A protocol for EBT3 radiochromic film dosimetry using reflection scanning.

Pavlos Papaconstadopoulos1, Gyorgy Hegyi1, Jan Seuntjens1, Slobodan Devic2.   

Abstract

PURPOSE: To evaluate the performance of the EBT3 radiochromic film dosimetry system using reflection measurements and to suggest a calibration protocol for precise and accurate reflection film dosimetry.
METHODS: A set of 14 Gafchromic EBT3 film pieces were irradiated to various doses ranging from 0 to 8 Gy and subsequently scanned using both the reflection and transmission mode. Scanning resolution varied from 50 to 508 dpi (0.5-0.05 mm/pixel). Both the red and green color channels of scanned images were used to relate the film response to the dose. A sensitivity, uncertainty, and accuracy analysis was performed for all scanning modes and color channels. The total uncertainty, along with the fitting and experimental uncertainty components, was identified and analyzed. A microscope resolution target was used to evaluate possible resolution losses under reflection scanning. The calibration range was optimized for reflection scanning in the low (<2 Gy) and high (>2 Gy) dose regions based on the reported results.
RESULTS: Reflection scanning using the red channel exhibited the highest sensitivity among all modes, being up to 150% higher than transmission mode in the red channel for the lowest dose level. Furthermore, there was no apparent loss in resolution between the two modes. However, higher uncertainties and reduced accuracy were observed for the red channel under reflection mode, especially at dose levels higher than 2 Gy. These uncertainties were mainly attributed to saturation effects which were translated in poor fitting results. By restricting the calibration to the 0-2 Gy dose range, the situation is reversed and the red reflection mode was superior to the transmission mode. For higher doses, the green channel in reflection mode presented comparable results to the red transmission.
CONCLUSIONS: A two-color reflection scanning protocol can be suggested for EBT3 radiochromic film dosimetry using the red channel for doses less than 2 Gy and the green channel for higher doses. The precision and accuracy are significantly improved in the low dose region following such a protocol.

Entities:  

Mesh:

Year:  2014        PMID: 25471974     DOI: 10.1118/1.4901308

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  5 in total

1.  Scanning methodology for contact lens-type ocular in vivo dosimeter (CLOD) dosimetry applying a silicone material.

Authors:  Jaeman Son; Jin Dong Cho; Seongmoon Jung; Chang Heon Choi; Jong Min Park; Jung-In Kim
Journal:  Radiat Oncol       Date:  2022-05-07       Impact factor: 4.309

2.  Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using Gafchromic™ EBT3 films.

Authors:  Ning Wen; Siming Lu; Jinkoo Kim; Yujiao Qin; Yimei Huang; Bo Zhao; Chang Liu; Indrin J Chetty
Journal:  Radiat Oncol       Date:  2016-10-04       Impact factor: 3.481

3.  Gafchromic EBT3 film dosimetry in electron beams - energy dependence and improved film read-out.

Authors:  Petri Sipilä; Jarkko Ojala; Sampsa Kaijaluoto; Ilkka Jokelainen; Antti Kosunen
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

4.  Comparison of EBT and EBT3 RadioChromic Film Usage in Parotid Cancer Radiotherapy.

Authors:  M T Bahreyni Toossi; F Khorshidi; M Ghorbani; N Mohamadian; D Davenport
Journal:  J Biomed Phys Eng       Date:  2016-03-01

5.  Clinical utility of Gafchromic film in an MRI-guided linear accelerator.

Authors:  Ilma Xhaferllari; Joshua P Kim; Ruchira Liyanage; Chang Liu; Dongsu Du; Anthony Doemer; Indrin J Chetty; Ning Wen
Journal:  Radiat Oncol       Date:  2021-06-26       Impact factor: 3.481

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.