Percutaneous device closure of patent ductus arteriosus with pulmonary artery hypertension: long-term results.

Ishwarappa Balekundri Vijayalakshmi1, Natraj Setty, Chitra Narasimhan, Vivek Singla, Cholenahalli Nanjappa Manjunath.   

Abstract

UNLABELLED: Device closure of patent ductus arteriosus (PDA) is treatment of choice. But device closure in presence of pulmonary artery hypertension (PAH) remains a challenge. Data on patient selection, technical considerations, and complications are limited. AIM: To know the challenges and efficacy of device closure of PDA with PAH. MATERIALS AND
RESULTS: Out of 1,325 cases of device closure of PDA, 246 (18.6%) with PAH formed the study material. To test the feasibility, chosen device is used to occlude PDA for ten minutes without oxygen inhalation. The device is released only if PAH reduced. PAH decreased in all except in 1 patient after closure with muscular ventricular septal occluder (MVSDO), pulmonary artery pressure (PAP) transiently increased (became supra-systemic), without significant reduction in aortic pressure. Device embolized in 8 patients (3.3%). Percutaneous retrieval was done in 4 (by snare in 2 and by fixing the cable to device in 2) and replaced with bigger devices. The surgical removal of the embolized MVSDO and ligation was done in 4 cases. All patients were on oral sildenafil and bosentan until PAP regressed to normal. Follow up was from 6 months to 9 years. No residual shunt in any patient on follow-up. The PAP regressed to normal in all except 5 cases (2.03%) of Down's syndrome with systemic PAP.
CONCLUSIONS: Device closure of PDA with PAH is feasible, safe in all age groups. Temporary PDA occlusion with device is effective and time saving for evaluating pulmonary vascular reactivity. Device embolization in aorta is higher with severe PAH. Novel method of retrieval is effective.
© 2014, Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2014        PMID: 25490863     DOI: 10.1111/joic.12156

Source DB:  PubMed          Journal:  J Interv Cardiol        ISSN: 0896-4327            Impact factor:   2.279


  3 in total

1.  Safety and Outcomes of Transcatheter Closure of Patent Ductus Arteriosus in Children With Pulmonary Artery Hypertension.

Authors:  Arash Salavitabar; Usha S Krishnan; Mariel E Turner; Julie A Vincent; Alejandro J Torres; Matthew A Crystal
Journal:  Tex Heart Inst J       Date:  2020-08-01

2.  A novel method of creation of a fenestration in nitinol occluder devices used in closure of hypertensive patent arterial ducts.

Authors:  Anil Kumar Singhi; Kothandam Sivakumar
Journal:  Ann Pediatr Cardiol       Date:  2016 Jan-Apr

3.  Outcomes of transcatheter closure of patent ductus arteriosus with the off-label use of large occluders (≥16 mm).

Authors:  Kewal Kanabar; Dinakar Bootla; Navjyot Kaur; C R Pruthvi; Darshan Krishnappa; Krishna Santosh; Vivek Guleria; Manoj Kumar Rohit
Journal:  Indian Heart J       Date:  2020-04-07