Patients with symmetric IOP were randomized by a coin flip as to the eye selected
for the TUG treatment. If there was a significantly higher IOP in one eye, this
eye was selected for the TUG treatment.If patients were presently on pharmaceutical agents, a bilateral washout period
was performed. Prostaglandin analog medications were washed out for
1 month. Other agents were washed out for 1 week.The treatment was
performed after a measurement of the IOP was taken, and an ocular examination
was conducted to determine the baseline IOP. The baseline IOP was an average of
the two IOP measurements prior to the TUG treatment. In addition, an evaluation
was performed to characterize any baseline signs of inflammation. The following
steps were performed during the surgery illustrated in Figure 2. A drop of tetracaine was applied. The patient was then
placed in the supine position. A lid speculum was used for exposure. A marking
pen was then used on the sclera to mark four equally spaced quadrants at the
limbus. The eye was then covered with lidocaine 3.5% gel (Akorn). This served as
the necessary transmission gel, an added anesthetic. Additionally, it maintained
corneal moisture. A function generator (Tektronix AFG3101 Single Channel
Arbitrary/Function Generator, calibrated in October of 2008) was then tuned to
generate the proper power and frequency. This was fed by the single-channel
output to a power amplifier (E&J RF Power Amplifier, Model #2100 L,
Serial #1045). The “EyeSonix” hand piece of 40 kHz was then
attached. The treatment was performed once the correct power output with zero
reflectance was determined by tuning the frequency.
Figure 2
The investigator is performing a glaucoma procedure with the TUG
device.
The investigator is performing a glaucoma procedure with the TUG
device.A tip cover was then put over the hand piece. Next, the tip was placed
sequentially at the treatment sites. The application was at a point 0.5 mm
on the scleral side of the limbus. Applications were 45 s in duration at
each clock hour position for 12 clock hours. The four previously placed marks
allowed for an equal positioning of the spots of three per quadrant. Throughout
the procedure, the power output was monitored and adjusted to maintain the
correct delivery of energy to the hand piece. The intraocular pressure was
measured 1 h after the treatment as a safety precaution, since after laser
trabeculoplasty there is a small chance of a pressure spike after the treatment.
No significant post-treatment pressure elevations were seen. Post-treatment, the
patients received Nevanac (nepafenac ophthalmic suspension) 0.1% three times a
day for 1 day. This non-steroidal medication was used if there were
symptoms of inflammatory response: soreness, photophobia, or signs of
inflammation; the treatment continued until the symptoms and signs resolved.
Steroids, although more powerful, were avoided as the post-inflammatory response
was desired. The patients were seen 14 times post-surgically: on the following
day, 1 week later, and after that once a month for 12 months. The
measurements of IOP were not taken in 17 patients for a total of 47 missed
follow-up visits to the investigator’s office. Two patients were lost to
the follow-up visits after the 6-month appointment, and two dropped out after
being in the study for 10 months.Results of the treatment were tabulated at each follow-up visit into two general
categories. The first category was that of the intraocular pressure. The
pre-treatment (or baseline) IOP reading was an average of the two most recent
readings prior to the washout. These were Goldmann applanation measurements.
Post-treatment readings of both eyes were obtained by the technician before the
investigator entered the room. Post-treatment tonometry was performed in two
different methods. “Tonopen” (Tono-Pen XL Medtronic) measurements
were taken by well-trained technicians with at least four separate readings.
These readings were then averaged. After the tonopen measurements were recorded,
the technician performed Goldmann applanation tonometry. A second Goldmann
applanation tonometry was performed by the investigator. The two Goldman
readings were then averaged with the average of the tonopen readings; therefore,
the Goldman readings were weighted at 66% of the average for the single overall
reported average. The Goldmann tonometer was calibrated at least once a
week.The second category was of an analysis of the inflammatory reaction from the
treatment. This evaluation included the subjective symptoms of irritation,
discomfort, and pain and an objective slit lamp evaluation for signs of
conjunctival injection and signs of anterior chamber cells and flare. Each of
the parameters was graded on a 0 to 4 scale (minimal to maximal presence). The
subjective symptoms were elicited by staff technicians and verified by the
investigator and were always asked in the same manner. The slit lamp evaluation
was performed by the investigator. Each of the markers was characterized on a
scale of 0 to 4 with 0 being minimally present and 4 the worst possible. These
subjective responses were obtained by assistants and verified by the
investigator. The signs were observed at the same slit lamp at each visit. The
conjunctival injection was observed under low magnification in a low-light room
condition whereas cell and flare were evaluated with high power in a dark
room.
Results
Intraocular pressure
The IOP was analyzed separately in medication-free patients, in medication-free
non-normotensive glaucomapatients, and in those who were previously on
medication.
Medication-free patients
For the nine patients, the mean age was
73.3 ± 9.0 years (min = 62 years,
max = 89 years). Eight patients were Caucasian and one was
African-American. One patient was female and eight patients were male. Two
patients had primary open-angle glaucoma, four had normotensive glaucoma,
and one had pigmentary glaucoma. In Figure 3, the
graph of an average IOP decrease from the baseline for the treatment and
control groups is presented. In both groups, there is an overall downward
trend in average IOP, with the control group profile remaining above but
shadowing that of the treatment group.The average percent decrease in the
IOP from the baseline value is represented in Figure 4. The trend line from the time of treatment over the 12-month
period reveals a decrease in pressure in both the treated and the
contralateral eye which persists throughout the duration of follow-up. Note
that for these medication-free patients, the average percent IOP reduction
from the baseline is over 20% 1 year post-treatment in either eye.
Figure 3
Average IOP decrease (in mmHg) from the baseline in
medication-free patients in treated v. control eyes,
= 9.
Figure 4
Average percent IOP decrease from the baseline in medication-free
patients in treated v. control eyes,
= 9.
Average IOP decrease (in mmHg) from the baseline in
medication-free patients in treated v. control eyes,
= 9.Average percent IOP decrease from the baseline in medication-free
patients in treated v. control eyes,
= 9.
When reviewing the findings, it became apparent that the decrease in
intraocular pressure was less evident with those whose pre-treatment
pressure was below 19.5 mmHg. Thus, the analysis was redone without the
four normotensive glaucomapatients in order to obtain further perspective
in considering the treatment on patients with intraocular pressures of 20 or
greater.For the remaining five patients, the mean age was
71.8 ± 11.8 years (min = 62 years,
max = 89 years). All patients were Caucasian males. Four
patients had primary open-angle glaucoma and one had pigmentary glaucoma. Of
these five patients, one missed two follow-up visits, and thus, the data
were unavailable in the two cases.Figures 5 and
6 show, respectively, the average IOP decrease and
average percent IOP decrease from the baseline for this group of patients.
On both graphs, larger decreases in absolute units as well as in percent
intraocular pressure are observed in the medication-free patient group (cf.
Figures 3 and 4). The
average percent IOP decrease from the pre-treatment value approaches 25% as
opposed to 20% for all medication-free patients.
Figure 5
Average IOP decrease (in mmHg) from the baseline in
medication-free non-normotensive glaucoma patients in treated v.
control eyes, = 5.
Figure 6
Average percent IOP decrease from the baseline in medication-free
non-normotensive glaucoma patients in treated v. control eyes,
= 5.
Average IOP decrease (in mmHg) from the baseline in
medication-free non-normotensive glaucomapatients in treated v.
control eyes, = 5.Average percent IOP decrease from the baseline in medication-free
non-normotensive glaucomapatients in treated v. control eyes,
= 5.
Example of a medication-free non-normotensive glaucoma patient
Figure 7 presents an IOP profile for a
non-normotensive glaucoma study participant. This person had been a patient
in our practice for 3 years with normal pressures and cup disc ratios.
He was seen for routine examination in 2010 and found to have increased
optic nerve head cupping and an elevated IOP of 32 mmHg by Goldmann
applanation tonometry. He was then given options for treatment and chose the
TUG study. The IOP measurements were taken at pre-treatment, 1 day and
1 week post-surgery, and then every month for 1 year. His central
corneal thickness (CCT) measurements were 526 and 530 μm in the
treated and the control eye, respectively.
Figure 7
Individual profile for a medication-free non-normotensive
glaucoma patient: IOP measurements in treated v. control eyes
prior to the TUG treatment and during the study follow-up.
Individual profile for a medication-free non-normotensive
glaucomapatient: IOP measurements in treated v. control eyes
prior to the TUG treatment and during the study follow-up.Note that the control eye exhibits a similar behavior in terms of the IOP
measurements to the treated eye. Noteworthy is the length of this effect in
both eyes after only one treatment in one eye for this patient.
Medication washout patients
The mean age of the 17 patients was 73.1 ± 13.2 years
(min = 51 years, max = 92 years). Fourteen
patients were Caucasian, one African-American, and one Hispanic. Six females
and 11 males were treated. Eight patients had primary open-angle glaucoma,
seven had normotensive glaucoma, and two had pigmentary glaucoma. Twelve
patients missed a follow-up, but came back for subsequent visits. Two
patients dropped out after being in the study for 6 months, and two
dropped out after 10 months.In this group of 17 medication washout patients, three patients had a second
TUG treatment within 1 year. Such retreatments were performed if the
intraocular pressure approached the washout pressure. These retreatments
occurred at either the fifth or the sixth month post-initial TUG treatment.
Also in this group of patients, four of the 17 had a reintroduction of
medication. It should be noted that the washout of the medication was for
both eyes. One of the patients had a resumption of medication in the control
eye, but not in the treated eye. Of the group that went back on medication,
one patient used the medication for only 1 month and found the
medication again led to unacceptable irritation. She then was one of the
four who had a second TUG treatment.The baseline IOP value for these patients was an average of the IOP
measurements for two visits prior to the medication washout period. The
effect of the TUG treatment in the IOP reduction was compared to the IOP on
the pharmaceutical regimen.Figure 8 shows the
average change in IOP from the baseline after the initial TUG treatment.
Where the graph is above zero, the post-TUG treatment IOP was higher, below
zero—the medication IOP was higher. Note that the two IOP measurements
deviate very little from each other always remaining within
±2 mmHg.Figure 9 represents the
same results as in Figure 8 but in terms of the
average percent of change in IOP from the patient’s pharmaceutical
regimen. The positive values on the graph indicate that the post-TUG
treatment IOP exceeds that of the pharmaceutical regimen.To analyze the
number of patients who retained clinical control of pressure after washout,
a connected graph of percent of medication washout patients with IOP change
from baseline of at most 10% is constructed (see Figure 10).
Figure 8
Average IOP difference from pharmaceutical control in washout
patients in treated v. control eyes,
= 17.
Figure 9
Average percentage change in IOP from the pharmaceutical
regimen.
Figure 10
Percent of patients in the medication washout group
( = 17) whose IOP change
from the baseline value was within 10% or below the
pharmaceutical control: treated v. control eyes,
= 17.
Average IOP difference from pharmaceutical control in washout
patients in treated v. control eyes,
= 17.Average percentage change in IOP from the pharmaceutical
regimen.Percent of patients in the medication washout group
( = 17) whose IOP change
from the baseline value was within 10% or below the
pharmaceutical control: treated v. control eyes,
= 17.
Examples of medication washout patients
Figure 11 demonstrates an historical perspective
of IOP measurements in both eyes across the visits for a patient in the
medication washout group. This patient was among the 13 (74%) patients who
needed only a single TUG treatment to last 1 year. In the past, he had
been on latanoprost which controlled his IOP to approximately 16. He had a
washout of this medication. At the end of the 1-year period, he was off
medication in both eyes with an IOP around 16 mmHg. Even longer
follow-up shows his IOP to be maintained at a significant decrease from
baseline and even from pharmaceutical control.The next graph
(Figure 12) illustrates an IOP historical
perspective for a patient who was compelled to return to medication with
additive effect and later chose to have a TUG retreatment which showed an
enhanced effect from the second treatment over the original. This patient
was on latanoprost and then changed to Combigan as a result of side effects
of the latanoprost. Her IOP on medication was 21 mmHg. She had a 1-week
washout of the Combigan with a resultant IOP of 23 mmHg. Her TUG
treatment led to a decrease in IOP to 17 mmHg, but it gradually
increased to 23 mmHg over a 5-month period of time. She was offered the
option of returning to medication. She was reintroduced to medication with a
substantial decrease to 12.5 mmHg. But she again was bothered by the
side effects of Combigan and elected to have a second TUG treatment after
washout. The second TUG treatment was performed, and after 7 months
post-TUG #2 (1 year after TUG#1), the IOP was 15 mmHg without
medication.The first subject to have bilateral treatment is illustrated
(Figure 13). His pressure gradually elevated
over time after SLT laser. He elected to have TUG rather than a repeat SLT
or initiation of pharmaceutical agents. After coin flip, the first treated
was the right eye (OD). There was a profound effect on this treated eye. The
non-treated eye appeared to have a modest effect. After 7 months, the
non-treated eye returned to baseline. At this time, the left eye (OS) was
treated. The graph shows that for close to 3 years after the treatment
of the second eye, the intraocular pressures are significantly lower than
baseline in both eyes.
Figure 11
Historical perspective for a medication washout patient: IOP
measurements in both eyes prior to the TUG treatment and during
the study follow-up.
Figure 12
Historical perspective for a medication washout patient who
resumed medication and underwent a second TUG treatment: IOP
measurements in both eyes prior to the first TUG treatment and
during the study follow-up.
Figure 13
The first subject to undergo the TUG treatment in each eye.
The second eye was treated 7 months after the first eye when
the intraocular pressure in that eye had increased to the original
baseline.
Historical perspective for a medication washout patient: IOP
measurements in both eyes prior to the TUG treatment and during
the study follow-up.Historical perspective for a medication washout patient who
resumed medication and underwent a second TUG treatment: IOP
measurements in both eyes prior to the first TUG treatment and
during the study follow-up.The first subject to undergo the TUG treatment in each eye.
The second eye was treated 7 months after the first eye when
the intraocular pressure in that eye had increased to the original
baseline.
Tolerability of the procedure
The intraocular pressure was measured at a time approximately 2 h after the
treatment to look for any possible pressure spikes. No pressure spikes were
found in the patients treated.The tolerability of the treatment with the TUG device was evaluated at each
visit. Patients were queried about three markers of patient symptoms:
irritation, discomfort, and pain. Similarly, there were three findings which we
associated with signs of inflammatory response: conjunctival injection, anterior
chamber cells, and flare.The values of symptoms scores were as follows:1. Irritation scores ranged between 1 and 2.5 in three
patients at the 1-day visit. One week after the treatment, irritation was
gone.2. Discomfort was present in four patients at the 1-day
visit. One had a score of 2, and three were at 1. At 1 week post-treatment,
one patient was at 1, one at 1.5, and the rest were at 0. Afterwards, this
symptom has disappeared.3. Pain was experienced by one patient at the level of 2 at
the 1-day visit. Besides this instance, pain was absent for each and every
patient at each and every visit.The values of signs scores are listed below:1. Injection of the conjunctiva was a frequent but
short-lived finding. At the first visit, 1 day after the treatment, it was
found in 24 of the 26 patients. One had a score of 3, one had a score of 2.5,
and seven were at 2. There were five at 1.5, seven at 1, and the other three
scored 0.5. By the 1-week visit, only eight had any such finding: two at 0.5 and
six at 1. This sign was not apparent afterwards.2. Cells in the aqueous humor were not observed in any of the
patients at any visit.3. Flare was observed in the treated eye of nine patients on
the first day. In five patients, the score was 0.5, and the other four had a
score of 1. At the 1-week visit, a flare score of 0.5 was measured in a single
patient. In the other patients, this sign was absent.In summary, there is a frequent finding of conjunctival injection with a
subjective feeling of irritation that fades over several days. No symptom or
sign reached the level of 3 at any time. The more serious symptom of pain and
sign of cells in the aqueous humor were remarkable for their absence at each and
every visit including the first day (with one exception of pain on day 1).The
average symptoms and signs scores are shown in Figures 14 and 15, respectively. As can be seen on
the graphs, the minimal findings have dissipated by 1 month. It is doubtful
that new signs or symptoms will appear in the longer follow-up.
Figure 14
Average symptoms scores (irritation, discomfort, and pain) in TUG
patients over the duration of the study.
Figure 15
Average signs scores (injection, cells, and flare) in TUG patients
over the duration of the study.
Average symptoms scores (irritation, discomfort, and pain) in TUGpatients over the duration of the study.Average signs scores (injection, cells, and flare) in TUGpatients
over the duration of the study.From the graphs, the patients tolerated the procedure quite comfortably with the
only common comment of mild irritation which was consistent with a mild to
moderate conjunctival injection. These typically resolved in a few days in
almost all cases without treatment. If the patient had these findings, Nevanac
(nepafenac ophthalmic suspension) was offered. Most patients declined to use the
medication and stated that the discomfort did not warrant any pharmaceutical
treatment.
Discussion
The armamentarium of treatment modalities for glaucoma is increasing. At the present
time, pharmaceutical agents are typically the first line of treatment with the use
of laser seen as an adjunct method. A new method of decreasing intraocular pressure
is described. This is the initial study to be reported using a low-frequency,
low-power, focused ultrasound for the treatment of glaucoma.The mechanism of action has yet to be described fully, but the design of the
ultrasound was with the purpose of creating a focal area of hyperthermia within the
anterior chamber angle. It was felt that such an effect could trigger inflammatory
cytokines analogous to the effect of SLT laser. The finding of a concurrent decrease
in the intraocular pressure in the contralateral eye may support this mechanism of
action.We hypothesized that ultrasound energy applied externally near the limbus may have at
least three modes of action all of which could trigger a decrease in the intraocular
pressure. None of these potential modes of action are mutually exclusive, and other
modes are certainly possible.Firstly, there may be a sonomechanical, or vibratory, effect transmitted to the
trabecular meshwork, loosening debris and flushing out blockages [18-24]. Work in Sweden by Björn Svedbergh resulted in a patented device
with the expressed concept of shaking debris to loosen trabecular meshwork blockage
(Ultrasound Probe—US Patent 6162193, filing date Sep 15, 1997, issue date Dec
19, 2000). This device utilized a fluid-filled chamber with a membrane for applying
non-focused ultrasound transmission to the external eye.A second mechanism, that of a localized hyperthermia, may trigger heat shock proteins
and potentially beneficial cytokines. This would hypothetically be similar to a
cytokine response evoked by laser trabeculoplasty [25-33]. Specific cytokines have been demonstrated to lower the IOP after both
argon and selective laser trabeculoplasty and have also been found to be triggered
by ultrasound of a frequency similar to that used in phacoemulsification [34].A third mode of effect may be an induction of cytokines through integrins [35,36]. These receptors absorb ultrasound energy and in turn induce cytokine
activity which may be beneficial in lowering the IOP.Ultrasound energy applied externally to the eye has a significant advantage compared
with many other treatment modalities in that it can be applied in a non-invasive
manner. Previous treatments to use ultrasound to treat glaucoma have been directed
more posteriorly [37-42]. Coleman et al. revealed that high-intensity focused ultrasound (HIFU)
led to an ablation of the ciliary body and a subsequent thinning of the sclera. The
energy was directed from the outside of the eye towards the ciliary body with an
attempt to ablate the tissue and thereby decrease aqueous production. This
high-powered ultrasound had an additional effect in leading to a thinning of the
sclera overlying the ciliary body. The use of this ultrasound model was for
intransigent glaucoma, and its use fell from favor with the advent of other superior
treatment methods. Recently, another group (EyeTechCare of Lyon, France) has
reported an ultrasound device, “EyeOP1”, used to coagulate the ciliary
epithelium. This device uses a single-treatment strategy with a circular tip with a
multiple port array. These ports focus high-intensity ultrasound into the ciliary
body to decrease aqueous inflow in patients with refractory glaucoma [17].In a single-center pilot study of a novel glaucoma treatment technique (TUG), we
found a significant decrease in intraocular pressure. The effect on intraocular
pressure in the “naïve” group was significant with an average
decrease in the total group of almost 20% and almost 25% when not including the
normotensive glaucomapatients. The effect of the treatment appeared to be more
significant in the higher baseline intraocular pressures. In over 74%, the effect
lasted for 1 year. A second treatment was offered if the IOP rose to baseline.
Subsequent TUG treatments typically lasted far longer than the first treatment
(beyond the 1 year reported in this study).In those subjects washed out of their pharmaceutical glaucoma medication, the
intraocular pressure measurements after the TUG treatment were equal (within 10%) to
or even better (70% of the visits) than their pharmaceutical control for the year of
the study.The subjects tolerated the procedure well with only minimal discomfort noted on the
first day post-treatment. There was typically a symptom of only a mild feeling of
irritation and a slit lamp finding of mild to moderate conjunctival hyperemia.A method to treat glaucoma which reduces the issues of pharmaceutical compliance,
allergy, and side effects and has the potential for portability could be a
significant contribution. It has even more potential for areas of the third world
where glaucoma is more prevalent and treatment with medication and/or surgery is
unavailable. An updated prototype weighing less than 5 lbs with the same
characteristics as that used for this study has now been produced and is in early
multicenter clinical trials.
Conclusion
This is a report of a first-in-human trial of a low-power, low-frequency ultrasound
instrument to treat open-angle glaucoma. The findings are supportive of a
well-tolerated procedure with a significant decrease in intraocular pressure.
Further studies are needed for validation.
Abbreviations
IOP: intraocular pressure; OD: right eye; OS: left eye; SLT: laser selective laser
therapy; TUG: therapeutic ultrasound for glaucoma.
Competing interests
Dr. Schwartz is the Founder and President of EyeSonix, the company that has developed
the TUG treatment. The other authors declare that they have no competing
interests.
Authors’ contributions
DS was responsible for the design of the study, the clinical work necessary, and the
initial review and analysis of the data. JS was responsible for the collaboration of
the design of the study and review of the manuscript. OK was responsible for the
review of the data and substantial help with the analysis of data results. All
authors read and approved the final manuscript.