In Vitro Comparison of Aerosol Delivery Using Different Face Masks and Flow Rates With a High-Flow Humidity System.

Hui-Ling Lin1, Robert J Harwood2, James B Fink3, Lynda T Goodfellow2, Arzu Ari2.   

Abstract

BACKGROUND: Aerosol drug delivery to infants and small children is influenced by many factors, such as types of interface, gas flows, and the designs of face masks. The purpose of this in vitro study was to evaluate aerosol delivery during administration of gas flows across the range used clinically with high-flow humidity systems using 2 aerosol masks.
METHODS: A spontaneous lung model was used to simulate an infant/young toddler up to 2 y of age and pediatric breathing patterns. Nebulized salbutamol by a vibrating mesh nebulizer positioned at the inlet of a high-flow humidification system at gas flows of 3, 6, and 12 L/min was delivered via pediatric face masks to a pediatric face mannequin attached to a filter. Aerosol particle size distribution exiting the vibrating mesh nebulizer and at the mask position distal to the heated humidifier with 3 flows was measured with a cascade impactor. Eluted drug from the filters and the impactor was analyzed with a spectrophotometer (n = 3). Statistical analysis was performed by analysis of variance with a significant level of P < .05.
RESULTS: The inhaled mass was between 2.8% and 8.1% among all settings and was significantly lower at 12 L/min (P = .004) in the pediatric model. Drug delivery with pediatric breathing was greater than with infant breathing (P = .004). The particle size distribution of aerosol emitted from the nebulizer was larger than the heated humidified aerosol exiting the tubing (P = .002), with no difference between the 3 flows (P = .10).
CONCLUSIONS: The flows of gas entering the mask and breathing patterns influence aerosol delivery, independent of the face mask used. Aerosol delivery through a high-flow humidification system via mask could be effective with both infant and pediatric breathing patterns.
Copyright © 2015 by Daedalus Enterprises.

Entities:  

Keywords:  aerosol delivery; aerosol mask; gas flows; high-flow humidification; particle size distribution; pediatric; vibrating mesh nebulizer

Mesh:

Substances:

Year:  2014        PMID: 25492959     DOI: 10.4187/respcare.03595

Source DB:  PubMed          Journal:  Respir Care        ISSN: 0020-1324            Impact factor:   2.258


  5 in total

1.  Development of an Inline Dry Powder Inhaler for Oral or Trans-Nasal Aerosol Administration to Children.

Authors:  Dale Farkas; Michael Hindle; Serena Bonasera; Karl Bass; Worth Longest
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2019-08-29       Impact factor: 2.849

2.  High-Efficiency Nose-to-Lung Aerosol Delivery in an Infant: Development of a Validated Computational Fluid Dynamics Method.

Authors:  Karl Bass; Susan Boc; Michael Hindle; Kelley Dodson; Worth Longest
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2018-12-15       Impact factor: 2.849

Review 3.  Creation of a protective pulmonary bioshield against inhaled organophosphates using an aerosolized bioscavenger.

Authors:  Yvonne J Rosenberg; James B Fink
Journal:  Ann N Y Acad Sci       Date:  2016-07-02       Impact factor: 5.691

4.  High-Efficiency Dry Powder Aerosol Delivery to Children: Review and Application of New Technologies.

Authors:  Karl Bass; Dale Farkas; Amr Hassan; Serena Bonasera; Michael Hindle; P Worth Longest
Journal:  J Aerosol Sci       Date:  2020-10-14       Impact factor: 3.433

5.  Initial Development of an Air-Jet Dry Powder Inhaler for Rapid Delivery of Pharmaceutical Aerosols to Infants.

Authors:  Connor Howe; Michael Hindle; Serena Bonasera; Vijaya Rani; P Worth Longest
Journal:  J Aerosol Med Pulm Drug Deliv       Date:  2020-08-04       Impact factor: 2.849

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.