Unexpected triggering of the dialysate blood leak detector by haemolysis.

E Lindley, D Finney, P Jones, A Lewington, A O'Reagan, G Webb.   

Abstract

This case showed that it is possible for haemoglobin released by haemolysis in the extracorporeal circuit to pass through a high permeability (albumin retaining) dialyser membrane. In the incident described, the blood leak detector of the dialysis machine was activated before the patient became symptomatic. Haemolysis should be considered as a possible cause of blood leak alarms during dialysis with high flux membranes.

Entities:  

Keywords:  Blood leak alarm; Haemodialysis,; Haemolysis,; High-permeability membrane,

Mesh:

Substances:

Year:  2014        PMID: 25468361     DOI: 10.1179/2295333714Y.0000000105