| Literature DB >> 25489199 |
E O Odari1, N L M Budambula2, H Nitschko3.
Abstract
BACKGROUND: Development of "combination" assays detecting in parallel, within a single test, Hepatitis C Virus (HCV) antigens and antibodies, not only reduces the window period in HCV-infection but also costs. Reduction of costs is important for developing countries where money and personal resources are limited.Entities:
Keywords: Ag-Ab Combination assay; ELISA; Hepatitis C Virus; Monolisa HCV Ag-Ab Ultra
Mesh:
Substances:
Year: 2014 PMID: 25489199 PMCID: PMC4248034 DOI: 10.4314/ejhs.v24i4.10
Source DB: PubMed Journal: Ethiop J Health Sci ISSN: 1029-1857
Suspected Hepatitis C samples tested with AXSYM and Monolisa Ag/Ab ELISA kits (n=73)
| Test samples | AXSYM | ||||
| Sample type | Total | Reactive | Non-Reactive | Reactive | Non- Reactive |
| PCR positive samples | 31 | 3(100%) | 0 | 31(100%) | 0 |
| Antibody positive- PCR negative | 9 | 9(100%) | 0 | 8(88.9%) | 1(11.1%) |
| HCV negative samples | 26 | 1(3.8%) | 25(96.2) | 0 | 26(100%) |
| Bonemarrow transplant patient | 7 | 4(57.1%) | 3(42.9%) | 4(57.1%) | 3(42.9%) |
Sensitivity, specificity and predictive values of Monolisa Ag/Ab and AxSYM HCV version 3.0 ELISA assays
| ELISA assay | 95% confidence interval | |||
| Sensitivity | Specificity | PPV % | NPV % | |
| Overall Analysis | 91.5(43/47) | 100(26/26) | 100 | 86.7 |
| (79.6 – 97.6) | (86.8 –100) | (91.8 – 100) | (69.3 – 96.2) | |
| Partial analysis | 97.5(39/40) | 100 | 100 | 96.3 |
| (86.8 – 99.9) | (86.8 – 100) | (91–100) | (81 – 99.9) | |
| AxSYM HCV version 3.0 | ||||
| Overall Analysis | 93.6 | 96.2 | 97.8 | 89.3 |
| (82.5 – 98.7) | (80.4 – 99.9) | (88.2 –99.9) | (71.8 –99.7) | |
| Partial analysis | 100(40/40) | 96.1 | 97.6 | 100 |
| (91.2 – 100) | (80.4 –99.9) | (87.1 – 99.9) | (86.3 –100) | |
7 samples in the BMT sample panel excluded from the analysis
Figure 1shows no correlation (R2 (32) = 0.097) between viral load and optical density signals for Monolisa Ag-Ab Ultra assay. One sample of approximately 10 × 106 IU/mL was not detected (red arrow) since OD fell below the cut off value, whereas one sample below 1 × 105 IU/mL (blue arrow) was detected.
Figure 2shows no correlation (R2 (32) = 0.082) between viral load and optical density signals for AxSYM, antibody ELISA. One sample of approximately 10 × 106 IU/mL (circled in a red ring) was not detected since the OD was below the cut off value (s/co = 1.0), whereas on sample below 1 × 105 IU/mL was detected.
Figure 3shows results of ELISA tests from plasma of a BMT patient obtained pre and post bone marrow transplant. All samples obtained before bone marrow transplant were positively detected (Figure 3a and 3b), whereas all samples collected more than 200 days after transplantation were not detected by the 2 assays.