Literature DB >> 25471963

The influence of anatomical noise on optimal beam quality in mammography.

Björn Cederström1, Erik Fredenberg1.   

Abstract

PURPOSE: Beam-quality optimization in digital mammography traditionally considers detection of a target obscured by quantum noise in a homogeneous background. This does not correspond well to the clinical imaging task because real mammographic images contain a complex superposition of anatomical structures, resulting in anatomical noise that may dominate over quantum noise. The purpose of this paper is to assess the influence on optimal beam quality in mammography when anatomical noise is taken into account.
METHODS: The detectability of microcalcifications and masses was quantified using a theoretical ideal-observer model that included quantum noise as well as anatomical noise and a simplified model of a photon-counting mammography system. The outcome was experimentally verified using two types of simulated tissue phantoms.
RESULTS: The theoretical model showed that the detectability of tumors and microcalcifications behaves differently with respect to beam quality and dose. The results for small microcalcifications were similar to what traditional optimization methods yield, which is to be expected because quantum noise dominates over anatomical noise at high spatial frequencies. For larger tumors, however, low-frequency anatomical noise was the limiting factor. Because anatomical structure noise has similar energy dependence as tumor contrast, the optimal x-ray energy was found to be higher and the useful energy region was wider than traditional methods suggest. A simplified scalar model was able to capture this behavior using a fitted noise mixing parameter. The phantom measurements confirmed these theoretical results.
CONCLUSIONS: It was shown that since quantum noise constitutes only a small fraction of the noise, the dose could be reduced substantially without sacrificing tumor detectability. Furthermore, when anatomical noise is included, the tube voltage can be increased well beyond what is conventionally considered optimal and used clinically, without loss of image quality. However, no such conclusions can be drawn for the more complex mammographic imaging task as a whole.

Entities:  

Mesh:

Year:  2014        PMID: 25471963     DOI: 10.1118/1.4900611

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  2 in total

1.  The effect of anatomical noise on perception of low contrast in intra-oral radiographs: an in vitro study.

Authors:  Lars Olsson; Mats Nilsson; Björn Svenson; Kristina Hellén-Halme
Journal:  Dentomaxillofac Radiol       Date:  2016-02-19       Impact factor: 2.419

2.  Virtual assessment of stereoscopic viewing of digital breast tomosynthesis projection images.

Authors:  Gezheng Wen; Ho-Chang Chang; Jacob Reinhold; Joseph Y Lo; Mia K Markey
Journal:  J Med Imaging (Bellingham)       Date:  2018-01-17
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.