Reproducibility of first- and second-trimester uterine artery pulsatility index measured by transvaginal and transabdominal ultrasound.

A E G M T Ferreira1,2, F Mauad Filho1,2, P S G Abreu2, F M Mauad2,3, E Araujo Júnior4, W P Martins1.   

Abstract

OBJECTIVES: To examine and compare the reproducibility of measurement of first- and second-trimester uterine artery pulsatility index (UtA-PI) using transabdominal (TAS) and transvaginal (TVS) ultrasound.
METHODS: This was an observational study including women ≥ 18 years of age, with a singleton pregnancy, in the first trimester (between 11 + 0 and 13 + 6 weeks' gestation) or second trimester (between 20 and 26 weeks' gestation). UtA-PI and angle of insonation were assessed by two observers (one with 15 and the other with 3 years of experience) using both TAS and TVS. The more experienced observer performed two scans alternated by the other observer. The acquisitions were completely independent and the observers were blinded to each other and to their own measurements. Reproducibility of the measurements by TVS and TAS was assessed using the concordance correlation coefficients (CCCs), intraclass correlation coefficients (ICCs) and limits of agreement (LoA).
RESULTS: We analyzed data from 97 women in the first trimester and 96 in the second trimester. The mean ± SD UtA-PI was significantly higher when measured using TVS, compared with TAS, in both the first (1.60 ± 0.49 vs 1.52 ± 0.63, respectively; P = 0.03) and second (1.07 ± 0.33 vs 0.96 ± 0.32, respectively; P < 0.001) trimesters. The median angle of insonation was significantly lower when using TVS, compared with TAS, in both the first (8.0° (interquartile range (IQR), 2.5-16.3°) vs 12.5° (IQR, 2.5-20.0°), respectively; P = 0.04) and second (10.0° (IQR, 4.5-16.5°) vs 17.5° (IQR, 5.0-27.9°), respectively; P < 0.001) trimesters. Both ultrasound techniques had similar reproducibility: the intraobserver CCC ranged from 0.93 to 0.95 and the interobserver CCC ranged from 0.81 to 0.86; and the ICCs of both techniques were highly comparable (the intraobserver LoA was approximately ± 20-30% and the interobserver LoA was approximately ± 30-40%).
CONCLUSIONS: When measuring UtA-PI, assessment by TVS provides higher values and better insonation angle compared with TAS. The reproducibility of the ultrasound methods in both first and second trimesters of pregnancy was comparable and should not be considered as good. Future studies examining technical improvements with the aim of increasing the reproducibility of this technique should be encouraged.
Copyright © 2014 ISUOG. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  Doppler ultrasonography; pre-eclampsia; reproducibility of results; uterine artery

Mesh:

Year:  2015        PMID: 25504919     DOI: 10.1002/uog.14762

Source DB:  PubMed          Journal:  Ultrasound Obstet Gynecol        ISSN: 0960-7692            Impact factor:   7.299


  2 in total

1.  Reference range for uterine artery Doppler pulsatility index using transvaginal ultrasound at 20-24w6d of gestation in a low-risk Brazilian population.

Authors:  Alberto Borges Peixoto; Taciana Mara Rodrigues Da Cunha Caldas; Gabriele Tonni; Priscilla De Almeida Morelli; Larissa D'amico Santos; Wellington P Martins; E Araujo Júnior
Journal:  J Turk Ger Gynecol Assoc       Date:  2016-01-12

2.  Uterine Artery Doppler Flow Indices in Pregnant Women During the 11 Weeks + 0 Days and 13 Weeks + 6 Days Gestational Ages: a Study of 168 Patients.

Authors:  Voicu Dascau; Gheorghe Furau; Cristian Furau; Cristina Onel; Casiana Stanescu; Liliana Tataru; Cristina Ghib-Para; Cristina Popescu; Luminita Pilat; Maria Puschita
Journal:  Maedica (Bucur)       Date:  2017-01
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.