Impact of silicone and metal port-a-cath implants on superficial hyperthermia treatment quality.

Citlalli Jessica Trujillo-Romero1, Margarethus Marius Paulides, Tomas Drizdal, Gerard C van Rhoon.   

Abstract

PURPOSE: A port-a-cath is a device implanted under the skin for continuous drug administration. It is composed of a catheter and a silicone or metal reservoir. A simulation study was done to assess the impact of a port-a-cath implant on the quality of superficial hyperthermia treatments applied using the Lucite cone applicator (LCA).
METHODS: Specific absorption rate (SAR) and temperature distributions were predicted using SEMCAD-X (version 14.8). We simulated 72 arrangements: two LCA-implant set-ups (central port-a-cath or at an edge below the LCA footprint), six translations of the LCA per set-up, two LCA orientations (Parallel or perpendicular electric field direction) per set-up, two implant materials (silicon or metal) and a control without port-a-cath. Treatment quality was quantified by the average 1 g SAR coverage (CV25%), i.e. volume within the 25% iso-SAR surface, and the volume within the 40 °C iso-temperature surface (CV40 °C).
RESULTS: CV25% reduced with a silicon port-a-cath located below the LCA footprint. In the worst scenario, only 64% of the CV25% of the control set-up was achieved. For a metal port-a-cath below the LCA aperture, dramatic reductions of CV25% were predicted: worst scenario down to 12.1% of the control CV25%. For the CV40 °C the worst case values were 74.5% and 6.5%, for silicon and metal implants, respectively.
CONCLUSIONS: A silicone port-a-cath below the LCA had a smaller effect on treatment quality than a metal implant. Based on this study we recommend verifying heating quality by 3D patient-specific treatment planning when a port-a-cath is located below the footprint of the applicator.

Entities:  

Keywords:  Implants; port-a-cath; superficial hyperthermia; thermal modelling; treatment planning

Mesh:

Substances:

Year:  2014        PMID: 25495267     DOI: 10.3109/02656736.2014.985748

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  3 in total

Review 1.  Heating technology for malignant tumors: a review.

Authors:  H Petra Kok; Erik N K Cressman; Wim Ceelen; Christopher L Brace; Robert Ivkov; Holger Grüll; Gail Ter Haar; Peter Wust; Johannes Crezee
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

2.  Modelling Curved Contact Flexible Microstrip Applicators for Patient-Specific Superficial Hyperthermia Treatment Planning.

Authors:  H Petra Kok; Jort Groen; Akke Bakker; Johannes Crezee
Journal:  Cancers (Basel)       Date:  2020-03-11       Impact factor: 6.639

Review 3.  Current state of the art of regional hyperthermia treatment planning: a review.

Authors:  H P Kok; P Wust; P R Stauffer; F Bardati; G C van Rhoon; J Crezee
Journal:  Radiat Oncol       Date:  2015-09-17       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.