Dual-source RF transmission in cardiac SSFP imaging at 3 T: systematic spatial evaluation of image quality improvement compared to conventional RF transmission.

Michael Rasper1, Bettina M Gramer2, Marcus Settles3, Karl-Ludwig Laugwitz4, Tareq Ibrahim5, Ernst J Rummeny6, Armin Huber7.   

Abstract

The purpose of this investigation was to systematically evaluate the spatial distribution of image quality improvement with dual-source radiofrequency (RF) transmission in cardiac steady-state free precession sequences at 3.0 T. Imaging with and without dual-source RF transmission was performed in 30 patients. Contrast-to-noise ratio for the left ventricular myocardium was significantly higher using dual-source RF transmission, but improvement was not uniformly distributed. The posterior myocardium showed significantly less contrast-to-noise ratio gain than all other cardiac regions. Signal-to-noise ratio increase was higher in the right than in the left ventricle. Subjective image quality was significantly enhanced by parallel RF transmission.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Cardiac MRI; Dual-source RF transmission; MultiTransmit; SSFP

Mesh:

Substances:

Year:  2014        PMID: 25481220     DOI: 10.1016/j.clinimag.2014.11.007