Influences of spherical phantom heterogeneities on dosimetric charactristics of miniature electronic brachytherapy X-ray sources: Monte Carlo study.

Hojatollah Karimi Jashni1, Habib Safigholi2, Ali S Meigooni3.   

Abstract

Dose calculations in current brachytherapy treatment planning systems (TPS) are commonly based on TG-43U1 formalism. These TPS are obtained by superposition principle of single-source dosimetric parameters in liquid water, neglecting the effects of tissue heterogeneity. In this work, the sensitivity of the TG-43U1 based radial dose function (g(r)) of Miniature Electronic Brachytherapy X-ray Sources (MEBXS) to bone-heterogeneity was examined. To quantify the heterogeneity effects for g(r), a series of Monte Carlo (MC) based radiation transport simulations at the center of homogeneous and heterogeneous spherical phantoms were performed using the MCNP5 code. The ratio of the g(r) in the heterogeneius phantom to the uniform soft tisuue phantom as a function of the bone thickness was determined. These results indicated that for 40keV beam, the maximum ratios for thicknesses of 1cm and 2cm were 3.36 and 3.27, respectively. These values changed to 4.28 and 4.06, for 60keV beam, respectively. Introduction of 0.5cm or 1cm red marrow, into the interior of the cortical bone changed the maximum variations to, 3.54, and 3.57 for 40keV, and 4.28, and 4.25, for 60keV, respectively.
Copyright © 2014 Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Heterogeinety; MEBXS; Miniature electronic brachytherapy X-ray source; Monte Carlo; TG-43U1

Year:  2014        PMID: 25464186     DOI: 10.1016/j.apradiso.2014.10.014

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  4 in total

1.  Anode optimization for miniature electronic brachytherapy X-ray sources using Monte Carlo and computational fluid dynamic codes.

Authors:  Masoud Khajeh; Habib Safigholi
Journal:  J Adv Res       Date:  2015-04-20       Impact factor: 10.479

2.  Assessment of two hemispherical and hemispherical-conical miniature sources used in electronic brachytherapy using Monte Carlo Simulation.

Authors:  Barat Barati; Mansour Zabihzadeh; Mohammad Javad Tahmasebi Birgani; Nahid Chegini; Mojtaba Hoseini Ghahfarokhi; Jafar Fatahiasl
Journal:  Electron Physician       Date:  2017-02-25

3.  HDR brachytherapy in vivo source position verification using a 2D diode array: A Monte Carlo study.

Authors:  Joel Poder; Dean Cutajar; Susanna Guatelli; Marco Petasecca; Andrew Howie; Joseph Bucci; Anatoly Rosenfeld
Journal:  J Appl Clin Med Phys       Date:  2018-06-01       Impact factor: 2.102

4.  Optimum radiation source for radiation therapy of skin cancer.

Authors:  Habib Safigholi; William Y Song; Ali S Meigooni
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.