MRI compatible head phantom for ultrasound surgery.

Georgios Menikou1, Tetiana Dadakova2, Matt Pavlina2, Michael Bock2, Christakis Damianou3.   

Abstract

OBJECTIVE: Develop a magnetic resonance imaging (MRI) compatible head phantom with acoustic attenuation closely matched to the human attenuation, and suitable for testing focused ultrasound surgery protocols.
MATERIALS AND METHODS: Images from an adult brain CT scan were used to segment the skull bone from adjacent cerebral tissue. The segmented model was manufactured in a 3-D printer using (Acrylonitrile Butadiene Styrene) ABS plastic. The cerebral tissue was mimicked by an agar-evaporated milk-silica gel (2% w/v-25% v/v-1.2% w/v) which was molded inside a skull model.
RESULTS: The measured attenuation of the ABS skull was 16 dB/cm MHz. The estimated attenuation coefficient of the gel replicating brain tissue was 0.6 dB/cm MHz. The estimated agar-silica gel's T1 and T2 relaxation times in a 1.5 Tesla magnetic field were 852 ms and 66 ms respectively. The effectiveness of the skull to reduce ultrasonic heating was demonstrated using MRI thermometry.
CONCLUSION: Due to growing interest in using MRI guided focused ultrasound (MRgFUS) for treating brain cancer and its application in sonothrombolysis, the proposed head phantom can be utilized as a very useful tool for evaluating ultrasonic protocols, thus minimizing the need for animal models and cadavers.
Copyright © 2014 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Brain; MRI; Ultrasound

Mesh:

Year:  2014        PMID: 25482534     DOI: 10.1016/j.ultras.2014.11.004

Source DB:  PubMed          Journal:  Ultrasonics        ISSN: 0041-624X            Impact factor:   2.890


  8 in total

Review 1.  Surgical applications of three-dimensional printing: a review of the current literature & how to get started.

Authors:  Don Hoang; David Perrault; Milan Stevanovic; Alidad Ghiassi
Journal:  Ann Transl Med       Date:  2016-12

2.  Three-dimensional printing of MRI-visible phantoms and MR image-guided therapy simulation.

Authors:  Dimitris Mitsouras; Thomas C Lee; Peter Liacouras; Ciprian N Ionita; Todd Pietilla; Stephan E Maier; Robert V Mulkern
Journal:  Magn Reson Med       Date:  2016-02-11       Impact factor: 4.668

3.  Evaluation of ultrasonic scattering in agar-based phantoms using 3D printed scattering molds.

Authors:  Antria Filippou; Christakis Damianou
Journal:  J Ultrasound       Date:  2022-01-08

4.  Design and validation of a phantom for transcranial ultrasonography.

Authors:  Denis Leonov; Maria Kodenko; Daria Leichenco; Anastasia Nasibullina; Nicholas Kulberg
Journal:  Int J Comput Assist Radiol Surg       Date:  2022-04-10       Impact factor: 3.421

Review 5.  3D printing in neurosurgery: A systematic review.

Authors:  Michael Randazzo; Jared M Pisapia; Nickpreet Singh; Jayesh P Thawani
Journal:  Surg Neurol Int       Date:  2016-11-14

6.  Printed Receive Coils with High Acoustic Transparency for Magnetic Resonance Guided Focused Ultrasound.

Authors:  Joseph Corea; Patrick Ye; Dongjin Seo; Kim Butts-Pauly; Ana Claudia Arias; Michael Lustig
Journal:  Sci Rep       Date:  2018-02-21       Impact factor: 4.379

7.  Acoustic and thermal characterization of agar based phantoms used for evaluating focused ultrasound exposures.

Authors:  Georgios Menikou; Christakis Damianou
Journal:  J Ther Ultrasound       Date:  2017-06-01

8.  MR relaxation times of agar-based tissue-mimicking phantoms.

Authors:  Anastasia Antoniou; Leonidas Georgiou; Theodora Christodoulou; Natalie Panayiotou; Cleanthis Ioannides; Nikolaos Zamboglou; Christakis Damianou
Journal:  J Appl Clin Med Phys       Date:  2022-04-12       Impact factor: 2.243

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.