LC-gradient used for the separation of the plasma samples
| Time [min] | Eluent A [%] | Eluent B [%] |
|---|---|---|
| 0 | 100 | 0 |
| 0.1 | 80 | 20 |
| 8 | 75 | 25 |
| 10 | 0 | 100 |
| 11 | 0 | 100 |
| 11.1 | 100 | 0 |
| 25 | 100 | 0 |
Figure 1Investigation of matrix effects by the post column infusion method defined by Bonfiglio et al. A: Scheme of the post-column infusion system B: Total ion chromatogram C: Overlay chromatograms showing the matrix effects due to purified plasma on the detection of angiotensin II.
Figure 2Overview about sample preparation for the quantification of angiotensin II.
Figure 3Detection of native angiotensin II by ESI-MS A: positive full scan spectrum B: positive MS/MS spectrum of the parent ion 523.8 m/z. The b6 fragment ion (784.4 m/z) was used for the quantification.
Figure 4Detection of stable isotope labeled angiotensin II by ESI-MS A: positive full scan spectrum B: positive MS/MS spectrum of the parent ion 528.8 m/z. The b6 fragment ion (794.4 m/z) was used for the quantification.
Figure 5Linear calibration curve for stable isotope labeled angiotensin II prepared in plasma (3–120 fmol on column; N = 3).
Accuracy and precision of quality control samples
| Intra-day (n = 3) | Inter-day (n = 9) | |||||
|---|---|---|---|---|---|---|
| QC | Concentration (nM) | Accuracy (%)1 | Precision (%)2 | Accuracy (%)1 | Precision (%)2 | |
|
| LLOQ | 6 | 106.4 | 16.0 | 112.5 | 14.5 |
| LQC | 30 | 96.6 | 6.6 | 91.5 | 8.1 | |
| MQC | 120 | 105.3 | 7.6 | 103.1 | 4.6 | |
| HQC | 240 | 92.7 | 7.8 | 94.6 | 9.0 | |
1calculated as (mean determined amount/nominal amount × 100).
2calculated as % CV. (SD/mean) × 100.
Figure 6Analysis of purified plasma sample from healthy volunteer by ESI-MS. A: extracted ion chromatogram of the ion 784.4 ± 0.5 m/z corresponding to the native angiotensin II and a characteristic fragment spectrum with labeling of the b6-fragment (grey shaded area) B: extracted ion chromatogram of the ion 794.4 ± 0.5 m/z corresponding to the stable isotope labeled angiotensin II and a characteristic fragment spectrum with labeling of the b6-fragment (grey shaded area). Confirmation of the identity of the native ngiotensin II was performed by using both: retention time and fragment ion of co-eluting stable isotope ngiote ngiotensin II.
Endogenous ngiotensin II concentrations in 9 individual healthy subjects and CKD-5D patients
| Endogenous angiotensin II concentrations [pM] in individual subjects | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Group | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | mean |
| Healthy subjects | 30.2 | 9.7 | 19.2 | 9.5 | 10.2 | 33.5 | 24.9 | 22.7 | 6.6 | 18.5 |
| CKD-5D patients | 16.8 | 10.4 | 13.4 | 205.4 | 260.4 | 8.4 | 14.3 | 6.3 | 45.4 | 64.5 |