Literature DB >> 25469990

Removal of active-fixation coronary sinus leads using a mechanical rotation extraction device.

Alexander Kypta1, Hermann Blessberger, Karim Saleh, Simon Hönig, Jürgen Kammler, Clemens Steinwender.   

Abstract

BACKGROUND: Active fixation coronary sinus (CS) leads are widely used in cardiac resynchronization therapy (CRT). Due to their low dislodgement rates they are an attractive option for implanters. However, extraction of active fixation leads is a complex procedure that bears potential fatal risks for patients.
METHODS: We analyzed all patients undergoing StarFix® (Medtronic Inc., Minneapolis, MN, USA) extraction because of severe infection at our institution. Indication for extraction was severe device infection. Procedural outcomes as well as patient characteristics are reported. For removal of the leads a 9-Fr Evolution® (Cook Intravascular Inc., Leechburg, PA, USA) mechanical rotation extraction sheath was used.
RESULTS: Between 2011 and 2014, six patients underwent extraction of StarFix® leads. One patient was female, whereas the others were males (mean age 64 ± 12 years). Implant duration was 46.5 ± 8.2 months before removal. All leads could be successfully extracted totally with the Evolution®. In one patient (16.6%) pericardial tamponade occurred immediately requiring surgical intervention. After establishment of cardiopulmonary bypass a 5-mm sharp cut in the lateral vein of the CS could be identified. It was fixed and the patient survived. On all leads, significant tissue growth between the fixation lobes could be observed.
CONCLUSION: As illustrated by our cases, Attain StarFix® leads implanted years ago can be extracted with the help of a mechanical extraction sheath. However, this procedure bears potential risks and should only be performed with a cardiac surgery standby. If clinicians evaluate the implantation of Attain StarFix® leads in patients who are young or at high risk for device infection, they should be aware of these findings.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  active fixation leads; device infection; lead extraction; resynchronization therapy

Mesh:

Year:  2014        PMID: 25469990     DOI: 10.1111/pace.12552

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  First experience in quadripolar active fixation coronary sinus lead extraction: a case report.

Authors:  Elhosseyn Guella; Michael Brack; Khalid Abozguia; Christopher John Cassidy
Journal:  Eur Heart J Case Rep       Date:  2020-07-22

2.  Editorial: What makes transvenous extraction more difficult?

Authors:  Hideo Okamura
Journal:  J Cardiol Cases       Date:  2015-12-04

3.  Safety and effectiveness of coronary sinus leads extraction - single high-volume centre experience.

Authors:  Andrzej W Kutarski; Wojciech Jacheć; Łukasz Tułecki; Konrad Tomków; Paweł Stefańczyk; Wojciech Borzęcki; Dorota Nowosielecka; Marek Czajkowski; Maciej Polewczyk; Anna Polewczyk
Journal:  Postepy Kardiol Interwencyjnej       Date:  2019-09-18       Impact factor: 1.426

4.  Hybrid transvenous and surgical approach for the extraction of coronary sinus leads: A case series.

Authors:  Ashraf Ahmed; Gianmarco Arabia; Luca Bontempi; Manuel Cerini; Francesca Salghetti; Claudio Muneretto; Gianfranco Mitacchione; Antonino Milidoni; Antonio Curnis
Journal:  Pacing Clin Electrophysiol       Date:  2022-06-10       Impact factor: 1.912

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.